PROCEDURE
1.
Set up the equipment as shown in figure 4-3 and
set the controls as follows:
5.
Set NEG-POS-MAN switch to NEG and repeat
steps 3 and 4 observing the pulses. The Model
8081 A output shou Id trigger on the negative
half-cycles of the oscillator output.
8081 A:
SPECIFICATION
RATE range
100KHz-10KHz
Rate VERNIER
CCW
Operating Mode switch
TRIG
EXT INPUT LEVEL
center
NEG-POS-MAN switch
pas
4-10
EXTERNAL GATE MODE (SLOW)
TEST
Oscilloscope:
Gate signal turns rep. rate generator on synchronously.
Sensitivity (both channels)
1V/div
Trigger. ,
external
Timebase main
20ps/div
Oscillator:
DESCRIPTION
The function of the Model 8081 A is tested in external
gate mode using an external sinewave generator to apply
a gate signal of 10KHz.
Frequency
10KHz
Amplitude
1VRMs
3.
Check that during the positive slope of the oscilla·
tor output a positive pu Ise output occurs.
4.
Turn EXT INPUT LEVEL control from positive to
negative and observe waveforms.
so
OHM TEE
SQOHM
TERMINATION
REAl. ,IME
OSCU.LOSCOPE
Figure 4-4. External Gate Mode (Slow) Test Setup
EQUIPMENT
Oscillator
, HP 651 B
Real Time Oscilloscope
HP 182C Mainframe
with 1804A and 1821 A plug-ins.
50 Ohm Feedthru Termination
HP 10100C
i*--
Propagation delay
I
Display oscillator output and Model 8081 A output
on screen
8081A
OUTPUT
2.
OSCILLATOR
OUTPUT
PROCEDURE
OSCILLATOR
OUTPUT
8081 A
OUTPUT
Threshold lever""" +1 V
-1V
1.
Set up the equipment as shown in figure 4-4 and
set the controls as follows:
8081A:
RATE range. ,
100KHz-10KHz
Rate VERNIER
CCW
Operating Mode switch
GATE EXT
EXT INPUT LEVEL
center
NEG-paS-MAN switch
pas