North America ............................... 47 8.4.6 Regulatory statements ..........................48 8.4.7 Declaration of Hazardous Substances (DoHS) ................49 8.4.8 Other regulations and standards ......................51 Health and Safety Declaration Form ...................... 52 Index ........................... 54 Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
About this chapter This chapter contains important user information and intended use of Electrophoresis Power Supply EPS 301. In this chapter Section See page Important user information About this manual Associated documentation Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Two electrophoresis units can be connected to EPS 301 and run with the same programmed method at one time. EPS 301 is intended for research use only, and shall not be used in any clinical proce- dures, or for diagnostic purposes.
Page 5
• You must understand the concepts of electrophoresis. • The system must be installed according to the instructions in Chapter 4 Installation, on page 21 of these Operating Instructions. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
This manual provides information needed to install, operate and maintain the product in a safe way. Scope of this manual This manual is valid for EPS 301. The image below shows the product. Notes and tips Note: A note is used to indicate information that is important for trouble-free and optimal use of the product.
In addition to above documentations, the product package also include product docu- mentation binders containing detailed specifications and traceability documents. Component documentation Documentation for components produced both by Cytiva and by a third-party are, if existent, also included in the product documentation. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
All users must read and understand the entire contents of this general safety chapter, and the specific safety precau- tions information in each subsequent chapter of this manual to become aware of the hazards involved. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
2 Safety instructions 2.1 Safety precautions Safety precautions Introduction EPS 301 is powered by mains voltage and handles materials that may be hazardous. Before installing, operating or maintaining the system, you must be aware of the hazards described in this manual. Definitions This user documentation contains safety notices (WARNING, CAUTION, and NOTICE) concerning the safe use of the product.
Page 10
WARNING Always follow these General Precautions to avoid injury when using EPS 301. WARNING Operation and user maintenance of EPS 301 should be performed by properly trained personnel only. WARNING Protective earth. The product must always be connected to protective earth when energized.
Page 11
Installing and moving WARNING Do not block the ventilation inlets or outlets on the system. CAUTION Do not use the product in a dusty atmosphere or close to spraying water. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Page 12
The high voltage power supply must always be disconnected when the safety lid of the electrophoresis unit is taken off. The high voltage power supply must never be switched on unless the safety lid is on the electrophoresis unit. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Page 13
2.1 Safety precautions WARNING Do not operate the instrument in extreme humidity (above 95%). Avoid condensation by letting the unit equilibrate to ambient temperature when taking EPS 301 from a colder to a warmer envi- ronment. Maintenance WARNING The instrument covers must not be opened by the user. The instru- ment contains electrical circuits which can give a lethal electric shock.
The table below describes the various safety labels that may be found on the product. Symbol/text Description Warning! Read the user documentation before using the system. Do not open any covers or replace parts unless specifi- cally stated in the user documentation. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Restart after emergency shutdown or power failure Power failure to EPS 301 will cause the run to stop immediately in an undefined state. Turn on the mains power if EPS 301 is switched off. The program continues when power is resumed. The message -PF- can be removed by pressing any key. No other special actions are needed.
3 System description System description About this chapter This chapter gives an overview of EPS 301, and a brief description of its function. In this chapter Section See page Materials Front panel Rear panel Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
3 System description 3.1 Materials Materials The main components of EPS 301 Power Supply are made of the following materials. Component Material Instrument box Aluminium Display holder Polypropene (PP). UL94V0 Keyboard Polyester Output sockets Acetalplastic (POM). UL94V0 Rubber feet Polyurethane...
SET/ENTER allows you to make changes in the program during a run. Valid values are: • Voltage 5 to 300 V • Current 10 to 400 mA • Time 0.01 to 23.59, or timer off - - - - Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Page 19
(HV on) are switched off. By pressing VIEW the end parameter values are displayed. Output sockets There are two sets of output sockets to allow two electrophoresis units to be connected and run at the same time. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
3. A switch for voltage range. The left position corresponds to 100 to 120 V and the right to 200 to 240 V. 4. Vents. MAINS Disconnect before servicing 100-120 220-240 Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
CAUTION Do not use the product in a dusty atmosphere or close to spraying water. Site requirements The installation site must comply with the following specifications: Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Page 22
The product is designed for indoor use only. Unpacking EPS 301 is delivered in protective packing material and must be unpacked with care. • Check the contents against the packing list supplied. • Record on the receiving documents if there is any apparent damage on the delivery box.
Page 23
Switch on the power. Each time the instrument is turned on a self diagnostic test is done. If an error is detected during the test a message will appear on the display and an alarm will sound. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Programming Running Quick guide Precautions WARNING Operation and user maintenance of EPS 301 should be performed by properly trained personnel only. WARNING Lethal electric shock. This instrument can develop sufficient voltage and current to produce a lethal shock. Extreme caution must be exercised in its operation.
Page 25
Do not operate the instrument in extreme humidity (above 95%). Avoid condensation by letting the unit equilibrate to ambient temperature when taking EPS 301 from a colder to a warmer envi- ronment. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Two electrophoresis units can be run at the same voltage simultaneously. When two electrophoresis units are run at the same time, double the current. Voltage is the same whether one or two units are run. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Press STOP in program mode to save the displayed value. The program then enters the END mode. Note: When the EPS 301 power supply is switched on, the programmed values from the last run can be used directly by pressing run. Set limits for voltage, current and...
• Press VIEW in RUN mode to change between the actual values for voltage, current or elapsed time. • Press VIEW for more than three seconds to switch automatically between these values. • Stop the automatic switching by pressing VIEW or RUN. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Page 29
Post electrophoresis Disconnect the leads and proceed with post-electrophoretic techniques. Since the quality of your separation will deteriorate due to diffusion, you should remove the gel and begin staining or blotting immediately. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Refer also to the separate EPS 301 quick guide included with the power supply. We recommend you keep this separate quick guide close by the instrument.
Page 31
Press ↑↓ to change the values. Press SET/ENTER. End process Note: The program stops automatically on completion. Step Action Press STOP to stop the process. Press VIEW after the run to check the end parameters. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Page 32
Step by view h.m. 0.23 view 6. Stop the run STOP automatically or manually. 7. View end parameters. view View and step by view view 1.05 h.m. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
WARNING Use only approved parts. Only spare parts and accessories that are approved or supplied by Cytiva may be used for maintaining or servicing the product. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Page 34
If wetted, unplug the power supply until the instrument is dry. Servicing All servicing should be entrusted to qualified personnel only. Please contact your local Cytiva representative for more service information. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
7 Troubleshooting Troubleshooting About this chapter This chapter contains information of troubleshooting when using EPS 301 system. Troubleshooting If an error occurs, either during a run or when switching on the power supply, the output is switched off if necessary and an error message will be displayed. The following messages can be displayed.
Reference information About this chapter This chapter contains reference information for EPS 301. In this chapter Section See page Specifications Ordering information Recycling information Regulatory information Health and Safety Declaration Form Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
8 Reference information 8.1 Specifications Specifications Technical specification The table below lists the technical specifications for EPS 301. Component Description Regulation Maximum voltage and current with automatic cross-over at preset limits Output mode Contante Voltage: 0 to 300 V DC...
Page 38
Frequency 50/60 Hz Maximum supply voltage ± 10% fluctuation Power consumption Max 120 W Pollution degree Transient level Overvoltage category II Dimensions (WxDxH) 250 x 215 x 95 mm Weight 3.0kg Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
8 Reference information 8.2 Ordering information Ordering information Product Code No Electrophoresis Power Supply EPS 301 18113001 For further information, visit cytiva.com. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Disposal of the product When taking Electrophoresis Power Supply EPS 301 out of service, the different mate- rials must be separated and recycled according to national and local environmental regulations.
European Union and European Economic Area 8.4.3 Great Britain 8.4.4 Eurasian Economic Union (Евразийский экономический союз) 8.4.5 North America 8.4.6 Regulatory statements 8.4.7 Declaration of Hazardous Substances (DoHS) 8.4.8 Other regulations and standards Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
The table below summarizes the required manufacturing information. Requirement Information Name and address of manufacturer Cytiva Sweden AB Björkgatan 30 SE 751 84 Uppsala Sweden Telephone number of manufacturer + 46 771 400 600 Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
• used according to the Operating Instructions or user manuals, and • used in the same state as it was delivered, except for alterations described in the Operating Instructions or user manuals. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
• used according to the Operating Instructions or user manuals, and • used in the same state as it was delivered, except for alterations described in the Operating Instructions or user manuals. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Telephone: +7 985 192 75 37 E-mail: rucis@cytiva.com Информация о производителе и импортере В следующей таблице приводится сводная информация о производителе и импортере, согласно требованиям Технических регламентов Таможенного союза и (или) Евразийского экономического союза. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Page 46
Адрес электронной почты: rucis@cytiva.com Shelf life and manufacturing date For information regarding shelf life please contact your local Cytiva representative. For information regarding manufacturing date, see year and month of production on the product. Description of symbol on the system label Описание...
Note: The user is cautioned that any changes or modifications not expressly approved by Cytiva could void the user’s authority to operate the equip- ment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules.
When used in a residential environment, there is a concern of radio interference. 유의사항 A급 기기 (업무용 방송통신 기자재) 이 기기는 업무용환경에서 사용할 목적으로 적합성평가를 받 은 기기 로서 가정용 환경에서 사용하는 경우 전파간섭의 우려가 있습 니다. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
EFUP shall be done in accordance with the Product Maintenance Procedures. This product must not be disposed of as unsorted municipal waste, and must be collected separately and handled properly after decommissioning. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
Page 50
Indicates that this hazardous substance contained in at least one of the homogeneous materials used for this part is above the limit requirement in GB/T 26572 • Data listed in the table represents best information available at the time of publication. Electrophoresis Power Supply EPS 301 Operating Instructions 29115049 AC...
This section describes the additional standards that apply to the product. Regulatory compliance of connected equipment Any electrical equipment connected to Electrophoresis Power Supply EPS 301 should meet the safety requirements of EN/IEC 61010-1, or relevant standards. Within EU, connected equipment must be CE marked.
Service Ticket #: To make the mutual protection and safety of Cytiva service personnel and our customers, all equipment and work areas must be clean and free of any hazardous contaminants before a Service Engineer starts a repair. To avoid delays in the servicing of your equipment, complete this checklist and present it to the Service Engineer upon arrival.
Page 53
To make sure the mutual protection and safety of Cytiva personnel, our customers, transportation personnel and our environment, all equipment must be clean and free of any hazardous contaminants before shipping to Cytiva. To avoid delays in the processing of your equipment, complete this checklist and include it with your return.
Need help?
Do you have a question about the EPS 301 and is the answer not in the manual?
Questions and answers