JEOL 2010F MANUAL
Quick check list
1. Check the accelerating voltage, must be at 200 kV (right screen), HT (µA) (left panel) at
0.96-0.97 and Emission (left panel) at 155-160.
2. Check the vacuum sequence (left drawer), V17, V13, V5B and V2 on. Vacuum measured
-5
on SIP (SIP unit, second vacuum meter from the top) must be better than 2 x 10
Pa (blue
numbers, 150 l/s). The vacuum in the gun (SIP unit, vacuum meter at the top) must be better
-6
than x 10
Pa.
3. If needed, fill the reservoir with LN2. Be sure that the viewing screen cover is on before
filling the trap.
4. Insert specimen holder into TEM (Insert holder in airlock, set air/pump switch to pump
(yellow light turns ON), when green light turns ON, rotate fully clockwise in two stages and
push in completely)
5. Check the vacuum sequence again, (left drawer), V17, V13, V5B, V8, V21 and V2 on.
Vacuum measured on SIP (SIP unit, second vacuum meter from the top) must be better than 2
-5
x 10
Pa (blue numbers, 150 l/s). The vacuum in the gun (SIP unit, vacuum meter at the top)
-6
must be better than x 10
Pa.
6. If vacuum is ok, open the valve between the column and the gun by clicking button
VALVE (left panel controls).
7. If needed, align the microscope. Work!

Advertisement

loading
Need help?

Need help?

Do you have a question about the 2010F and is the answer not in the manual?

Questions and answers

Summary of Contents for JEOL 2010F

  • Page 1 JEOL 2010F MANUAL Quick check list 1. Check the accelerating voltage, must be at 200 kV (right screen), HT (µA) (left panel) at 0.96-0.97 and Emission (left panel) at 155-160. 2. Check the vacuum sequence (left drawer), V17, V13, V5B and V2 on. Vacuum measured...
  • Page 2 1. Initial check 1. Check the vacuum sequence (left drawer), V17, V13, V5B and V2 on. Vacuum measured in SIP (SIP unit, second vacuum meter from the top) must be better than 2 x Pa (blue numbers, 150 l/s). The vacuum in the gun (SIP unit, vacuum meter at the top) must be better than x 10 2.
  • Page 3 3. Electron Beam Generation 1. Check the vacuum sequence (left drawer), V17, V13, V5B, V8, V21 and V2 on. Vacuum measured on SIP (SIP unit, second vacuum meter from the top) must be better than 2 x 10 Pa (blue numbers, 150 l/s). The vacuum in the gun (SIP unit, vacuum meter at the top) must be better than x 10 Pa.
  • Page 4 2. Set the SPOT SIZE switch to 1 and focus the beam using the BRIGHTNESS knob. 3. Click Gun button in the right-hand drawer. 4. Center the beam on the fluorescent screen using the SHIFT X and Y knobs. 5. Click Gun button again to deselect it. 6.
  • Page 5 9. Select SHIFT Y. Unless the ratio of the currents in the 1st and 2nd deflection coils is adjusted properly, the caustic spot splits into two parts in the Y direction. Unify the split spot using the SHIFT Y knob. 10.
  • Page 6 If you are the last user of the day and cold-trap was used, heat the trap up to ambient temperature. 1. Remove all the apertures! 2. Insert the heater assembly into the coolant reservoir and insert the heater plug into the HTR socket on the connector box.