Functional Verification Test
Functional
Verification Test
Procedure
Note
Functional
Verification Test
Example
Chapter 4
The Functional Verification Test quickly verifies that the module is
operational. Perform this test any time to verify or check the operation.
Perform this test before beginning any Quick Verification or Performance
Verification Tests.
Use the following procedure to verify the module's operation.
Install the module in the mainframe and apply power.
1.
Verify the module passed its power-on self-test by executing the
2.
SYStem:ERRor? query and entering the results.
A passed self-test is indicated by the return
+0, " No Error"
any other return value indicates a power-on failure.
If an incorrect module address is used, the module will not respond.
Verify the module's address before troubleshooting.
Execute the full self-test by sending the *TST? query and entering
3.
the result.
A passed self-test is indicated by the return
+0
Any other return value indicates a failure. Additional information
about the self-test results and self-test error codes is given in
Chapter 6 beginning on page 73.
An example of the Functional Verification Test is shown beginning on
page 27.
Verification Tests 33]