UNI
MP110 Series
Quick Start Guide
mPower Electronics Inc.
2910 Scott Blvd. Santa Clara, CA 95054
www.mpowerinc.com
info@mpowerinc.com
PN:M027-4002-000
Read Before Operating
The User's Guide must be carefully read by all
individuals who have or will have the responsibility of
using, maintaining, or servicing this product. The
product will perform as designed only if it is used,
maintained, and serviced in accordance with the
manufacturer's instructions.
User Interface
The UNI's user interface consists of the LCD
display, LEDs, an alarm buzzer, one push-button key,
an alligator clip, and a chemical sensor.
LED alarm
window
LCD
Display
Buzzer
Sensor and
gas inlet
LITE
v1.1
Alligator clip
Concentration
Display
Battery
status
Programming
& On/Off key
Sensor ID
Product label
with Sensor ID
Warning
Never operate the monitor when the cover is removed.
Remove the monitor cover and battery only in area known
as non-hazardous.
Use only mPower's lithium battery part number M500-
0038-000 (EVE 14335) (3.6 V, 1650 mAh, 2/3 AA size).
This instrument has not been tested in an explosive gas/air
atmosphere having oxygen concentration greater than 21%.
Substitution of components will impair suitability for
intrinsic safety and void warranty.
It is recommended to bump test with a known
concentration gas to confirm the instrument is functioning
properly before use.
Before use, ensure that the colorless ESD layer on the
display is not damaged or peeling. (The blue protective film
may be removed.)
Avertissement
N'utilisez jamais le moniteur lorsque le couvercle est enlevé.
Retirer le couvercle du moniteur et la batterie uniquement
dans une zone connue comme non dangereuse.
Utilisez uniquement le numéro de pièce de la batterie au
lithium mPower M500-0038-000 (EVE 14335) (3.6 V, 1650
mAh, 2/3 AA grandeur).
Cet instrument n'a pas été testé dans une atmosphère
explosive gaz / air ayant une concentration en oxygène
supérieure à 21%.
La substitution de composants compromettra l'aptitude à la
sécurité intrinsèque et annulera la garantie.
Il est recommandé de tester avec un gaz de concentration c
onnu pour confirmer que l'instrument fonctionne correctem
ent avant de l'utiliser.
Avant l'utilisation, assurez-vous que la couche ESD incolore
de l'écran n'est pas endommagée ou épluchée. (Le
film protecteur bleu peut être enlevé.)
Turning the Unit On
Press and hold the Operation Key (
seconds, until the LCD displays "On" and the unit
enters a self-test sequence. The alarm buzzer,
vibrator, and LEDs are tested, and the firmware
version, (e.g. "VER","1.0.0.0") and the programmed
alarm limits, STEL, TWA, etc. are shown.
After the start-up is completed, the instrument
enters the detection mode and displays the real-time
gas concentration.
Turning the Unit Off
In normal display mode, press and hold the Key
for a 3-second count down until the unit powers off.
In normal mode, the gas concentration is
displayed on the LCD continuously and the unit
alarms if a pre-set limit is exceeded. Press and hold
the Key until it beeps to test the function of the
buzzer, LEDs, and vibration alarm. Or short press
the Key to cycle through values of STEL, TWA,
MIN (O
only), and PEAK, and view EVT LOG.
2
From the EVT LOG screen, long-press the
button until it beeps to show the latest alarm event
A1 and then short-press the button to cycle through
the last 10 alarm events. Up to 50 events can be
3
viewed using mPower Suite.
Normal Mode
Event Log
2
) for 3
4

Advertisement

loading
Need help?

Need help?

Do you have a question about the UNI LITE MP110 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for mPower Electronics UNI LITE MP110 Series

  • Page 1 2/3 AA grandeur). • Cet instrument n'a pas été testé dans une atmosphère explosive gaz / air ayant une concentration en oxygène mPower Electronics Inc. supérieure à 21%. • La substitution de composants compromettra l'aptitude à la 2910 Scott Blvd. Santa Clara, CA 95054 sécurité...
  • Page 2 Enter Configuration Mode Exit Configuration Mode Scroll to “EXIT?” and long-press to exit and With the unit off, double-click the Operation Key and the unit shows “PROG”. Short-press the Key to return to Normal Mode. prompt password entry with the first digit flashing. Alarm Limits Short-press the Key to increase the number, and long- Alarms are triggered when readings are above...
  • Page 3 Maintenance and Service Proper Product Disposal Battery (continued): Until the battery is drained, at The End Of Life the screen will display "bAT LoW" and the instrument The Waste Electrical and readings will no longer be displayed. After the battery Electronic Equipment dies, it will continue to sound and light alarm for 1 (WEEE) directive...