...6 Final checks
6 .4 Test tables
Table 14 AK104 (GAH100 – single-range system)
Upper display
mV input
Display
reading
(%)
Zero
See: Zero
See:
(mV)
Display
zero
________
______
Range
Span
See: Span
See:
(mV)
Display
span
________
______
Active
4-20 mA
Relay
on-screen
output
open or
alarm?
reading
closed?
LOW
TB3 9+10 -
TB3 1+2
alarm
4 mA
- Closed
None
TB3 9+10 -
TB3 1+2
20 mA
- Open
6 .5 Test the switches
The switches can be independently verified if required.
• On AK101, use TB4.
• On AK102 use TB3 for the upper switch and TB4 for the lower
swich.
6 .5 .1 Switch position 1
Move switch to position 1.
– On terminals 13 & 14, you should get a closed circuit
– On terminals 13 & 15, you should get an open circuit
– On terminals 13 & 16, you should get an open circuit
6 .5 .2 Switch position 2
Move switch to position 2.
– On terminals 13 & 14, you should get an open circuit
– On terminals 13 & 15, you should get a closed circuit
– On terminals 13 & 16, you should get an open circuit
6 .5 .3 Switch position 3
Move switch to position 3.
– On terminals 13 & 14, you should get an open circuit
– On terminals 13 & 15, you should get an open circuit
– On terminals 13 & 16, you should get a closed circuit