Table 9. mA Stepping Pushbuttons
Pushbutton
X
Adjusts up to the next higher 25 %
M
step
% STEP
% STEP
Adjusts down to the next lower 25
% step
G
W
F
Sets to 100 % value
Span Check
Span Check
Sets to 0 % value
E
26
Adjustment
Table 10. mA Step Values
Value (for each span setting)
Step
4 to 20 mA
0 %
4.000 mA
25 %
8.000 mA
50 %
12.000 mA
75 %
16.000 mA
100 %
20.000 mA
125 %
24.000 mA
120 %

Auto Ramping the mA Output

Auto ramping gives the ability to continuously apply a
varying current stimulus from the meter to a transmitter,
while hands remain free to test the response of the
transmitter. Select either sourcing or simulating by
choosing the SOURCE or SIMULATE jacks.
0 to 20 mA
0.000 mA
5.000 mA
10.000 mA
15.000 mA
20.000 mA
24.000 mA