INSTRUCTIONS Labels and Symbols Important Information — Please Read Before THUNDERBEAT 5 mm, 45 cm, Pistol Grip Chapter 1 Checking the Package Contents TB-0545PC Chapter 2 Instrument Nomenclature and Specifications THUNDERBEAT 5 mm, 35 cm, Pistol Grip TB-0535PC Chapter 3 Preparation and Inspection THUNDERBEAT 5 mm, 45 cm, Inline Grip TB-0545IC...
Contents Contents Labels and Symbols ......................1 Important Information — Please Read Before Use ............2 Intended use .......................... 2 Application of high-frequency treatment ................2 Instruction manual ......................... 3 User qualification ........................3 Instrument compatibility ......................3 Reprocessing and storage ..................... 3 Repair and modification ......................
Page 4
Contents Chapter 5 Storage and Disposal ............49 Storage ........................49 Disposal ........................49 Chapter 6 Troubleshooting ..............51 Troubleshooting ..................... 51 Appendix ....................53 System chart ........................53 THUNDERBEAT series ....................... 55 EMC information ......................57 Temperature of the shaft, grasping section and probe tip during activation ..... 61 TB-0545PC/IC/FC, TB-0535PC/IC/FC, TB-0520IC, TB-0510IC INSTRUCTION MANUAL...
Labels and Symbols Labels and Symbols Safety-related labels and symbols are attached to the instrument at the locations shown below. If labels and symbols are missing or illegible, contact Olympus. THUNDERBEAT instrument (Pistol grip) Symbol 1 for THUNDERBEAT Lot number ...
Important Information — Please Read Before Use Important Information — Please Read Before Use Intended use The THUNDERBEAT hand instruments are intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer, (TD-TB400). Seal &...
Before use, thoroughly review this manual and the manuals of all equipment which will be used during the procedure and use the equipment as instructed. Keep all instruction manuals in a safe, accessible location. If you have any questions and comments about any information in this manual, contact Olympus. User qualification This manual does not explain or discuss clinical surgical procedures.
Important Information — Please Read Before Use Repair and modification This THUNDERBEAT instrument contains no user-serviceable parts. Do not modify or attempt to repair; patient or surgeon, surgical staff injury and/or equipment damage can result. Signal words The following signal words are used throughout this manual: Indicates an imminently hazardous situation which, if not avoided, will result in DANGER death or serious injury.
Important Information — Please Read Before Use Dangers, warnings, and cautions Follow the dangers, warnings and cautions described below when handling the THUNDERBEAT instrument. This information is supplemented by the dangers, warnings and cautions described in each chapter. DANGER •...
Page 10
Important Information — Please Read Before Use TB-0545PC/IC/FC, TB-0535PC/IC/FC, TB-0520IC, TB-0510IC INSTRUCTION MANUAL...
If the sterile pack and/or the THUNDERBEAT instrument are damaged, a component is missing or you have any questions, do not use the THUNDERBEAT instrument; contact Olympus for assistance. NOTE Torque wrench and stabilizer are also available separately (MAJ-1983 Torque Wrench/Stabilizer Kit).
2.1 Symbols and descriptions Chapter 2 Instrument Nomenclature and Specifications Symbols and descriptions Ch.2 Symbol Description Symbol Description Refer to instructions. Single use only Use by (expiration date) Sterilized using ethylene oxide Sterilization lot number Lot number Keep away from sunlight Keep dry Do not use if package is Do not resterilize...
2.2 Nomenclature Nomenclature THUNDERBEAT instrument (Pistol grip) Ch.2 Nomenclature Description Probe tip During output, ultrasonic vibration or bipolar output is activated. Grasping section During output, bipolar output is activated. (jaw) Shaft Shaft covered with insulated tube. Rotation knob Rotates the shaft. Transducer Connect the THUNDERBEAT Transducer.
Page 15
2.2 Nomenclature THUNDERBEAT instrument (Inline grip) Ch.2 Nomenclature Description Probe tip During output, ultrasonic vibration or bipolar output is activated. Grasping section During output, bipolar output is activated. (jaw) Shaft Shaft covered with insulated tube. Rotation knob Rotates the shaft. Control handle By opening or closing the control handle, grasping section is opened or closed.
Page 16
2.2 Nomenclature THUNDERBEAT instrument (Front-actuated grip) Ch.2 Nomenclature Description Probe tip During output, ultrasonic vibration or bipolar output is activated. Grasping section During output, bipolar output is activated. (jaw) Shaft Is inserted into the abdominal cavity. Rotation knob Rotates the shaft. Transducer Is connected to the transducer.
Page 17
2.2 Nomenclature Closeup of the distal end of the shaft Ch.2 Nomenclature Description Upper surface of Coated with insulating material excluding the gray area. the grasping section Non-insulated area – PTFE pad – TB-0545PC/IC/FC, TB-0535PC/IC/FC, TB-0520IC, TB-0510IC INSTRUCTION MANUAL...
Page 18
2.2 Nomenclature Torque wrench (provided) For connecting the transducer (available separately) to the THUNDERBEAT instrument. NOTE The torque wrench is also available separately (MAJ-1983 Torque Wrench/Stabilizer Kit). Ch.2 Nomenclature Description Head Mounts on the knob of the instrument. Assembly index THIS SIDE UP.
Page 19
2.2 Nomenclature Stabilizer (provided) For connecting the transducer (available separately) to the THUNDERBEAT instrument. NOTE The stabilizer is also available separately (MAJ-1983 Torque Wrench/Stabilizer Kit). Ch.2 Nomenclature Description Grip During use, it is held. TB-0545PC/IC/FC, TB-0535PC/IC/FC, TB-0520IC, TB-0510IC INSTRUCTION MANUAL...
Page 20
2.2 Nomenclature THUNDERBEAT Transducer (separately available) Ch.2 Nomenclature Description THUNDERBEAT Is connected to the instrument. instrument connection Transducer cord Electrical signal is output from the ultrasonic generator to the instrument. Transducer plug Connects to the THUNDERBEAT socket of the ultrasonic generator. Symbol Indicates that this is a THUNDERBEAT transducer.
2.3 Specifications Specifications Transportation, storage, and operating environments Operating Ambient temperature 10 – 40C (50 – 104F) environment Relative humidity 30 – 85% Atmospheric Ch.2 70 – 106 kPa pressure Altitude 3000 m or less Transportation and Ambient temperature –40 to +60C (–40 to +140F) storage environment Relative humidity 10 –...
Page 22
2.3 Specifications Medical Devices Directive This device complies with the requirements of Directive 93/42/EEC concerning medical devices. Classification: Class II b Applied standard IEC 60601-1-2: 2007 • This instrument complies with the EMC standard for medical electrical equipment; edition 3 (IEC 60601-1-2: 2007). However, when connecting to an instrument that complies with the EMC Ch.2 standard for medical electrical equipment, edition 1...
Chapter 8, “Troubleshooting” in the instruction manual for the ultrasonic generator. If the irregularity is still not resolved after troubleshooting, contact Olympus. Using the THUNDERBEAT instrument while any irregularity is observed, may cause malfunction and may injure the surgeon, surgical staff and/or patient.
Page 24
3.1 Warnings and cautions: preparation and inspection WARNING • To prevent malfunction or damage, be careful not to apply excessive force to the transducer connecting sections during assembly. If the connection is difficult, it is probable that a part is crushed, bent or otherwise deformed. Check the THUNDERBEAT instrument sufficiently and do not use it if an irregularity is observed.
3.2 Preparation of the equipment Preparation of the equipment Prepare the THUNDERBEAT instrument, the ultrasonic generator, compatible electrosurgical generator, the transducer compatible ancillary equipment (shown in the “System chart” in the Appendix) and other equipment to be used with the THUNDERBEAT instrument. The required personal protective equipment, such as eyewear, a face mask, moisture-resistant clothing, and chemical-resistant gloves that fit properly to prevent skin exposure.
Page 26
3.3 Connection of the transducer • When attaching the head of the torque wrench to the rotation knob, do not damage the shaft insulation. Confirm that the “Use by” date has not passed and the sterile pack shows no signs of damage such as a break, peeled stickers, contamination, or the presence of moisture.
Page 27
3.3 Connection of the transducer Turn the rotation knob clockwise with a light force, holding the transducer by the other hand until the rotation knob stops. Front-actuated Pistol grip Inline grip grip Ch.3 Figure 3.2 Mount the stabilizer on the transducer. Front-actuated Pistol grip Inline grip...
Page 28
3.3 Connection of the transducer Grasp the stabilizer grip mounted on the transducer and attach the head of the torque wrench down onto the rotation knob. Front-actuated Pistol grip Inline grip grip Ch.3 Figure 3.5 Grasp both the grip of the torque wrench and the stabilizer mounted on the transducer and then rotate the rotation knob clockwise slowly with the torque wrench until the torque wrench clicks.
Page 29
3.3 Connection of the transducer CAUTION • In the case of the inline grip model, do not hold the grip handle and transducer together while rotating the rotation knob with the torque wrench. Ch.3 Inline grip Figure 3.7 • Do not dispose of the torque wrench and the stabilizer but keep them in a sterile condition, because they will be used for changing and withdrawing the transducer during the procedure and post-procedure.
Electrosurgical Generator USG-400 ESG-400 Table 3.1 Prepare and inspect the generators according to their instruction manuals. Should any irregularity be observed, do not use them; contact Olympus. Connection to ultrasonic generator Connecting/disconnecting the transducer plug to/from the ultrasonic generator...
Page 31
3.5 Connection to ultrasonic generator CAUTION • Do not attempt to clean the contacts inside the transducer plug with a sharp object such as the tips of tweezers. Also, when cleaning the contacts with a brush, do not poke or rub the contacts with the metal tip on the bristles of the brush. Otherwise, deformation or damage of the contact will cause conduction failure, making energy delivery impossible.
If these symbols are not identical, the ultrasonic generator or transducer may malfunction. If this occurs, immediately stop using the system and contact Olympus. Prepare and connect ancillary equipment to be used with the THUNDERBEAT by referring to their respective instruction manuals.
Page 33
3.6 Inspection of ancillary equipment Output mode Output Output Level Energy Main Purpose Characteristics Mode (Default Level) • This mode is same as the setting 39 of FineCoag mode of the ESG-400. • The output setting that has been validated Tissue for vessel sealing is level 1.
Page 34
3.6 Inspection of ancillary equipment The cutting speed indicator bars THUNDERBEAT Set screen of the ultrasonic generator Figure 3.9 Ch.3 TB-0545PC/IC/FC, TB-0535PC/IC/FC, TB-0520IC, TB-0510IC INSTRUCTION MANUAL...
3.6 Inspection of ancillary equipment Verification of the high-frequency (RF bipolar) output WARNING • The output should always be verified outside the body cavity to avoid unintended tissue damage. Otherwise, tissues may be burnt. • Verify the high-frequency (RF bipolar) energy delivery before surgery. Otherwise, the THUNDERBEAT may not function properly during the surgery.
Page 36
3.6 Inspection of ancillary equipment NOTE When the high frequency output activated for a long time (more than 15 seconds) the SEAL incomplete error is occur, but the THUNDERBEAT doesn’t have any problem. Because it is normal. Prepare a sterile container (like a bowl) with dimensions similar to those shown in Figure 3.10.
Page 37
3.6 Inspection of ancillary equipment Confirm the following items during the SEAL mode output: • During activating, the activation in “All screen” or “Set screen” is displayed on the touch-screen of the ultrasonic generator, no error window (see Figure 3.12). •...
3.6 Inspection of ancillary equipment Inspecting the combination with the trocar WARNING • The outer diameter of the shaft of the THUNDERBEAT instrument is 5.5 mm. Use a trocar with a corresponding size. Be sure to confirm the THUNDERBEAT instrument and trocar compatibility prior to use.
If the irregularity remains after troubleshooting, replace the transducer. If this does not resolve the issue, replace the THUNDERBEAT instrument. If the irregularity remains unresolved, contact Olympus. • Use the THUNDERBEAT instrument properly. Improper use may cause the probe...
Page 40
4.1 Warnings and cautions Typical damaged phenomenon of PTFE pad Partial separation of the PTFE pad Exposed metal part due to severe wear or the PTFE pad Figure 4.1 Ch.4 WARNING • The probe tip, the PTFE pad, and the grasping section of the THUNDERBEAT instrument wear due to ultrasonic vibrations.
Page 41
4.1 Warnings and cautions WARNING • If the THUNDERBEAT is used to treat a patient with blood hypertension, coronary disease, arteriosclerosis, diabetes, and/or cirrhosis, or a patient with blood vessel irregularities such as calcification, sufficient sealing performance may not be possible.
Page 42
4.1 Warnings and cautions WARNING • When applying energy to a blood vessel, confirm that the distal end of the shaft is not submerged in fluids such as pooled blood or saline. Such conditions could potentially reduce the effectiveness of the THUNDERBEAT and could cause unintended tissue damage.
Page 43
4.1 Warnings and cautions WARNING • If the alarm tone sounds and an error window is displayed during the procedure, immediately stop the procedure. During surgical procedures, withdraw the THUNDERBEAT instrument from the body cavity. Do not remove the transducer plug from the ultrasonic generator.
Page 44
4.1 Warnings and cautions WARNING • During TLH colpotomy or LSH amputation, avoid inserting the probe tip vertically and deeply into the uterine cervix and activating the THUNDERBEAT instrument. Do not overfill the grasping section with a large bite of tissue. This may result in damage to the probe tip.
Page 45
4.1 Warnings and cautions WARNING • Whenever possible, avoid contacting the shaft other than the grasping section to the tissue as the temperature of the shaft can become elevated and could cause unintentional burns. Please refer to “Temperature of the shaft, grasping section and probe tip during activation”...
Page 46
4.1 Warnings and cautions WARNING • Do not pinch or contact the transducer cord with sharp object. Otherwise the cord may break and malfunction, which could cause burns due to a possible leakage of RF Bipolar current. • Before activating the output, be sure that neither the grasping section nor the probe tip comes into contact with the surrounding tissue.
Page 47
4.1 Warnings and cautions WARNING • For hemostasis in the SEAL mode, grasp a bleeding point with the grasping section and the probe tip. Do not use the outside surface of the grasping section for hemostasis because that surface coated with insulating material is not conductive so as not to stop bleeding.
4.2 Operation NOTE • When the THUNDERBEAT instrument is withdrawn from the patient due to the occurrence of an error display or other irregularities, the probe tip may break and fall off due to friction between the trocar and the instrument or due to the stress caused by cleaning the probe tip.
Page 49
4.2 Operation Setting the output level Confirm that the output level displayed on the touchscreen is appropriate for the procedure. Press the plus or minus button to change the output level, if desired. Set the appropriate output level by referring to “ Output mode” on page 29. When both the THUNDERBEAT and SONICBEAT are connected to the ultrasonic generator, press the “THUNDERBEAT”...
Page 50
4.2 Operation Output mode SEAL & CUT mode • In the SEAL & CUT mode, stop the output after finding that the vessel or tissue is transected. • Always grasp a blood vessel in the middle of the grasping section. •...
4.3 Procedure after use Procedure after use CAUTION • Do not withdraw the trocar and the THUNDERBEAT simultaneously. Otherwise, the grasping section or the probe tip may damage the surrounding tissue, or the THUNDERBEAT itself may be damaged. • Do not pull the transducer cord when disconnecting the transducer from the THUNDERBEAT instrument.
Page 52
Figure 4.4 If a THUNDERBEAT instrument cannot be disconnected from the transducer, contact Olympus. Dispose of the THUNDERBEAT instrument, torque wrench, and stabilizer, and clean and sterilize the transducer as described in Chapter 5, “Reprocessing: General Policy” and Chapter 6, “Cleaning, Disinfection, and Sterilization Procedures” of the instruction manual for the transducer.
5.1 Storage Chapter 5 Storage and Disposal Storage WARNING • Do not store the sterile instrument packs in a place where they may be damaged, the stickers may be peeled off or they may be exposed to moisture. Otherwise, sterility of the THUNDERBEAT instrument may be compromised and inflammation of tissue and infection may result.
6.1 Troubleshooting Chapter 6 Troubleshooting Troubleshooting If any irregularity is observed, take proper remedial actions by referring to Chapter 8, “Troubleshooting” in the instruction manual for the ultrasonic generator. WARNING Do not use the THUNDERBEAT instrument on the patient if any irregularity is observed with the equipment.
The recommended combinations of equipment and accessories that can be used with this instrument are listed below. New products released after the introduction of this instrument may also be compatible for use in combination with this instrument. For further details, contact Olympus. WARNING If combinations of equipment other than those shown below are used, the full responsibility is assumed by the medical treatment facility.
Page 58
System chart CISPR 11 Class A Electrosurgical Generator THUNDERBEAT (ESG-400) series (TB-XXXXXX) Communication Cable 0.25 m (MAJ-1871) Docking Fixture (MAJ-1876) THUNDERBEAT Transducer (TD-TB400) Energy cart (TC-E400) Ultrasonic Generator (USG-400) Communication Cable SONICBEAT 10 m (MAJ-1872) Transducer (TD-SB400) Communication Cable 0.25 m (MAJ-1871) App.
EMC information EMC information This model is intended for use in the electromagnetic environments specified below. The user and the medical staff should ensure that it is used only in these environments. Magnetic emission compliance information and recommended electromagnetic environments Emission standard Compliance Guidance...
Page 62
EMC information Electromagnetic immunity compliance information and recommended electromagnetic environments IEC 60601-1-2 Immunity test Compliance level Guidance test level Electrostatic Contact: Same as left Floors should be made of wood, concrete, or 2, 4, 6 kV discharge (ESD) ceramic tile that hardly produces static. If IEC 61000-4-2 floors are covered with synthetic material that Air:...
Page 63
EMC information Cautions and recommended electromagnetic environment regarding portable and mobile RF communications equipment, such as cellular phones IEC 60601-1-2 Compliance Immunity test Guidance test level level Conducted RF 3 Vrms 3 V (V Formula for recommended separation distance IEC 61000-4-6 (150 kHz –...
Page 64
EMC information Recommended separation distance between portable and mobile RF communications equipment and this instrument Separation distance according to frequency of transmitter (m) (calculated as V =3 and E Rated maximum output 150 kHz – 80 MHz 80 MHz – 800 MHz 800 MHz –...
Temperature of the shaft, grasping section and probe tip during activation Temperature of the shaft, grasping section and probe tip during activation Nomenclature Description Grasping section Over 200C and probe tip Distal end of the Over 135C by the continuous activation. shaft Shaft up to 60 mm Over 60C by the continuous activation.
Page 66
Temperature of the shaft, grasping section and probe tip during activation App. TB-0545PC/IC/FC, TB-0535PC/IC/FC, TB-0520IC, TB-0510IC INSTRUCTION MANUAL...