AEU-6000-70V & AEU-6000 Quick Reference Guide
Manual Operation:
• Select handpiece ratio:
- 1:5, 1:3, 1:2, 1:1, or 20:1 Ratios - Unit
- 8:1 Ratio - Unit automatically switches to
• Select speed, torque setting and irrigation
setting by using the up/down buttons.
• Select Motor Direction
• Select Torque: Auto Stop (Green Light)
• Select Flow On/Off
Preset Operation:
• Select Preset by pressing one of the
buttons numbered 1-5
Changing Presets:
• Select desired handpiece ratio, speed, torque, and irrigation
settings as indicated above in Manual Operation.
• Press and hold Preset button where you want new settings to be
stored.
• Follow Prompt:
- "Save To Preset (#)?"
Press Preset button #1 (YES)
• Observe Confirmation:
- "Preset (#) Setting Saved"
NOTE: Software Version: Appears briefly when unit is turned on.
AEU-6000-70V & AEU-6000 Quick Reference Guide
Manual Operation:
• Select handpiece ratio:
- 1:5, 1:3, 1:2, 1:1, or 20:1 Ratios - Unit
- 8:1 Ratio - Unit automatically switches to
• Select speed, torque setting and irrigation
setting by using the up/down buttons.
• Select Motor Direction
• Select Torque: Auto Stop (Green Light)
• Select Flow On/Off
Preset Operation:
• Select Preset by pressing one of the
buttons numbered 1-5
Changing Presets:
• Select desired handpiece ratio, speed, torque, and irrigation
settings as indicated above in Manual Operation.
• Press and hold Preset button where you want new settings to be
stored.
• Follow Prompt:
- "Save To Preset (#)?"
Press Preset button #1 (YES)
• Observe Confirmation:
- "Preset (#) Setting Saved"
NOTE: Software Version: Appears briefly when unit is turned on.
automatically switches to Implant Mode
Endo Mode.
Maximum (Orange Light)
Limiting (No Lights)
automatically switches to Implant Mode
Endo Mode.
Maximum (Orange Light)
Limiting (No Lights)
Calibrating Handpiece
• Preselect ratio of handpiece being calibrated by using Ratio
Up/Down buttons:
(Note: Unit switches to Implant Mode when 1:5, 1:3, 1:2, 1:1, or 20:1
ratios are selected or Endo Mode when 8:1 ratio is selected.)
• Insert a file, bur, or drill into the handpiece.
• Press CAL button.
• Follow Prompts:
- "Add Handpiece to Motor"
Press Preset button #1 (NEXT).
- "Free Run in Progress, Please Wait..."
- "Calibration Successful!, The Result is Saved"
- "Calibration Failed!"
Press Preset button #1 (RETRY), or button #3 (EXIT)
Installing new software (takes 30 seconds):
• Turn off unit
• Pull back blue plug from slot and insert new
software memory card.
• Turn unit On
• "Memory Card Detected. Re-program?"
Press Preset Button #1 (YES) to install new program.
• "Presets will be erased. Continue?" Press button #1 (YES).
• "Programming successful. Eject card."
• System reboots.
Calibrating Handpiece
• Preselect ratio of handpiece being calibrated by using Ratio
Up/Down buttons:
(Note: Unit switches to Implant Mode when 1:5, 1:3, 1:2, 1:1, or 20:1
ratios are selected or Endo Mode when 8:1 ratio is selected.)
• Insert a file, bur, or drill into the handpiece.
• Press CAL button.
• Follow Prompts:
- "Add Handpiece to Motor"
Press Preset button #1 (NEXT).
- "Free Run in Progress, Please Wait..."
- "Calibration Successful!, The Result is Saved"
- "Calibration Failed!"
Press Preset button #1 (RETRY), or button #3 (EXIT)
Installing new software (takes 30 seconds):
• Turn off unit
• Pull back blue plug from slot and insert new
software memory card.
• Turn unit On
• "Memory Card Detected. Re-program?"
Press Preset Button #1 (YES) to install new program.
• "Presets will be erased. Continue?" Press button #1 (YES).
• "Programming successful. Eject card."
• System reboots.
...OR:
...OR: