Page 1
Intrinsically Safe Explosion-Proof High-Precision Tuning Fork Scale FZ-Ex Series Manual Operation IMPORTANT • To ensure safe and proper use of the scale, please read this manual carefully. • After reading this manual, store it in a safe place near the scale, so you can review it as needed.
Page 3
Preface Thank you very much for having purchased our dust- and water-proof intrinsic safety explosion-proof structure electronic scale. This document is the Operation Manual for the following dust- and water-proof intrinsic safety explosion-proof structure electronic scale. In the first place, install this product properly referring to the Installation Manual attached separately, and then read this document.
Page 4
Instructions ● The copyright of this document belongs to SHINKO DENSHI CO., LTD. Reprinting or duplicating of all or part of this document without notice shall not be allowed. ● Please note that product improvement or modification may cause partial discrepancy between the product and the description of this document.
Important Notice ・It should be known that this product contains potential danger. And so please be sure to observe this document when installing, operating or servicing this product. ・If the product is used in a manner not specified by the manuals or other accompanying documents, the protection provided by the product may be impaired.
How to use this document ■ Symbols used in this document Understand the meaning of the following symbols and observe the instructions of this document. Symbols Meaning Used for high-risk point concerning the operations that may lead to death or severe physical injury to persons if proper precautions are not taken.
Page 7
■About how to read this document This document consists of the following contents: Describes about operating precautions, names 1 When beginning to use and functions of each section, etc. Please be sure to read this section when using this product for the first time.
Contents Preface ..........................i Important Notice ........................iii How to use this document ....................iv Contents ..........................vi 1 Prior to use ........................1 Precautions ........................1 Names and functions of each section ................4 Performance of operation keys ..................5 How to interpret the display .................... 6 1-4-1 Main LCD ........................
Page 9
Stability discrimination width ..................31 Stability discrimination frequency ................. 32 Response speed ......................32 Weight renewal interval ....................33 5 User information setting ....................34 Hierarchy of user information setting ................34 Measurer’s ID setting ....................34 Product name ID setting ....................35 Lot number setting ......................
Page 10
Scale ID setting ......................61 Maintenance setting ..................... 61 Date and time setting ....................62 Date indication format ....................62 Output character setting ....................62 Password control ......................63 8-10 Password change ......................63 8-11 Password cancellation history ..................64 8-12 Operation of minimum weight indication ..............
FZ- Ex series operetion manual 1 Prior to use 1 Prior to use Precautions ■ No disassembling or modification. Unless specifically stated in this document, disassembling or modification of this product, mounting or removal of an undesignated component no longer maintains the function of the explosion-proof structure, leading to a serious accident or physical injury.
Page 12
1 Prior to use FZ-Ex series operetion manual ■ Do not move the device with a sample to be weighed set on the scale. That may cause the sample to fall from the weighing pan, leading to a physical injury or destruction of the sample.
Page 13
FZ- Ex series operetion manual 1 Prior to use ■ Do not give a shock to the scale. It may cause breakage or failure. Place a sample to be weighed softly. ■ Do not let an overload situation (o-Err indication) continue. It may cause breakage or failure.
FZ- Ex series operetion manual 1 Prior to use Performance of operation keys Type / name of a key Performance [On / Off] Turns on and off the power for the scale. [Direction] Used for function setting. [Transfer] Used for outputting. [Function F] Used for function calling.
1 Prior to use FZ-Ex series operetion manual How to interpret the display 1-4-1 Main LCD Symbol Name Description Gram Represents gram unit. Kilogram Represents kilogram unit. Percent Lights when in the percent scale mode. Zero point Indicates the zero point. Plus Plus Minus...
FZ- Ex series operetion manual 1 Prior to use 1-4-2 Sub LCDs ( i03 only) ■Upper sub LCD ■Lower sub LCD Name Description Symbol Gram Represents gram unit. Kilogram Represents kilogram unit. Percent Lights when in the percent scale mode. 7-segment display Displays numbers and simple letters.
2 Basic usage FZ-Ex series operetion manual 2 Basic usage Turning on / off the power, and checking for the operation Turn on and off the power for this product. Turn on the power. Turn on the power for the power supply box.
FZ-Ex series operetion manual 2 Basic usage Turn off the power for the scale. Press and hold the [On/Off] key. The product becomes standby status and the symbol 「 」 lights. Pressing holding [On/Off] key obtains the standby status from any operation status. Making a zero adjustment Adjusting the indication to zero is called “zero adjustment.”...
2 Basic usage FZ-Ex series operetion manual 2-2-1 Zero adjustment range when in use Zero adjustment range when in use is limited in this product. The available zero adjustment range when in use is shown below: Model Lower limit (g) Upper limit (g) -9.3 FZ623Ex...
FZ-Ex series operetion manual 2 Basic usage Weighing an object placed on a container (tare) When weighing an object to be weighed with the object placed on a container (tare), the weight of the container must be subtracted from the total weight to get the actual weight of the object to be weighed.
2 Basic usage FZ-Ex series operetion manual Weighing with an object to be weighed added Place an added object to be weighed and weigh the weight of the added object. Performing the tare weight subtraction with the object to be weighed which has been already weighed makes it possible to weigh the mass of a next object to be weighed with the previous object to be weighed remaining placed.
FZ-Ex series operetion manual 2 Basic usage Selecting the main LCD indication The main LCD and the sub LCD can be used in combination with each other. The content of main LCD indication changes in the following sequence: In the case the percent scale function and the adding function are set, selection (switching) of the main LCD indication is available.
2 Basic usage FZ-Ex series operetion manual Basic operation The menu of this product is divided into two as described below: (1)Setting menu The menu to set a variety of functions (2)Execution menu The menu not to set but only to execute the program. 2-7-1 Hierarchy of a setting menu The setting menu of this product is divided into four, from the first layer to the third layer and...
FZ-Ex series operetion manual 2 Basic usage 2-7-2 Operation of the setting menu, setting of various functions To perform settings for various functions from the state of weighing, chiefly execute the following procedure. (1)Press [Function F] key to enter respective setting from the state of weighing. (2)Shift to the intended setting item using the [Direction] key.
Page 26
2 Basic usage FZ-Ex series operetion manual To return to the state of weighing after setting various functions, chiefly execute the following procedure. (1) Press the [Shift] and then [Clear Home] keys at any of the first, second or third layer. -16-...
FZ-Ex series operetion manual 2 Basic usage 2-7-3 Operation of the setting menu, inputting of numeric values Numeric value inputting is limited to seven digits at a maximum. ■Example) When inputting 12345 Input “1”. Press the [Numeric keypad] “1.” Number “1” is displayed on the extreme right like 「...
2 Basic usage FZ-Ex series operetion manual 2-7-4 Operation of the setting menu, inputting of characters ■Operation of character input Press the [Numeric keypad] “ 2 ” Press the [Numeric keypad] “2”. Number “2”is displayed on the extreme 」. left like「2 Press the [Numeric keypad] “2”...
Page 29
FZ-Ex series operetion manual 2 Basic usage Input “ C ” Press “Right” of the [Direction] key. The cursor moves to the right. Press the [Numeric keypad] “2” several times till the indication changes to 「ABC 」. Discontinue inputting. Press “Right” of the [Direction] key. The cursor moves to the right.
3 Functions related to the operation FZ-Ex series operetion manual 3 Functions related to the operation Settings to change the scale operations. Hierarchy of functions related to the operation *1 Unit that can be used on the model of Max 100-300 kg is only 「111.UA.2」 : “kg”. *2 “...
3 Functions related to the operation FZ-Ex series operetion manual Unit setting Unit in the weight mode can be set either in “g” or “kg”. Select a setting menu. Select the unit setting. (“2-7-2 Operation of the setting menu, setting of various functions”...
3 Functions related to the operation FZ-Ex series operetion manual Adding function Weighs a plurality of objects to be weighed in sequence and indicates its total value. The adding function includes two ways of calculating method. Method of weighing objects to be weighed while replacing the objects Addition accumulating function Method of weighing objects to be weighed without replacing the Net adding function...
3 Functions related to the operation FZ-Ex series operetion manual 3-4-1 Weighing by means of the plus side addition Place a first object to be weighed. After 「 」 appears, press [Enter] key. The weighed value is stored and 「 」...
3 Functions related to the operation FZ-Ex series operetion manual 3-4-2 Weighing by means of the minus side addition Place an object to be weighed and perform the tare weight subtraction. After 「 Tare 」 appears, press [Enter] key. The weighed value is stored and 「 」...
3 Functions related to the operation FZ-Ex series operetion manual Comparator function It is possible to preset threshold values and determine whether or not a measured value is within the range defined by the preset values. The comparator function can be used in any scale mode, i.e. weight scale mode, percent scale mode, and coefficient scale mode.
3 Functions related to the operation FZ-Ex series operetion manual 3-5-3 Comparator function setting Select a setting menu. Select comparator function. (“2-7-2 Operation of the setting menu, setting of various functions” and “3-1 Hierarchy of functions related to the operation”) Set comparator function.
3 Functions related to the operation FZ-Ex series operetion manual Buzzer setting This is a convenient function for key inputting and use of the comparator function. Select a setting menu. Select the buzzer setting function. (“2-7-2 Operation of the setting menu, setting of various functions”...
3 Functions related to the operation FZ-Ex series operetion manual Conditions for stability waiting Can not be used. Set when to indicate the weighed value after the zero adjustment or tare weight subtraction; either after or before the weighed value stabilises. Select a setting menu.
3 Functions related to the operation FZ-Ex series operetion manual 3-10 Direct start Setting to the direct start makes it possible to turn on and off the power with the switch on the power supply box without pressing the [On/Off] key. Select a setting menu.
4 Function srelated to the performance FZ-Ex series operetion manual Function srelated to the performance Set the scale indication stability and response speed. Hierarchy of functions related to the performance ” “ can not be used. -30-...
FZ-Ex series operetion manual 4 Function srelated to the performance Zero tracking Setting to the zero tracking function makes it possible to automatically correct the zero point fluctuation caused by the temperature fluctuation, etc. that is likely to occur when “0” is indicated, through which the “0”...
4 Function srelated to the performance FZ-Ex series operetion manual Stability discrimination frequency Can not be used. The larger numeric value is set, the higher stability is obtained. Select a setting menu. Select the stability discrimination frequency. (“2-7-2 Operation of the setting menu, setting of various functions”...
FZ-Ex series operetion manual 4 Function srelated to the performance Weight renewal interval Can not be used. This is a function to output data at regular intervals. Select a setting menu. Select the sampling time. (“2-7-2 Operation of the setting menu, setting of various functions”...
5 User information setting FZ-Ex series operetion manual 5 User information setting Set various user IDs and upper and lower limit values. Hierarchy of user information setting (1) It is possible to register a number for an ID up to a maximum of ten characters. (2) Up to 100 IDs that have been set can be registered individually from “001”...
FZ-Ex series operetion manual 5 User information setting Product name ID setting An ID can be provided for each product name. Select a setting menu. Select the product name ID. (“2-7-2 Operation of the setting menu, setting of various functions” and “5-1 Hierarchy of user information setting”) Set the product name ID.
5 User information setting FZ-Ex series operetion manual Preset tare weight setting Inputting, registration and calling of a preset tare weight value can be performed. 5-6-1 Inputting of a preset tare weight value There are two ways of inputting a reference value and upper and lower limit values as described below: - Numeric value setting method: Inputting a setting value directly via [Numeric keypad] operation...
Page 47
FZ-Ex series operetion manual 5 User information setting The preset tare weight setting can be performed out of the scale mode as a shortcut according to the following procedure: Set a tare weight value. [Numeric value setting method] Press the [Preset tare] key in the scale mode.
5 User information setting FZ-Ex series operetion manual 5-6-2 Registration of a preset tare weight value Nine preset tare weight values can be registered. Set a preset tare weight value. (“2-7-2 Operation of the setting menu, setting of various functions” and “5-1 Hierarchy of user information setting”) Register the preset tare weight value.
FZ-Ex series operetion manual 5 User information setting Setting of a percent scale reference value The weight of an object to be weighed is indicated in percent relative to the reference weight. Set the reference weight by way of either the numeric value setting method, which requires the inputting of a numeric value, or the actual value setting method, which requires the weighing of a sample.
5 User information setting FZ-Ex series operetion manual An indication of “L-Err” signifies that the reference weight is below the limit weight and that the weight is unmeasurable. Percent scale limit weight FZ623Ex 0.1 g FZ3202Ex, FZ6202Ex FZ15001Ex FZ30K0.1GEx, FZ60K0.1GEx FZ100K1GEx, FZ200K1GEx 100 g FZ150K1GFEx, FZ300K1GFEx...
Page 51
FZ-Ex series operetion manual 5 User information setting Select the upper limit value setting. Press the [Direction] key. Select 「37.TS.」. Press the [Direction] key. Select 「372.HI.」. Press the [Enter] key. Set an upper limit value. Input an upper limit value with [Numeric keypad].
Page 52
5 User information setting FZ-Ex series operetion manual The discrimination value setting of the comparator function can be performed out of the scale mode as a shortcut according to the following procedure: Select the reference value setting. (In the case of the relative value discrimination) Press the [Target] key in the scale mode.
FZ-Ex series operetion manual 5 User information setting 5-8-2 Actual value setting method Select a setting menu. Select the discrimination value setting of the comparator function. (“2-7-2 Operation of the setting menu, setting of various functions”) Select the reference value setting. (In the case of the relative value discrimination) Press the [Direction] key.
Page 54
5 User information setting FZ-Ex series operetion manual Set a lower limit value. Press the [Function F] key. Place an object to be weighed that is equivalent to the lower limit value. Press the [Enter] key. The lower limit value is stored. The operation mode returns to the scale mode.
Page 55
FZ-Ex series operetion manual 5 User information setting The discrimination value setting of the comparator function can be performed out of the scale mode as a shortcut according to the following procedure: Select the reference value setting. (In the case of the relative value discrimination) Press the [Target] key in the scale mode.
5 User information setting FZ-Ex series operetion manual Coefficient value setting Can not be used. The value that is obtained by multiplying a measured weight by a predetermined coefficient can be indicated. For example, in the case the coefficient is “2.35” and the measured weight is “2000 g”, a value of “4700 g”...
FZ-Ex series operetion manual 6 External input / output functions 6 External input/output functions This function is used for communication through the external peripheral devices. Hierarchy of the external input/output functions “ ” can not be used. -47-...
6 External input / output functions FZ-Ex series operetion manual Connecter terminal numbers and their functions Input/output to and from an external device such as a personal computer via the RS-232C is available on the power supply box. The RS-232C interface for this product is the D-SUB9P type.
6 External input / output functions FZ-Ex series operetion manual 6-4-4 Input command composition Composed of four characters including a terminator (CR=0DH / LF=0AH). 6-4-5 Transmission procedure Send an input command from an external device to the scale. Since transmission and reception are performed by way of full-duplex communication method, the input command can be transmitted irrespective of the transmission timing from the scale.
FZ-Ex series operetion manual 6 External input / output functions 6-4-6 Command format Please take care not to take alphabetical “O” for Arabic number “0.” When the command of O0 to O7 is sent and the scale setting is switched once, the setting will be maintained, so the same command need not be sent repeatedly.
6 External input / output functions FZ-Ex series operetion manual GZII format This is different from “6-4 GZIII format” only in the operation of the T-command. In the GZII format, the tare weight subtraction / zero adjustment is executed by the T-command. For other specifications, please refer to “6-4 GZIII format”.
FZ-Ex series operetion manual 6 External input / output functions Response 6-6-1 Response command format (when set to the A00, Exx format) Composed of five characters including a terminator (CR=0DH / LF=0AH) 6-6-2 Response command Code(A1) Code(A2) Code(A3) Description Normal completion Command error (Abnormal command received) 0 - 9...
6 External input / output functions FZ-Ex series operetion manual Power supply box communication setting Perform the power supply box communication setting in line with the peripheral device to which the output is transmitted. Select a setting menu. Select the power supply box communication setting.
Page 65
FZ-Ex series operetion manual 6 External input / output functions Set the output conditions. Press the [Direction] key. Select 「413.OC.」. Input a setting value. 「413.OC. 0」 : Output stop 「413.OC. 1」 : Continuous output at all Inputting of the setting value times 「413.OC.
6 External input / output functions FZ-Ex series operetion manual Set the stop bit. Press the [Direction] key. Select 「418.ST.」. Input a setting value. 「418.ST. 1」 :1 bit 「418.ST. 2」 :2 bits Inputting of the setting value Set unused high order digit. Press the [Direction] key.
FZ-Ex series operetion manual Functions related to the lock 7 Functions related to the lock Perform the setting for the prohibition of change of menu items and the disabling of key operation, etc. Hierarchy of functions related to the lock Locking of functions related to the operation Various setting menus can be locked.
Functions related to the lock FZ-Ex series operetion manual Set the user information setting lock. Press the [Direction] key. Select 「513.UL.」. Input a setting value. 「513.UL. 0」 :Modifiable Inputting of the setting value 「513.UL. 1」 :Unable to Read/Write 「513.UL. 2」 :Unable to Write Set the external input/output setting lock.
FZ-Ex series operetion manual 8 Controlling and adjustment functions Controlling and adjustment functions Perform setting of the scale ID, the span adjustment and the date and time. Hierarchy of controlling and adjustment functions “ ” can not be used. -59-...
8 Controlling and adjustment functions FZ-Ex series operetion manual Outputting of the span adjustment result Can not be used. The span adjustment result can be output to a dedicated printer. Select a setting menu. Select the outputting of the span adjustment result. (“2-7-2 Operation of the setting menu, setting of various functions”...
FZ-Ex series operetion manual 8 Controlling and adjustment functions Confirm the history. Press the [Enter] key. The indication changes to the “Category of adjustment”. With each pressing of the [Enter] key, the indication changes to “Date”, “Time”, “Temperature”, “Weight used” and “Minimum indication setting”...
8 Controlling and adjustment functions FZ-Ex series operetion manual Date and time setting Select a setting menu. Select the date and time setting. (“2-7-2 Operation of the setting menu, setting of various functions” and “8-1 Hierarchy of controlling and adjustment functions”) Set the date and time.
FZ-Ex series operetion manual 8 Controlling and adjustment functions Password control This function is used for controlling by a password. Select a setting menu. Select the password control setting. (“2-7-2 Operation of the setting menu, setting of various functions” and “8-1 Hierarchy of controlling and adjustment functions”) Set the password control.
8 Controlling and adjustment functions FZ-Ex series operetion manual 8-11 Password cancellation history This function is used for checking the password cancellation history. 100 history records are stored in all. Select a setting menu. Select the password cancellation history. (“2-7-2 Operation of the setting menu, setting of various functions”...
FZ-Ex series operetion manual 8 Controlling and adjustment functions 8-12 Operation of minimum weight indication Can not be used. This is to be set when using the minimum weighed value indication function. Select a setting menu. Select the operation of minimum weighed value indication.
8 Controlling and adjustment functions FZ-Ex series operetion manual 8-14 Designation of minimum indication Select a setting menu. Select the designation of minimum indication. (“2-7-2 Operation of the setting menu, setting of various functions” and “8-1 Hierarchy of controlling and adjustment functions”) Designate a minimum indication.
Page 77
FZ-Ex series operetion manual 8 Controlling and adjustment functions 【Minimum indication list by model】 Setting value FZ623Ex FZ3202Ex FZ6202Ex FZ15001Ex FZ30K0.1GEx 61D .DA. 1 0.001 g 0.01 g 0.01 g 0.1 g 0.1 g SEL .DA. 1 61D .DA. 2 0.002 g 0.02 g 0.02 g...
8 Controlling and adjustment functions FZ-Ex series operetion manual 8-15 Reset to the factory settings Select a setting menu. Select the factory settings. (“2-7-2 Operation of the setting menu, setting of various functions” and “8-1 Hierarchy of controlling and adjustment functions”) Reset to the factory settings.
Page 79
FZ-Ex series operetion manual 8 Controlling and adjustment functions Select weight used span After an indication of 「CALWE IT」 adjustment.① appears for one second, the indication changes to the indication of weight selection used for the span adjustment. Press the [Direction] key and select a weight used for the span adjustment.
Page 80
8 Controlling and adjustment functions FZ-Ex series operetion manual The span adjustment starts. After completion of the zero-point adjustment and the indication changing to “on F.S.”, place the weight in the centre of the weighing pan. The indication changes to 「PUSH F」. Press the [Function F] key.
FZ-Ex series operetion manual 8 Controlling and adjustment functions (1) List of weights used for the span adjustment by model Model FZ623Ex FZ3202Ex FZ6202Ex FZ15001Ex FZ30K0.1GEx name 620 g 3200 g 6200 g 15000 g 30000 g 600 g 3000 g 6000 g Selectable 500 g...
Execution menu FZ- Ex series operetion manual 9 Execution menu Operation of the execution menu To operate the execution menu from the state of weighing, chiefly execute the following procedure. (1) Press the [Shift] and [Function F] keys to enter the execution menu from the state of weighing.
FZ-Ex series operetion manual Execution menu Calling of the registered user information This is a function to call the setting that was registered in “9-3 Registration of user information”. Select the execution menu mode. Select the user information calling. (“9-1 Operation of the execution menu”) Select a user to be called.
Execution menu FZ- Ex series operetion manual Registration of user information This is a function to register the content set in “5 User information setting”. Select the execution menu mode. Select the user information registration. (“9-1 Operation of the execution menu”) Select user information to be registered.
FZ-Ex series operetion manual Execution menu Calling of device setting information This is a function to call from the backup memory the setting value of a setting menu item. Select the execution menu mode. Select the calling of device setting information. (“9-1 Operation of the execution menu”) Select the device setting information.
Execution menu FZ- Ex series operetion manual Printing of the GLP header This is a function to add the GLP header at the time of printing. Set the output operation to 「 412. OO. 1 」 in the “6-8 Power supply box communication setting”.
FZ-Ex series operetion manual Execution menu Program number and check sum indication Select the execution menu mode. Select the program number and check sum indication. (“9-1 Operation of the execution menu”) Indicate program number. Press the [Direction] key. Select 「7.PROG.NO」. Press the [Enter] key.
Execution menu FZ- Ex series operetion manual Outputting of weight data Output to Power supply box: Set the output operation to 「412. OO. 1」 in “6-8 Power supply box communication setting”. 9-9-1 Outputting of tare weight Select the execution menu mode. Select the output of tare weight.
FZ-Ex series operetion manual Execution menu 9-9-3 Outputting of accumulated value. Output an accumulated value Press [Transfer] when accumulated value is indicated in the main LCD. 9-10 Indication of minimum weighed value Can not be used. The minimum weighed value set in the “8-13 Minimum weight indication value setting” can be checked.
Troubleshooting FZ- Ex series operetion manual 10 Troubleshooting 10-1 Error messages Message Cause Countermeasures ・ The weight of an object to be weighed is o-Err ・ Remove the object to be weighed, in excess of the weight of the weighing divide it into two or more, and then capacity.
Page 91
FZ- Ex series operetion manual Troubleshooting Message Cause Countermeasures ・ Sample weight is too light in comparison L-Err with the memorised reference mass of the percent scale. ・ Addition computation was executed ・ Return the indication to zero, t-Err confirm that an asterisk「 」mark doubly due to erroneous adding operation.
Page 92
Troubleshooting FZ- Ex series operetion manual Message Cause Countermeasures ・ User password input is in the wrong. ・ Check the password and input a Err702 correct password. ・ The operation key was pressed at the ・ Do not press the operation key Err703 time of starting from the standby status.
FZ- Ex series operetion manual Troubleshooting 10-2 Troubleshooting Symptom Cause Countermeasures ・ ・ Nothing indicated in DC power supply cable not Check DC power spite of turning on connected supply cable power connection. ・ ・ Indication flashes Power supply box is not switched on. Make sure that power Erroneous weight is supplied for power...
FZ-Ex series operation manual Appendix Appendix 1-2 Functional specification Protection class IP65 (Indicator and Scale only) Type of scale Weight scale / Percent scale / Coefficient scale Functions Adding functions (addition accumulating, net addition, plus side addition, minus side addition) Comparator function (2-point setting, 3-point discrimination, absolute value / relative value discrimination) Buzzer setting, Direct start,...
Appendix FZ-Ex series operation manual Temperature: Operating temperature Weighing section and Indicator section: +5 C to +40 C / humidity Power Supply Box: 0 C to +40 C Humidity: 85%RH or lower (with no condensation) Altitude Not higher than 2000 m above sea level Overvoltage category Pollution degree Indicator and scale: 3...
FZ-Ex series operation manual Appendix Appendix 3 Setting menu hierarchy list ■ Hierarchy of functions related to the operation *1 Unit that can be used on the model of Max 100-300 kg is only 「111.UA.」 : “2” (kg). *2 “ ”...
Page 100
Appendix FZ-Ex series operation manual ■ Hierarchy of functions related to the performance ■ Hierarchy of user information setting “ ” can not be used. -90-...
Page 101
FZ-Ex series operation manual Appendix ■ Hierarchy of the external input/output functions “ ” can not be used. -91-...
Page 102
Appendix FZ-Ex series operation manual ■ Hierarchy of functions related to the lock -92-...
Page 103
FZ-Ex series operation manual Appendix ■ Hierarchy of controlling and adjustment functions “ ” can not be used. -93-...
Appendix FZ-Ex series operation manual Appendix 4 Print sample ■Span ajustment result English Japanese -94-...
Page 105
FZ-Ex series operation manual Appendix ■ Header English Japanese “MA” is “Minimum weight indication value setting”, See Capter 8 Controlling and adjustment functions. When you have not set up “Minimum weight indication value”, it is printed as “none”. ■ Footer English Japanese -95-...