Dc Current Calibration - East Tester ET3240 User Manual

4 1/2 / 5 1/2 benchtop digital multimeter
Table of Contents

Advertisement

5 1/2 Digital Multimeter User Manual
8.3 DC current calibration
calibration.
Two ways to enter the DC current calibration:
ca.
L_In the DC current measurement function, press the
Ne
wol key combination, and
The calibration values that need to be entered for each range are shown below:
1 200pA range need to enter 0 pA and 200 pA total of two values.
._2mA range need to enter 0 mA and 2 mA total of two values.
20mA range need to enter 0 mA and 20 mA total of two values.
- 200mA range need to enter 0 mA and 200 mA total of two valucs.
... 2A file need to enter 0 A and 2 A total of two values.
Aa
Warning:
Two ways to enter the AC current calibration:
ct
LML
key combination, and
In the AC current measurement function, press the
correctly enter the calibration password to enter the AC current calibration function. In this way, vou
can start calibration from any range or calibrate a particular range.
aw
Warning: