    CALIBRATION PROCEDURE NI 9252 8 AI, ±10 V, 24 bit, 50 kS/s/ch Simultaneous This document contains the verification and adjustment procedures for NI 9252. For more information about calibration solutions, visit ni.com/calibration. Contents Software............................ 1 Documentation.......................... 1 Test Equipment..........................2 Test Conditions..........................3 Initial Setup..........................
  • Page 3: Test Equipment

    NI cDAQ-9174/9178 USB Chassis Quick Start NI-DAQmx installation and hardware setup NI 9252 Getting Started Guide NI 9252 connection information NI 9252 Datasheet NI 9252 circuitry and specifications NI-DAQmx Readme Operating system and application software support in NI-DAQmx LabVIEW Help LabVIEW programming concepts and reference information about NI-DAQmx VIs and functions NI-DAQmx C Reference Help...
  • Page 4: Test Conditions

    NI 9252. The verification procedure assumes that adequate traceable uncertainties are available for the calibration references. Accuracy Verification Complete the following procedure to determine the As-Found status of the NI 9252. NI 9252 Calibration Procedure | © National Instruments | 3...
  • Page 5 Set the calibrator to Standby mode (STBY). Connect the NI 9252 to the calibrator as shown in the following figure. Figure 1. Analog Input Verification Connections AI x + AI x + AI x – AI x – Guard Calibrator NI 9252 Calibrator NI 9252...
  • Page 6: Adjustment

    Set the calibrator to a reference value determined by the array of recommended calibration voltages. Set the calibrator to Operate mode (OPR). Call and configure the NI 9252 adjustment function according to NI 9252 Adjustment Configuration table. NI 9252 Calibration Procedure | © National Instruments | 5...
  • Page 7: Eeprom Update

    Table 5. NI 9252 Adjustment Configuration Physical Channel Reference Value cDAQ1Mod8/AI0:7 The actual value of the voltage output of the calibrator The module acquisition is started each time the adjust function is Note called. There is a delay of approximately 550 µs from the acquisition start until module data becomes available.
  • Page 8: Worldwide Support And Services

    1 866 ASK MYNI (275 6964). For support outside the United States, visit the Worldwide Offices section of ni.com/niglobal access the branch office websites, which provide up-to-date contact information. Range equals 10.5 V. NI 9252 Calibration Procedure | © National Instruments | 7...
  • Page 9 For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your patents.txt ni.com/patents software, the file on your media, or the National Instruments Patent Notice at . You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer ni.com/legal/export-compliance...