5 1/2 Drgatal Mulumeter User Manual
Note:
The minimum frequency that can be measured is 1 Hz
7.8 Diode
Range
Input protection
|
Description
0-2V
250Vp
| input Current is about 0.75mA..
Table 7-8
_Diode measurement indicator
Note:
Open circuit voltage is about 3 2V-
7.9 Conductance
Range | Resolution | _ Input protection
Description
Whien the measured resistance is less than
300, the buzzer sounds; when the reading is
0-20
00010
250Vp
less than 2KQ, "-OL-" is displayed when it
exceeds. The threshold resistance can be set
in the range of 0 to 2kO.
Table 7-9
_Conductance measurement indicator
Note:
Open circuit voltage is about 3.2V.
7.10 Duty Cycle
7.1 Temperature
8 Calibration
Note:
1) Betore calibration, it is required to prcheat the instrument for more than 30 minutes. The
accuracy of the standard source must be better than 3 — 5 times the accuracy of the calibrated
instrument.
2) After a specific range was calibrated, the calibration data i
you can just calibrate a separate range
3) In order to prevent inaccurate calibration due to wrong operation, it is required to enter a
password before the calibration.
4) You can enter the calibration interface after input the correct password. Take the interface of
DCV calibration as an example
mediately written to FLASH. So