Periodic Maintenance
Kingon P2 Portable Oxygen
Concentrator

Advertisement

loading
Need help?

Need help?

Do you have a question about the P2 and is the answer not in the manual?

Questions and answers

Adrian Klein
March 26, 2025

System does nor charge.

1 comments:
Mr. Anderson
March 26, 2025

The Kingon P2 system may not charge due to a Power Supply Fail, where the input voltage of the power supply is insufficient to ensure normal operation. Additionally, if the battery is depleted and not connected to an adapter, it will need to be replaced or recharged. If the battery temperature is too cold or too hot, the system may also shut down, preventing charging. Moving the concentrator to a suitable environment or checking the power supply can help resolve the issue.

This answer is automatically generated

Subscribe to Our Youtube Channel

Summary of Contents for Kingon P2

  • Page 1 Periodic Maintenance Kingon P2 Portable Oxygen Concentrator...
  • Page 2 1. Filter Cotton (Suggest to replace once a month) Remove the pre-filter cover by hand and replace with a new filter cotton. Replace the the pre-filter cover, and it's done successfully. Note: depending on the device use, you may not need to replace if it’s not dirty. 2.
  • Page 3 3. Battery The removed battery and any spare batteries you are not regularly using should be maintained every 3 months. Charge the battery to 100%, then drain to ~60-80% and then store away. Note: if you don’t plan to use the POC for a long time, please remove the battery. 4.