Summary of Contents for Helmer Scientific HPR118-GX
Page 1
CVS Slimline Refrigerator Instructions for Use Horizon Series™ HPR118-GX 360453/B...
Page 2
Helmer Scientific. Helmer Scientific assumes no responsibility or liability for any errors or inaccuracies that may appear in the informational content contained in this material. For the purpose of clarity, Helmer Scientific considers only the most recent revision of this document to be valid.
Models are indicated by a distinguishing model number that corresponds to the series, type, number of doors, and capacity of the refrigerator. For example, “HPR118-GX” refers to a Horizon Series Pharmacy Refrigerator with one door and a capacity of 18 cu ft.
Helmer Scientific Horizon Series Refrigerator Instructions for Use ™ 1 .5 Avoiding Injury • Do not use mechanical devices or other means to accelerate the defrosting process, other than those recommended by the manufacturer. • Do not damage the refrigerant circuit.
Helmer Scientific Horizon Series Refrigerator Instructions for Use ™ Installation 2 .1 Location Keep all ventilation openings in the enclosure or, in the structure of building-in, clear of obstruction. ♦ Has a grounded outlet meeting electrical requirements listed on the product specification label.
Page 7
Helmer Scientific Horizon Series Refrigerator Instructions for Use ™ Primary Monitor Probe The primary monitor probe is located at the top left side of the refrigerator. Primary monitor probe Install Additional Probe Through Access Port 1. Peel back putty to expose the port. (If routing through the side port, open the port doors to remove the putty) 2.
Helmer Scientific Horizon Series Refrigerator Instructions for Use ™ Operation 3 .1 Initial Power-Up 1. Plug the power cord into a grounded outlet that meets the electrical requirements on the product specification label. 2. Switch the AC ON/OFF switch ON.
Helmer Scientific Horizon Series Refrigerator Instructions for Use ™ View recorded temperature timer Notes • The timer denotes the period of time that has elapsed. It does not display the time at which a minimum or maximum temperature occurred. • The maximum period of time the timer can record is 99:59 (99 hours and 59 minutes).
Helmer Scientific Horizon Series Refrigerator Instructions for Use ™ 3 .4 Set Parameter Values Press and hold the Up and Down Arrows simultaneously for three (3) seconds to enter program mode. The LED Display will show .C or .F to indicate Celsius or Fahrenheit.
Helmer Scientific Horizon Series Refrigerator Instructions for Use ™ 3.7 Temperature Calibration Offsets Temperature calibration offsets indicate an acceptable margin of error between the actual temperature value and the desired temperature value. Monitor Offset ♦ Value is factory-set to match a calibrated reference thermometer.
Helmer Scientific Horizon Series Refrigerator Instructions for Use ™ 3 .10 Light Operation Note • The light switch must be in the ON position to illuminate the lights when the door is opened. • The default time delay for the lights to remain on is five (5) minutes.
Helmer Scientific Horizon Series Refrigerator Instructions for Use ™ Product Specifications 4 .1 Operating Standards These units are designed to operate under the following environmental conditions: ♦ Indoor use only ♦ Altitude (maximum): 2000 m ♦ Ambient temperature range: 15 °C to 32 °C (59°F to 90°F) ♦...
Helmer Scientific Horizon Series Refrigerator Instructions for Use ™ Compliance 5 .1 Safety Compliance This product is certified to applicable UL and CSA standards by a NRTL. 5 .2 Environmental Compliance This device complies with the 2011/65/EU Directive for the Restriction of Hazardous Substances (RoHS).
Page 15
Helmer Scientific Horizon Series Refrigerator Instructions for Use ™ 360453/B...