Model 8447D/E/F
Page 7
Table 2, Recommended Test Equipment and Accessories (Cont*d)
Instrument Type
Minimum Specifications
Suggested
Model
Use*
EMS Voltmeter
Voltage Range: 1 mV to 1 V (full range)
Frequency Range: 10 Hz to 10 MHz
Meter Accuracy: ±5%
HP 3400A
P,A,
T
Power Meter
Power Range: —20 dBm to -+T0 dBm
Frequency Range: 10—1300 MHz
Meter Accuracy: ±1%
HP 432A/478A
P
Amplifier
Gain: 24 dB
Noise: < 8 dB
Frequency Range: 0.1 — 1300 MHz
HP 8447D
P
Digital Voltmeter
Voltage Range: 0 — 60 V
Accuracy: ±.05% ±1 digit
Resistance Range: 0 to 10 megohm
Accuracy: ±0.3% ± 1 digit
HP 3440A/3444A
A,T
Double Balanced
Mixers
Frequency Range: 200 —1300 MHz
Sage Laboratories
2500 Series
P
Low Pass Filter
Cutoff Frequency: 400 MHz
Attenuation: >60 dB at 800 and 1200 MHz
- - - -.-.. .... ...
.- -
-
Cir-Q>Tel
FLT/2 Series
P
Attenuator
10 dB, 0.1 -1300 MHz
HP 8491A
Option 10
P
Attenuator
20 dB, 0.1-1300 MHz
HP 8491A
Option 20
P
Termination
50 ohm (BNC)
HP 11593A
P
Termination
50 ohm (Type N)
HP 908A
P
Adapter
Type N Female to BNC Male
UG-349A/U
P
Adapter (2)
BNC Female to Type N Male
UG-201A/U
P
1
Tee
BNC
UG-274B/U
P
Cable Assembly
Type N Connectors
HP 11500A
P
Cable Assembly (5)
BNC Connectors
HP 10503A
P
Cable Assembly
BNC with test clips attached
HP 10501A
Adapter
BNC Female to Binding Post
HP 10111A
P,A,T
Adapter
BNC Male to BNC Male
UG-491A/U
p
*Use:
P = Performance Test; A * Adjustments; T ■ Troubleshooting
Need help?
Do you have a question about the 8447D and is the answer not in the manual?