Error Messages And Corrective Actions - Omron ZG2-DSU User Manual

Smart sensor data storage unit
Hide thumbs Also See for ZG2-DSU:
Table of Contents

Advertisement

Error Messages and Corrective Actions

The following shows error messages that are displayed on the LCD screen and their
corrective actions.
Error message
Display details
SYSTEM ERROR
BANK DATA
SYSTEM ERROR
CONNECT
SYSTEM ERROR
MAIN COM
File not found
Setting failed
Data error
Transfer failed
Dead Battery
Disp range Error
NO MEDIA
Insufficient
ZG2-DSU User's Manual
Cause
Internal bank data is in error.
The Sensor Controller is not
connected.
Internal error
The specified bank file does not
exist on the Memory Card.
The specified setting value is out
of range.
The specified bank file is in error. Select the correct bank file.
Bank file communication error
The clock backup battery has
run down.
The measurement results
exceed the displayed number of
digits.
The Memory Card is not
inserted.
There is not enough remaining
space on the Memory Card.
Error Messages and Corrective Actions
Remedy
Hold down the
key for 3
seconds, and then hold down the
key for 3 seconds.
After initialization ends, turn the
power OFF then back ON again
to restore the unit.
Connect the Sensor Controller.
Turn the unit OFF then back ON
again.
Change the bank file number to
an existing file name.
Set within the allowable setting
range.
Check the connection between
Sensor Controllers.
Please contact your OMRON
sales representative.
Change the number of digits
past the decimal point setting.
Insert the Memory Card.
•Replace the Memory Card.
•Delete unwanted files from the
Memory Card.
•Format the Memory Card.
77

Advertisement

Table of Contents
loading

Table of Contents