This recorder has been calibrated at the factory. It retains the temperature accuracy
and calibration even when there is a power supply interruption. Should there be a need
to recalibrate the recorder, please follow the procedures below.
a)Let the unit run continuously until it reaches a stable condition. The recorder has
registered temperature for at least two hours.
b)
Place a standard thermocouple in a 100 ml container filled with 10% glycerin solution.
Place the container and the temperature recorder sensor in the unit. Do not place the
recorder sensor in contact with any liquid.
c)After four hours of soak time at the stable freezer temperature, the container
temperature should match the freezer temperature. Then compare the thermocouple
reading with the temperature recorder reading. If there is a discrepancy, use the button
(#1) or the button (#2) on the recorder to move the pen on the chart to reflect the
correct temperature.
Note: The ink pen will operate 5 seconds after the button is released.
Caution: The temperature recorder is preset during production, so resetting of the
temperature recorder is not recommended unless absolutely necessary.
22

Safety Precautions

This manual and labels attached to the product include important safety precautions
required to operate this product. Customers are advised to follow instructions
provided to avoid potential damage to product and injury to operators.
Note: Use this product only in the way described in the product literature and this
manual. Before using it, verify that this product is suitable for its intended use.
Do not modify system components or use unauthorized parts as this will void the
product warranty.
This sign is directly related to operator safety and equipment safety. The
contents and instructions that follow the sign must be strictly followed.
This sign refers to prohibited actions. Violation may result in damage to the
freezer or compromise the operator's safety.
This label is adhered to a cover that protects a high voltage area to
reduce the chance of electric shock. Access to the cover should only be
handled by a qualified engineer or technician.
NOTICE!
1.Freezer must be placed at least 200mm away from the wall to allow maximum air
circulation.
2.
Changing of the Refrigerants should be done according to the Service Manual
NOTICE!
For the most efficient energy consumption and optimized compressor life,the
temperature should be set at the highest temperature suitable for the samples to be
stored,normally set at -50
to -80 .
When power to the unit is on, switch on the battery charger. When the unit
is not in use switch power off and then switch off the battery charger..
Wear Cryogloves before
opening
the door.
Grounding mark