  • Page 1 ® R&S NGE100B Power Supply User Manual (Æ1=[2) 5601134302 Version 10...
  • Page 2 ® This manual describes the following R&S NGE100B models with firmware version 1.55 and higher: ● ® R&S NGE102B 2-Channel PSU (5601.3800.02) ● ® R&S NGE103B 3-Channel PSU (5601.3800.03) In addition to the base unit, the following options are described: ●...
  • Page 3: Table Of Contents

    ® Contents R&S NGE100B Contents 1 Safety and regulatory information............9 Safety instructions......................9 Labels on R&S NGE100B................... 12 Warning messages in the documentation..............13 Korea certification class A..................14 2 Documentation overview..............15 Manuals........................15 Data sheet........................15 Calibration certificate....................15 Release notes, open source acknowledgment (OSA)..........16 3 Conventions used in the documentation...........17 Typographical conventions..................17 Notes on screenshots....................17...
  • Page 4 ® Contents R&S NGE100B 5.2.2 Rear panel........................29 Trying out the instrument...................32 5.3.1 Selecting the channels....................32 5.3.2 Setting the output voltage and current limit..............32 5.3.3 Activating the channel output..................33 5.3.4 Storing/Recalling of instrument settings................33 Instrument control...................... 33 5.4.1 Ways to operate the instrument..................34 5.4.2 Means of manual interaction..................
  • Page 5 ® Contents R&S NGE100B Advanced functions....................64 7.7.1 EasyArb.........................65 7.7.2 EasyRamp........................67 Store and recall......................69 Interfaces and protocols.................... 71 7.9.1 Network connection.......................71 7.9.1.1 LAN connection......................71 7.9.1.2 Wireless LAN connection....................74 7.9.2 USB connection......................77 7.9.3 LXI..........................79 7.9.4 SCPI..........................81 7.9.5 VISA..........................81 7.9.6 Setting up remote control connection................81 7.10...
  • Page 6 ® Contents R&S NGE100B Measurement commands..................117 Advanced operating commands................118 8.5.1 Arbitrary commands....................118 8.5.2 EasyRamp commands....................122 8.5.3 DIO commands......................122 Data and file management commands..............127 Firmware update....................... 128 9 Troubleshooting................. 129 Displaying status information..................129 Problems during firmware update................129 Problems with remote control over LAN..............
  • Page 7 ® Contents R&S NGE100B A.1.5 Responses to queries....................143 Command sequence and synchronization............. 144 A.2.1 Preventing overlapping execution................144 List of commands................147 Index....................149 User Manual 5601.1343.02 ─ 10...
  • Page 8 ® Contents R&S NGE100B User Manual 5601.1343.02 ─ 10...
  • Page 9: Safety And Regulatory Information

    ® Safety and regulatory information R&S NGE100B Safety instructions 1 Safety and regulatory information The product documentation helps you use the product safely and efficiently. Follow the instructions provided here and in the following chapters. Intended use The product is intended for the development, production and verification of electronic components and devices in industrial, administrative, and laboratory environments by personnel familiar with the potential risks of measuring electrical quantities.
  • Page 10 ® Safety and regulatory information R&S NGE100B Safety instructions Only people skilled in electrical work should connect, set up and use the product. Such persons have the education and experience needed to recognize risks and avoid haz- ards of working with electricity. These users also need sound knowledge of at least one of the languages in which the user interfaces and the product documentation are avail- able.
  • Page 11 ® Safety and regulatory information R&S NGE100B Safety instructions Connecting to power The product is an overvoltage category II product. Connect the product to a fixed installation used to supply energy-consuming equipment such as household applian- ces and similar loads. Keep in mind that electrically powered products have risks, such as electric shock, fire, personal injury or even death.
  • Page 12: Labels On R&S Nge100B

    ® Safety and regulatory information R&S NGE100B Labels on R&S NGE100B ● When operating measuring accessories, only use the cables delivered with the accessory. If you have to use cables from other manufacturers, make sure that they are of the required overvoltage category. Do not operate the product in series or parallel unless that setup is supported.
  • Page 13: Warning Messages In The Documentation

    ® Safety and regulatory information R&S NGE100B Warning messages in the documentation ● Product and environment safety, see Table 1-1. ● Device information is provided on a sticker attached to the rear panel of R&S NGE100B. The sticker contains a barcode and the device ID. The device ID is a combination of the order number and the serial number.
  • Page 14: Korea Certification Class A

    ® Safety and regulatory information R&S NGE100B Korea certification class A 1.4 Korea certification class A 이 기기는 업무용(A급) 전자파 적합기기로서 판매자 또는 사용자는 이 점을 주의하시기 바라며, 가정외의 지역에서 사용하는 것을 목적으로 합니다. User Manual 5601.1343.02 ─ 10...
  • Page 15: Documentation Overview

    ® Documentation overview R&S NGE100B Calibration certificate 2 Documentation overview This section provides an overview of the R&S NGE100B user documentation. 2.1 Manuals You find the documents on the R&S NGE100B product page at: www.rohde-schwarz.com/product/nge100b Getting started Introduces the R&S NGE100B power supply series and describes how to set up and start working with the instrument.
  • Page 16: Release Notes, Open Source Acknowledgment (Osa)

    ® Documentation overview R&S NGE100B Release notes, open source acknowledgment (OSA) 2.4 Release notes, open source acknowledgment (OSA) The release notes list new features, improvements and known issues of the current firmware version, and describe the firmware installation. The open source acknowledg- ment document provides verbatim license texts of the used open source software.
  • Page 17: Conventions Used In The Documentation

    ® Conventions used in the documentation R&S NGE100B Notes on screenshots 3 Conventions used in the documentation 3.1 Typographical conventions The following text markers are used throughout this documentation: Convention Description "Graphical user interface ele- All names of graphical user interface elements on the screen, such as ments"...
  • Page 18: Welcome To R&S Nge100B

    ® Welcome to R&S NGE100B R&S NGE100B 4 Welcome to R&S NGE100B The two or three-channel power supply series are based on a classical transformer concept with high efficiency electronic pre-regulators and secondary linear regulators. This concept allows the instrument to achieve the high output power within a minimum space, high efficiency and lowest residual ripple.
  • Page 19: Getting Started

    ® Getting started R&S NGE100B Preparing for use 5 Getting started 5.1 Preparing for use Here, you can find basic information about setting up the product for the first time. 5.1.1 Lifting and carrying "Lifting and carrying the product" on page 10. 5.1.2 Unpacking and checking 1.
  • Page 20: Setting Up The R&S Nge100B

    ® Getting started R&S NGE100B Preparing for use Electromagnetic compatibility classes The electromagnetic compatibility (EMC) class indicates where you can operate the product. The EMC class of the product is given in the data sheet. ● Class B equipment is suitable for use in: –...
  • Page 21: Mounting The R&S Nge100B In A Rack

    ® Getting started R&S NGE100B Preparing for use Left = Stacked correctly Middle left = Stacked incorrectly, too many products Middle right = Stacked incorrectly, different dimensions Right = Stacked incorrectly, folded-out feet 4. NOTICE! Overheating can damage the product. Prevent overheating as follows: ●...
  • Page 22: Considerations For Test Setup

    ® Getting started R&S NGE100B Preparing for use 3. Push the R&S NGE100B onto the shelf until the rack brackets fit closely to the rack. 4. Tighten all screws at the rack brackets with a tightening torque of 1.2 Nm to secure the R&S NGE100B at the rack.
  • Page 23 ® Getting started R&S NGE100B Preparing for use Weight NGE102B 4.9 kg (10.80 lb) NGE103B 5.0 kg (11.02 lb) Operating limits The R&S NGE100B is equipped with a protective overload feature. The protective overload feature prevents damage to the instrument and is intended to protect against a possible electrical shock.
  • Page 24: Connecting To Power

    ® Getting started R&S NGE100B Preparing for use 5.1.6 Connecting to power For safety information, see "Connecting to power" on page 11. Before connecting the instrument to the mains, check whether the mains voltage con- forms to the mains voltage range specified on the label located at the left of the power connector.
  • Page 25: Switching On Or Off

    ® Getting started R&S NGE100B Preparing for use Ground terminal If necessary, ground the instrument using the grounding connection located at the rear panel: 1. Unscrew the screw of the ground terminal using a cross-recess screw driver. 2. Attach a ground cable with a ring terminal and pass the screw through it. 3.
  • Page 26: Connecting To Lan

    ® Getting started R&S NGE100B Preparing for use 5.1.8 Connecting to LAN Establishing the LAN connection Instrument option R&S NGE-K101 (P/N: 5601.2204.03) option is required to connect the R&S NGE100B to a network via LAN connection. The R&S NGE100B provides Ethernet (LAN) connectivity. Provided the corresponding rights are assigned, you can use these interfaces for remote control and data transfer from a controller PC.
  • Page 27: Connecting Usb Devices

    ® Getting started R&S NGE100B Instrument tour Figure 5-2: Ethernet settings dialog 5.1.9 Connecting USB devices The USB Type-A connector is at the front panel. You can connect or disconnect all USB devices from the R&S NGE100B during operation. But do not remove an external USB memory stick while the instrument is performing firmware update and storing of screen captures, since it leads to unsuccessful updates and loss of data.
  • Page 28: Front Panel

    ® Getting started R&S NGE100B Instrument tour 5.2.1 Front panel The front panel of the R&S NGE100B is as shown in Figure 5-3. The function keys and navigation controls are located at the right side of the display. The various connectors are located below the display and function keys.
  • Page 29: Rear Panel

    ® Getting started R&S NGE100B Instrument tour For detailed description on function keys, see Chapter 6.1.1, "Function keys", on page 43. Rotary knob and arrow keys (3) Rotary knob and arrow keys are means of navigation and adjustment. When pressed or rotated, they perform tasks like navigation around the screen, adjustment of param- eter values or confirmation of entries.
  • Page 30 ® Getting started R&S NGE100B Instrument tour 7 = Ground terminal 8 = AC inlet with fuse holder 9 = Voltage selector 10 = Ethernet (LAN) connector 11 = USB connector 12 = Digital I/O connector 13 = Kensington lock Ground terminal (7) Protective ground terminal to secure the R&S NGE100B, e.g.
  • Page 31 ® Getting started R&S NGE100B Instrument tour Voltage selector (9) Setting voltage selector On your first power-on connection, a yellow label sticker is attached over the inlet. Before peeling off the yellow label sticker, make sure that the correct fuse rating is used for the mains voltage.
  • Page 32: Trying Out The Instrument

    ® Getting started R&S NGE100B Trying out the instrument Digital I/O connector (12) Risk of instrument damage Do not exceed the maximum voltage rating of the Digital I/O pins (5.25 V max) when supplying voltages to the pins. The Digital I/O is a 5-pin terminal block for external trigger input or output. Measurement control for Digital I/O can be achieved with an external input signal or as an output signal to trigger other instruments for measurements.
  • Page 33: Activating The Channel Output

    ® Getting started R&S NGE100B Instrument control Alternatively: 1. Press [Voltage] or [Current] key on the front panel. 2. Press the desired channel key to activate the respective voltage or current limit set- ting of that channel. The value on the respective channel becomes editable and is positioned by a blue cursor.
  • Page 34: Ways To Operate The Instrument

    ® Getting started R&S NGE100B Instrument control 5.4.1 Ways to operate the instrument You can operate an R&S NGE100B in two ways: ● Manual operation Use the front panel controls to configure the R&S NGE100B settings. Chapter 5.4.2, "Means of manual interaction", on page 34 for basic informa- tion on manual operation of the instrument.
  • Page 35 ® Getting started R&S NGE100B Instrument control Figure 5-6: Example of screen display Device status bar Channel status bar Channel display area Historical channel information The following sections explain the information areas as labeled. Status bar information There are two types of status bar information located at the top of the screen layout: ●...
  • Page 36 ® Getting started R&S NGE100B Instrument control Function Description DIO:xxxx Digital Trigger I/O (Digital Trigger I/O option R&S NGE-K103 must be installed). The "xxxx" refers to I/O status for DIO1, DIO2, DIO3, DIO4. If in use, the icon is highlighted in green. If enabled, the icon is highlighted in white.
  • Page 37 ® Getting started R&S NGE100B Instrument control The channel display area shows the voltage and current values and the corresponding operating mode (CC, CV) of the R&S NGE100B when the device output is turned on. "Operating mode" on page 37 for the different operating modes that the R&S NGE100B supported.
  • Page 38 ® Getting started R&S NGE100B Instrument control Figure 5-10: Color coding of difference operating conditions Color Operating mode Description OFF mode Output is OFF Editing mode A solid blue cursor is shown when an item is selected. All cursor keys are also lit up. CV mode Active outputs are operated in a constant voltage mode.
  • Page 39: Accessing The Functionality

    ® Getting started R&S NGE100B Instrument control Figure 5-11: Example of a menu selection The term dialog refers to the views that cover the parameters of a certain function. Figure 5-12: Example of a dialog ● On-Screen keypad The on-screen keypad appears if alphanumeric entry field is required. See "Enter- ing alphanumeric parameters"...
  • Page 40: Entering Data

    ® Getting started R&S NGE100B Instrument control Apart from the main menu, we use the term "dialog" to refer the editable windows in the instrument. To open the menu 1. Press [Menu] key at the front panel. The main menu of the instrument is displayed. 2.
  • Page 41 ® Getting started R&S NGE100B Instrument control Entering numeric parameters If the input field requires numeric input, you can use either the rotary knob or arrow keys on the front panel to adjust to required values. To correct value in a dialog 1.
  • Page 42: Remote Control

    ® Getting started R&S NGE100B Instrument control Pressing the rotary knob also confirms the action and returns to the previous screen. To abort the entry ► On the on-screen keypad, select "ESC". The dialog closes without changing the settings. To alternate between lowercase and uppercase ►...
  • Page 43: Operating Basics

    ® Operating basics R&S NGE100B Front panel keys 6 Operating basics 6.1 Front panel keys For an overview of the front panel keys, see Chapter 5.2, "Instrument tour", on page 27. 6.1.1 Function keys The following keys can be categorized based on their functions. Figure 6-1: Function keys of R&S NGE100B Output and channel controls Navigation keys...
  • Page 44: Navigation Keys

    ® Operating basics R&S NGE100B Front panel keys [Ch 1] / [Ch 2] / [Ch 3] Selects the respective channel for output. Chapter 7.1, "Setting the voltage and current", on page 52. Remote command: on page 100 INSTrument:NSELect on page 100 INSTrument[:SELect] [Output] The [Output] key controls the channel output settings of the instrument.
  • Page 45: Memory Functions

    ® Operating basics R&S NGE100B Front panel keys [Voltage] Sets the output voltage for the channel. See Chapter 7.1, "Setting the voltage and cur- rent", on page 52. Remote command: on page 101 [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] on page 103 APPLy [Current] Sets the output current limit for the channel. See Chapter 7.1, "Setting the voltage and current", on page 52.
  • Page 46: Navigation Controls

    ® Operating basics R&S NGE100B Front panel keys Remote command: on page 95 *SAV on page 95 *RCL [Store Recall] Saves/loads instrument settings. Chapter 7.8, "Store and recall", on page 69. Remote command: on page 95 *SAV on page 95 *RCL 6.1.2 Navigation controls Navigation and value settings are done via the rotary knob and arrow keys.
  • Page 47 ® Operating basics R&S NGE100B Front panel keys Access: 1. Press [Menu] key on the front panel. The main menu screen appears. Figure 6-3: Main menu 2. Navigate the menu via the rotary knob or arrow keys. 3. Press the rotary knob or [Enter] key to enter the submenu. 4.
  • Page 48 ® Operating basics R&S NGE100B Front panel keys Protection ← Output Configures the OVP and OPP protection settings for the instrument. See Chapter 7.5, "Protection", on page 58. Fuse ← Output Configures the fuse delay and fuse linking function of the channel. See Chapter 7.3.2, "Fuse delay, fuse linking",...
  • Page 49: Power Derating

    ® Operating basics R&S NGE100B Power derating General ← Device Configures general instrument settings. Chapter 7.10.2, "General instrument settings", on page 84. License ← Device Displays license information and install license options. See Chapter 7.10.3, "License management", on page 85. Reset ←...
  • Page 50: Operation Modes

    ® Operating basics R&S NGE100B Operation modes Figure 6-4: Output performance graph According to the basic electrical formula for power (P) = current (I) x voltage (V), the following results for the maximum power per channel: ● NGE102B: 33.6 W per channel (66 W max for the combination of two channels) ●...
  • Page 51 ® Operating basics R&S NGE100B Operation modes Voltage regulation Figure 6-5: Current limit CC mode The current I corresponds to the current setting adjustable in the instrument. If I reaches I , the instrument switches to CC mode, i.e. the output current remains constant and limited to I even if the load increases.
  • Page 52: Instrument Functions

    ® Instrument functions R&S NGE100B Setting the voltage and current 7 Instrument functions 7.1 Setting the voltage and current The R&S NGE100B comes with the following instrument models: Models Channels R&S NGE102B Ch 1, Ch 2 R&S NGE103B Ch 1, Ch 2, Ch 3 Toggle the respective channel key ([Ch 1], [Ch 2] or [Ch 3]) on the front panel to select these channels.
  • Page 53: Activating The Channel Output

    ® Instrument functions R&S NGE100B Activating the channel output Figure 7-2: Voltage and current settings in the instrument "To go into live-mode" on page 40 for alternative way to configure the voltage and current values. See also "Keypad" on page 85 for setting the keypad fallback time. Set output voltage and current See details in "[Voltage]"...
  • Page 54: Electronic Fuse

    ® Instrument functions R&S NGE100B Electronic fuse Figure 7-3: Output of Ch 3 in CV and CC mode Activating the channel output See details in "[Output]" on page 44. Remote command: on page 105 OUTPut:GENeral on page 104 OUTPut[:STATe] on page 106 OUTPut:SELect 7.3 Electronic fuse To protect a connected, sensitive output in general, the R&S NGE100B power supply...
  • Page 55: Fuse Delay, Fuse Linking

    ® Instrument functions R&S NGE100B Electronic fuse [Fuse] Activates fuse function. Figure 7-4: Fuse indicator on respective channel Remote command: on page 110 FUSE:TRIPped? on page 111 FUSE[:STATe] 7.3.2 Fuse delay, fuse linking The fuse delay and fuse linking functions provide the flexibility to handle a fuse tripped event when it occurs.
  • Page 56 ® Instrument functions R&S NGE100B Electronic fuse Figure 7-5: Fuse dialog 3. Set "Fuse Delay" to a desire value. You can set a value from 0 s to 10 s in a 0.01 s step size. 4. Configure the "Fuse Linking" setting. The R&S NGE100B provides the following options: Example for "Ch 1", the available options are as follows: ●...
  • Page 57: Tracking Function

    ® Instrument functions R&S NGE100B Tracking function "Ch 1 & 2" "Ch 1" and "Ch 2" are linked to selected channel. If selected channel fuse tripped, "Ch 1" and "Ch 2" are turned off. "Ch 1 & 3" "Ch 1" and "Ch 3" are linked to selected channel. If selected channel fuse tripped, "Ch 1"...
  • Page 58: Protection

    ® Instrument functions R&S NGE100B Protection Figure 7-6: Tracking dialog [Track] Activates tracking function. Remote command: n.a. 7.5 Protection The R&S NGE100B provides individual overvoltage protection (OVP) and overpower protection (OPP) settings for each channel. With both functions activated, the R&S NGE100B protects itself and the load in case of voltage or power overloads.
  • Page 59 ® Instrument functions R&S NGE100B Protection Figure 7-8: Protection dialog 3. Set the desired mode and voltage for OVP protection. The R&S NGE100B turns off the output channel when the measured voltage exceeded the preset value. 4. Set the desired mode and power for OPP protection. The R&S NGE100B turns off the output channel when the measured power excee- ded the preset value.
  • Page 60: Digital Trigger I/O

    ® Instrument functions R&S NGE100B Digital trigger I/O <Level> ← Overvoltage Protection (OVP) ← Protection Sets OVP voltage level. You can set a value from 0 V to 32.5 V. The R&S NGE100B turns off the respective channel when the measured voltage exceeded the preset value.
  • Page 61 ® Instrument functions R&S NGE100B Digital trigger I/O ● Trigger input The data lines of the digital I/O interface receive external trigger signal. The exter- nal trigger signal triggers the selected channel (Ch 1, Ch 2, Ch 3) with the set response when the trigger condition ("Active High"...
  • Page 62 ® Instrument functions R&S NGE100B Digital trigger I/O Trigger out parameters Trigger conditions Description Output Off Output the selected logic level when the [Output] is turned off at the selected chan- nel. Fuse Tripped Output the selected logic level when a fuse tripped event occurs on the selected chan- nel.
  • Page 63 ® Instrument functions R&S NGE100B Digital trigger I/O Figure 7-9: Digital IO dialog 3. Select the desired DIO interface for configuration. For configuration details, see "Digital IO" on page 63. Digital IO........................63 └ Master Enable....................63 └ DIO 1 / DIO 2 / DIO 3 / DIO 4.................
  • Page 64: Advanced Functions

    ® Instrument functions R&S NGE100B Advanced functions Channel ← DIO 1 / DIO 2 / DIO 3 / DIO 4 ← Digital IO Sets channel for DIO settings. Remote command: on page 124 TRIGger:IN:SOURce:DIO<IO> on page 126 TRIGger:OUT:SOURce:DIO<IO> Condition ← DIO 1 / DIO 2 / DIO 3 / DIO 4 ← Digital IO Sets trigger condition for the selected Digital I/O line.
  • Page 65: Easyarb

    ® Instrument functions R&S NGE100B Advanced functions 7.7.1 EasyArb EasyArb function The arbitrary function is available only in Ch 1. "Voltage, current settings" on page 52. LabVIEW driver The arbitrary functions in LabVIEW driver rshmc804x do not support the EasyArb fea- ture in the R&S NGE100B.
  • Page 66 ® Instrument functions R&S NGE100B Advanced functions └ EasyArb List table................... 66 └ N......................66 └ Voltage....................66 └ Current....................66 └ Duration....................67 └ Apply EasyArb Data..................67 └ Clear Data Points....................67 EasyArb Settings for EasyArb waveform generator. EasyArb Mode on Ch 1 ← EasyArb Sets EasyArb mode on the output channel.
  • Page 67: Easyramp

    ® Instrument functions R&S NGE100B Advanced functions Remote command: on page 120 ARBitrary:DATA Duration ← EasyArb List table ← EasyArb Sets the duration taken between two data points. Remote command: on page 120 ARBitrary:DATA Apply EasyArb Data ← EasyArb Saves EasyArb configuration. The EasyArb data points are sent to Ch 1.
  • Page 68 ® Instrument functions R&S NGE100B Advanced functions The R&S NGE100B displays the "EasyRamp" dialog. See Figure 7-12. Figure 7-12: EasyRamp dialog 2. Set the desired EasyRamp settings. For details, see "EasyRamp" on page 68. 3. Press [Enter] key to confirm the settings. EasyRamp........................68 └...
  • Page 69: Store And Recall

    ® Instrument functions R&S NGE100B Store and recall 7.8 Store and recall The [Store Recall] function is disabled when the instrument is in editing mode. "Operating mode" on page 37. Upon power-up, the instrument loads the stored settings from memory location M1 with the exception on EasyArb parameters.
  • Page 70 ® Instrument functions R&S NGE100B Store and recall Store instrument settings Access: 1. Long press [Store Recall] key to store instrument settings. The R&S NGE100B displays a message "Store device settings: select any [Mx] key" in the status bar. Figure 7-13: Store instrument settings 2.
  • Page 71: Interfaces And Protocols

    ® Instrument functions R&S NGE100B Interfaces and protocols The R&S NGE100B recalls the instruments settings from the selected memory location. [Store Recall] Saves/loads instrument settings. In the EasyArb and DIO dialog environment, only the EasyArb or DIO settings are saved or loaded, the instrument settings are not saved or loaded. See "Instrument set- tings"...
  • Page 72 ® Instrument functions R&S NGE100B Interfaces and protocols The R&S NGE100B is equipped with a network interface and can be connected to an Ethernet LAN (local area network). A LAN connection is necessary for remote control of the instrument, and for access from a computer using a web browser. Risk of network failure Before connecting the instrument to the network or configuring the network, consult your network administrator.
  • Page 73 ® Instrument functions R&S NGE100B Interfaces and protocols Figure 7-15: Ethernet dialog 3. Configure the required Ethernet settings. See details in "Ethernet" on page 73. 4. Select "Apply" to complete the configuration. 5. Press [Menu] key to exit the dialog. Ethernet.........................73 └...
  • Page 74: Wireless Lan Connection

    ® Instrument functions R&S NGE100B Interfaces and protocols "DHCP & Auto- Automatically assign an IP address to the instrument. IP" Remote command: on page 97 SYSTem:COMMunicate:SOCKet:DHCP[:STATe] IP Address ← Ethernet Address used in communication with the network. This address is manually configured if "IP Mode"...
  • Page 75 ® Instrument functions R&S NGE100B Interfaces and protocols WLAN connection The LAN connection has a higher priority over WLAN. If LAN is connected, the WLAN connection will be disconnected. When LAN is discon- nected, WLAN reconnects automatically. An alterative to connection in local area network is wireless LAN connection. With the presence of an authenticated Wi-Fi signal, the R&S NGE100B automatically connects to a network and navigation can be made via the web browser according to the WLAN IEEE 802.11 b/g/n standards.
  • Page 76 ® Instrument functions R&S NGE100B Interfaces and protocols Figure 7-17: WLAN connected 5. To disconnect, select "Disconnect" in the WLAN dialog. WLAN Settings for WLAN connection. MAC Address ← WLAN Displays MAC address of the R&S NGE100B. Remote command: n.a. Module ←...
  • Page 77: Usb Connection

    ® Instrument functions R&S NGE100B Interfaces and protocols Remote command: on page 97 SYSTem:COMMunicate:WLAN:CONNect on page 98 SYSTem:COMMunicate:WLAN:DISConnect IP Address ← WLAN "IP Address" on page 74. Remote command: on page 98 SYSTem:COMMunicate:WLAN:IPADdress? Subnet Mask ← WLAN "Subnet Mask" on page 74. Remote command: n.a.
  • Page 78 ® Instrument functions R&S NGE100B Interfaces and protocols 2. Set the desired "USB Mode". See details in "USB Mode" on page 79. 3. Press [Enter] key to confirm the settings. The R&S NGE100B displays message to prompt the user to restart the instrument to apply the changes.
  • Page 79: Lxi

    ® Instrument functions R&S NGE100B Interfaces and protocols HMExplorer The window application supports the communication via LAN, USB VCP and WLAN only. For more information of the VISA driver installation, see the installation guide of the VISA driver file. Once the driver is installed, the "USB Test and Measurement Device (IVI)" is displayed in the "Device Manager"...
  • Page 80 ® Instrument functions R&S NGE100B Interfaces and protocols front panel devices via a WEB interface. The central tool for this type of configuration is a browser. LXI-compatible devices are divided into three classes, A, B and C, where the function- ality of the classes is based on each other hierarchically: Figure 7-20: LXI classes ●...
  • Page 81: Scpi

    ® Instrument functions R&S NGE100B Interfaces and protocols Figure 7-21: Instrument web browser view 7.9.4 SCPI SCPI (Standard Commands for Programmable Instruments) commands - messages - are used for remote control. Commands that are not taken from the SCPI standard fol- low the SCPI syntax rules.
  • Page 82 ® Instrument functions R&S NGE100B Interfaces and protocols ● USB TMC USB0::0x0AAD:: 0x0197::5601.1414<device_variant>-<device_serial_number>:: INSTR where k03 is used for 3-ch device, k02 for 2-ch device. (e.g. USB0::0x0AAD::0x0197::5601.1414k03-100689::INSTR) ● USB VCP ASRL<COM>::INSTR, where "COM" refers to the COM port number shown on the controlling PC.
  • Page 83: General Instrument Settings

    ® Instrument functions R&S NGE100B General instrument settings 7.10 General instrument settings The following chapters provide the general instrument information and utilities services in R&S NGE100B. Access: 1. Press [Menu] key. 2. Select "Device" menu item to configure the required instrument settings. 3.
  • Page 84: General Instrument Settings

    ® Instrument functions R&S NGE100B General instrument settings Remote command: on page 93 *IDN? Serial Number ← Information Unique identification number for the instrument. Remote command: on page 93 *IDN? Material Number ← Information Instrument orderable part number. Remote command: on page 93 *IDN? Firmware Version ←...
  • Page 85: License Management

    ® Instrument functions R&S NGE100B General instrument settings Figure 7-23: General dialog General......................... 85 └ Keypad......................85 └ Display......................85 └ Sound......................85 General General instrument settings. Keypad ← General Selects the wait time before the instrument automatically leaves the editing mode. Select "Off"...
  • Page 86 ® Instrument functions R&S NGE100B General instrument settings For unregistered license key code, see also the supplement note delivered together with the unregistered key code. ► Select "License" menu item. The R&S NGE100B displays the "License" dialog. See Figure 7-24. Figure 7-24: License dialog Install via USB To install from an XML file, proceed as follows:...
  • Page 87: Reset Instrument

    ® Instrument functions R&S NGE100B General instrument settings License Installation of instrument license key and information display. Device ID ← License Displays of instrument device ID. Remote command: n.a. License Key Installation ← License Installation of license key settings. Remote command: n.a.
  • Page 88: System Test

    ® Instrument functions R&S NGE100B General instrument settings Figure 7-25: Reset dialog Reset Resets instrument to factory default settings. Remote command: n.a. 7.10.5 System test This function provides a system test to check the instrument working conditions for screen color display, key backlight, beeper and cooling fan. Access: ►...
  • Page 89: Updating The Firmware

    ® Instrument functions R&S NGE100B General instrument settings System Test Lists the system test for instrument. Display Test ← System Test System test to check if display is working. RGB and a list of supported colors should appear on the screen during the test. Remote command: on page 97 SYSTem:TEST:SCReen...
  • Page 90 ® Instrument functions R&S NGE100B General instrument settings 4. Press [Menu] key. 5. Select "Utilities" > "Update" menu item. If firmware file is found, the R&S NGE100B displays the "Update" dialog. See Fig- 7-27. Figure 7-27: Update dialog 6. Press [Enter] key to start firmware update process. The R&S NGE100B reboots automatically as part of the update process.
  • Page 91: Help

    ® Instrument functions R&S NGE100B General instrument settings Remote command: n.a. Start Update from USB ← Device ← Update Displays the firmware update status from USB. Remote command: on page 128 DIAGnostic:SERVice:FWUP 7.10.7 Help The "Help" dialog provides the internet link and QR code for the latest R&S NGE100B information.
  • Page 92: Remote Control Commands

    ® Remote control commands R&S NGE100B Common setting commands 8 Remote control commands This chapter provides the description of all remote commands available for the R&S NGE100B series. The commands are sorted according to the menu structure of the instrument. A list of commands in alphabetical order is given in the "List of Command"s at the end of this documentation.
  • Page 93 ® Remote control commands R&S NGE100B Common setting commands Parameters: <Value> Range: 0 to 255 *ESR? Event status read Returns the contents of the event status register in decimal form and then sets the reg- ister to zero. Return values: <Contents>...
  • Page 94 ® Remote control commands R&S NGE100B Common setting commands *RST Reset Sets the instrument to a defined default status. The default settings are indicated in the description of commands. We recommend to start a program by *RST in order to set the instrument to a defined status prior to starting a program.
  • Page 95: System Commands

    ® Remote control commands R&S NGE100B System commands Prevents servicing of the subsequent commands until all preceding commands have been executed and all signals have settled (see also command synchronization and *OPC). Usage: Event *SAV <Number> Save Stores the current instrument settings under the specified number in an internal mem- ory.
  • Page 96 ® Remote control commands R&S NGE100B System commands SYSTem:LOCal Sets the system to front panel control. The front panel control is unlocked. Usage: Event Manual operation: "[Remote]" on page 45 "[Remote]" on page 82 SYSTem:REMote Sets the system to remote state. The front panel control is locked. By pressing Remote, the front panel control will be activated.
  • Page 97 ® Remote control commands R&S NGE100B System commands Manual operation: "Fan Test" on page 89 SYSTem:TEST:KEYBacklight Runs the system test: Keypad Backlight Test. Usage: Event Manual operation: "Key Backlight Test" on page 89 SYSTem:TEST:SCReen Runs the system test: Screen Color Test. Usage: Event Manual operation:...
  • Page 98 ® Remote control commands R&S NGE100B System commands SYSTem:COMMunicate:WLAN:DISConnect Stops the connection to the predefined wireless access point. Usage: Event Options: R&S NGE-K102 Instrument with serial number below 110000 Manual operation: "Connect" on page 76 SYSTem:COMMunicate:WLAN:IPADdress? Queries the IP address of the device when the wireless interface works as a client. Return values: <arg0>...
  • Page 99: Configuration Commands

    ® Remote control commands R&S NGE100B Configuration commands SYSTem:COMMunicate:WLAN:SSID <arg0> Defines or queries the SSID string of the access point when the wireless interface works as a client. Parameters: <arg0> Options: R&S NGE-K102 Instrument with serial number below 110000 Manual operation: "SSID"...
  • Page 100 ® Remote control commands R&S NGE100B Configuration commands Example: Selecting a channel You can select a channel either with an OUTput parameter, or just by the channel number. This example lists all ways how you can select and query a selected channel. // ************************************************ // Select a channel // ************************************************...
  • Page 101: Voltage Setting

    ® Remote control commands R&S NGE100B Configuration commands Example: INSTrument OUTPut1 INSTrument OUTP1 Selects the output channel as Ch 1. INST? -> 2 Return the output channel as Ch 2. Manual operation: "[Ch 1] / [Ch 2] / [Ch 3]" on page 44 8.3.2 Voltage setting The SOURce:VOLTage subsystem contains the commands for setting the voltage of...
  • Page 102: Current Setting

    ® Remote control commands R&S NGE100B Configuration commands MAXimum 32.200E+01 Default unit: V Example: VOLT 10 Set the voltage value to 10 V. VOLT? -> 10 Return the voltage value of the selected channel. Manual operation: "[Voltage]" on page 45 "Set output voltage and current"...
  • Page 103: Combined Voltage And Current Settings

    ® Remote control commands R&S NGE100B Configuration commands Example: CURR 2 Set the selected channel current to 2 A. CURR? -> 2 Return current value of the selected channel. Manual operation: "[Current]" on page 45 "Set output voltage and current" on page 53 8.3.4 Combined voltage and current settings The APPLy subsystem provides a command that enables you to set the current and...
  • Page 104: Output Setting

    ® Remote control commands R&S NGE100B Configuration commands 8.3.5 Output setting The OUTPut subsystem contains the commands for activating the output channels. Example: Activating the channels You can activate a selected channel and turn on or off the outputs either individually or all outputs simultaneously.
  • Page 105 ® Remote control commands R&S NGE100B Configuration commands Example: INST OUT1 OUTP ON Ch 1 and output state will be activated; i.e. Ch 1 and Output key on the front panel will be illuminated. OUTP? -> 1 Return the output state. Manual operation: "[Output]"...
  • Page 106 ® Remote control commands R&S NGE100B Configuration commands Output mode changed to constant voltage mode. Output mode changed to constant current mode. *RST: Example: OUTP:MODE CV OUTP:MODE? -> CV Output mode has changed to constant voltage mode. OUTPut:SELect <arg0> This command activates or queries the instrument channel state of the previous selected channel.
  • Page 107: Fuse Setting

    ® Remote control commands R&S NGE100B Configuration commands Example: OUTP:TRIP? -> 1 OTP event occurs. Usage: Query only 8.3.6 Fuse setting The FUSE subsystem contains the commands for overcurrent protection parameters such as activating fuses and setting fuse parameters of the output channels. The default unit is A.
  • Page 108 ® Remote control commands R&S NGE100B Configuration commands Example: Configuring fuses This example contains all commands to configure and query the fuse states and set- tings. // ************************************************ // Activate a fuse // ************************************************ INST OUT1 // selects a channel and activates the overcurrent protection FUSE 1 // queries the state of the overcurrent protection in the selected channel FUSE?
  • Page 109 ® Remote control commands R&S NGE100B Configuration commands FUSE:LINK 2 // queries the combined overcurrent protection of the selected channel FUSE:LINK? 2 // response 1 // ************************************************ // Unlink linked overcurrent protection // ************************************************ FUSE:UNLink 2 // queries the combined overcurrent protection of the selected channel FUSE:LINK? 2 // response 0 FUSE:DELay <Delay>...
  • Page 110 ® Remote control commands R&S NGE100B Configuration commands Example: INST OUT1 FUSE:LINK 2 Fuse Ch 1 is linked with Ch 2. FUSE:LINK? 2 -> 1 If the fuse of Ch 1 is linked with Ch 2, a "1" is returned. FUSE:LINK? 3 ->...
  • Page 111: Ovp Setting

    ® Remote control commands R&S NGE100B Configuration commands FUSE[:STATe] <arg0> Defines or queries the fuse for the previous selected channel. Parameters: <arg0> {0 | 1} Fuse is activated. Fuse is deactivated. *RST: OFF | 0 Example: INST OUT1 FUSE ON Fuse of Ch 1 is activated.
  • Page 112 ® Remote control commands R&S NGE100B Configuration commands Example: Configuring the overvoltage protection // ************************************************ // Set the overvoltage protection value // ************************************************ INST OUT1 //activates the OVP of the previous selected channel VOLT:PROT 1 // selects a channel and sets the OVP VOLT:PROT:LEV 5 // queries the output overvoltage value of a channel VOLT:PROT:LEV?
  • Page 113 ® Remote control commands R&S NGE100B Configuration commands [SOURce:]VOLTage:PROTection:MODE <arg0> Defines or queries the OVP mode for the previous selected channel. Parameters: <arg0> MEASured | PROTected MEASured The OVP switches off if the measured value exceeds the thresh- old. PROTected If the adjusted threshold is exceeded the output of the instru- ment will be not switched on;...
  • Page 114: Opp Setting

    ® Remote control commands R&S NGE100B Configuration commands Example: INST OUT1 VOLT:PROT ON OVP of Ch 1 is activated. VOLT:PROT? -> 1 Return the OVP state of the selected channel. Manual operation: "Overvoltage Protection (OVP)" on page 59 [SOURce:]VOLTage:PROTection:LEVel <Voltage> Defines or queries the OVP value of the previous selected channel.
  • Page 115 ® Remote control commands R&S NGE100B Configuration commands Example: Configuring the overpower protection // ************************************************ // Set the overpower protection value // ************************************************ INST OUT1 //activates the OPP of the previous selected channel POW:PROT 1 // selects a channel and sets the OPP POW:PROT:LEV 5 // queries the output overvoltage value of a channel POW:PROT:LEV?
  • Page 116 ® Remote control commands R&S NGE100B Configuration commands Set the OPP value to maximum. Default unit: W Example: INST OUT1 POW:PROT:LEV 5 Set the OPP value of channel 1 to 5 W. POW:PROT:LEV? -> 5 Return the selected channel OPP value. Manual operation: "<Level>"...
  • Page 117: Measurement Commands

    ® Remote control commands R&S NGE100B Measurement commands [SOURce:]POWer:PROTection:CLEar Resets the OPP state of the selected channel. If the OPP has tripped, the red blinking OPP message on the display will be cleared for the selected channel. Usage: Event Manual operation: "Overpower Protection (OPP)"...
  • Page 118: Advanced Operating Commands

    ® Remote control commands R&S NGE100B Advanced operating commands Usage: Query only 8.5 Advanced operating commands The following shows the subsystem that contains the commands for configuring the arbitrary function, EasyRamp and Digital I/O functions. 8.5.1 Arbitrary commands The ARBitrary subsystem contains the commands for configuring an arbitrary sequence for the output channels.
  • Page 119 ® Remote control commands R&S NGE100B Advanced operating commands Example: Configuring an arbitrary sequence This programming example generates an arbitrary sequence for a selected channel. The sequence starts at 1 V and 1 A for 1 sec, and both values are incremented each second by 1.
  • Page 120 ® Remote control commands R&S NGE100B Advanced operating commands ARBitrary:DATA <arg0> Defines or queries the arbitrary points for channel 1. The dwell time between two points is from 10 ms to 600 s. Parameters: <arg0> <Voltage1, Current1, Time1, Voltage2, Current2, Time2,..., Volt- age128, Current128, Time128>...
  • Page 121 ® Remote control commands R&S NGE100B Advanced operating commands ARBitrary:POINts? Returns the number of arbitrary points of EasyArb waveform. Return values: <arg0> Usage: Query only Manual operation: "Number of Data Points" on page 66 ARBitrary[:STATe] <arg0> Sets or queries the EasyArb state ON, i.e. Ch 1 is ready to run arb waveform. Parameters: <arg0>...
  • Page 122: Easyramp Commands

    ® Remote control commands R&S NGE100B Advanced operating commands Manual operation: "Apply EasyArb Data" on page 67 8.5.2 EasyRamp commands The VOTage:RAMP subsystem contains the commands for configuring the ramp func- tion for the output channels. [SOURce:]VOLTage:RAMP:DURation <Time> Sets or queries the EasyRamp duration of the previous selected channel in seconds. 0.01 s to 10 s in steps of 0.01 s.
  • Page 123 ® Remote control commands R&S NGE100B Advanced operating commands TRIGger[:ENABle]:DIO<IO> <arg0> Sets or queries the state of the specified Digital I/O line. The state is disabled when any setting has been changed. Suffix: <IO> 1..4 Parameters: <arg0> {OFF | ON} Enable the state for Digital I/O line.
  • Page 124 ® Remote control commands R&S NGE100B Advanced operating commands TOGGle Output state is toggled. INHibit Output state remains off as long as the TRIG IN signal is activa- ted. ARBitrary Starts the EasyArb function. If level signal is selected, the Easy- Arb function is stopped when TRIG IN signal is deactivated.
  • Page 125 ® Remote control commands R&S NGE100B Advanced operating commands TRIGger:MASTer[:STATe] <arg0> Defines or queries the master enable state for the Digital I/O interface. The state is dis- abled when any setting has been changed. Parameters: <arg0> {OFF | ON} "Master Enable" state is activated for all DIO interfaces. "Master Enable"...
  • Page 126 ® Remote control commands R&S NGE100B Advanced operating commands Suffix: <IO> 1..4 Parameters: <arg0> OVP | OPP | OTP Trigger output is linked to the OVP event. Trigger output is linked to the OPP event. Trigger output is linked to the OTP event. *RST: Manual operation: "Condition"...
  • Page 127: Data And File Management Commands

    ® Remote control commands R&S NGE100B Data and file management commands TRIGger:OUT:VOLTlevel:DIO<IO> <Voltage> Sets or queries the threshold voltage level of the trigger output of the specified Digital I/O line. Triggers when output voltage level exceeds or equals to the set level. Suffix: <IO>...
  • Page 128: Firmware Update

    ® Remote control commands R&S NGE100B Firmware update Usage: Query only HCOPy:SIZE:X? Returns the horizontal expansion of the screenshots. Return values: <arg0> Usage: Query only HCOPy:SIZE:Y? Returns the vertical expansion of the screenshots. Return values: <arg0> Usage: Query only 8.7 Firmware update The following shows the subsystem that contains the commands for firmware update function.
  • Page 129: Troubleshooting

    ® Troubleshooting R&S NGE100B Problems during firmware update 9 Troubleshooting ● Displaying status information................129 ● Problems during firmware update................. 129 ● Problems with remote control over LAN..............130 ● Contacting customer support................130 9.1 Displaying status information The R&S NGE100B displays notifications to inform you on current settings and states. It distinguishes between status, information, and warning and error notifications.
  • Page 130: Problems With Remote Control Over Lan

    ® Troubleshooting R&S NGE100B Contacting customer support 9.3 Problems with remote control over LAN 9.3.1 Cannot establish a LAN connection If you have problems to establish a LAN connection as described in Chapter 7.9.1.1, "LAN connection", on page 71: 1. Check if the TCP/IP address information is valid. 2.
  • Page 131 ® Troubleshooting R&S NGE100B Contacting customer support Figure 9-1: QR code to the Rohde & Schwarz support page User Manual 5601.1343.02 ─ 10...
  • Page 132: Transport

    ® Transport R&S NGE100B 10 Transport Lifting and carrying "Lifting and carrying the product" on page 10. Packing Use the original packaging material. It consists of antistatic wrap for electrostatic pro- tection and packing material designed for the product. If you do not have the original packaging, use similar materials that provide the same level of protection.
  • Page 133: Maintenance, Storage And Disposal

    ® Maintenance, storage and disposal R&S NGE100B Disposal 11 Maintenance, storage and disposal The product does not require regular maintenance. It only requires occasional clean- ing. It is however advisable to check the nominal data from time to time. ● Cleaning........................133 ●...
  • Page 134 ® Maintenance, storage and disposal R&S NGE100B Disposal Figure 11-1: Labeling in line with EU directive WEEE Rohde & Schwarz has developed a disposal concept for the eco-friendly disposal or recycling of waste material. As a manufacturer, Rohde & Schwarz completely fulfills its obligation to take back and dispose of electrical and electronic waste.
  • Page 135: Applications

    ® Applications R&S NGE100B Parallel and series mode 12 Applications 12.1 Parallel and series mode It is assumed that only qualified and trained personnel service the power supplies and the connected loads. Use only the R&S NGE100B power supply series when operating the channels in a series or parallel mode to increase the output voltage and current.
  • Page 136: Parallel Mode

    ® Applications R&S NGE100B Parallel and series mode Figure 12-1: Example of series connection 12.1.2 Parallel mode If it is necessary to increase the total current, the power supply outputs must be wired in parallel. The maximum total current is the sum of the individual currents of all sour- ces connected in parallel.
  • Page 137 ® Applications R&S NGE100B Parallel and series mode Principles of operation in parallel mode Generally, a higher current is first supplied from the channel with the higher output volt- age. Once this channel reaches its power limit, the remaining current is made available by the channel that is connected in parallel.
  • Page 138: Annex

    ® Additional basics on remote control R&S NGE100B Messages and SCPI command structure Annex A Additional basics on remote control A.1 Messages and SCPI command structure A.1.1 Messages Instrument messages are employed in the same way for all interfaces, if not indicated otherwise in the description.
  • Page 139: Scpi Command Structure

    ® Additional basics on remote control R&S NGE100B Messages and SCPI command structure are device-specific, however, their syntax follows SCPI rules as permitted by the standard. Instrument responses Instrument responses (response messages and service requests) are messages which the instrument sends to the controller after a query. They can contain measurement results, instrument settings and information on the instrument status.
  • Page 140 ® Additional basics on remote control R&S NGE100B Messages and SCPI command structure Example: MEASure:CURRent? is equivalent to MEAS:CURR? Case-insensitivity Uppercase and lowercase notation only serves to distinguish the two forms in the man- ual, the instrument itself is case-insensitive. Numeric suffixes If a command can be applied to multiple instances of an object, e.g.
  • Page 141: Scpi Parameters

    ® Additional basics on remote control R&S NGE100B Messages and SCPI command structure A.1.3 SCPI parameters Many commands are supplemented by a parameter or a list of parameters. The parameters must be separated from the header by a whitespace (ASCII code 0 to 9, 11 to 32 decimal, e.g.
  • Page 142: Overview Of Syntax Elements

    ® Additional basics on remote control R&S NGE100B Messages and SCPI command structure Example: OUTP:STAT ON OUTP:STAT? Response: 1 Text Text parameters observe the syntactic rules for key words, i.e. they can be entered using a short or long form. Like any parameter, they have to be separated from the header by a white space.
  • Page 143: Responses To Queries

    ® Additional basics on remote control R&S NGE100B Messages and SCPI command structure Table A-2: Syntax elements A colon separates the mnemonics of a command. The semicolon separates two commands within a program message. It does not alter the path. A comma separates several parameters of a command.
  • Page 144: Command Sequence And Synchronization

    ® Additional basics on remote control R&S NGE100B Command sequence and synchronization Response: 1 A.2 Command sequence and synchronization A sequential command finishes the execution before the next command is starting. To make sure that commands are actually carried out in a certain order, each command must be sent in a separate command line.
  • Page 145 ® Additional basics on remote control R&S NGE100B Command sequence and synchronization For time consuming overlapped commands, you can allow the controller or the instru- ment to do other useful work while waiting for command execution. Use one of the fol- lowing methods: *OPC with a service request 1.
  • Page 146 ® Additional basics on remote control R&S NGE100B Command sequence and synchronization using SCPI commands. A thread waiting for an *OPC? response does not block the GUI or the communication with other instruments. User Manual 5601.1343.02 ─ 10...
  • Page 147: List Of Commands

    ® List of commands R&S NGE100B List of commands [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]..................102 [SOURce:]POWer:PROTection:CLEar......................117 [SOURce:]POWer:PROTection:LEVel......................115 [SOURce:]POWer:PROTection:TRIPped?.....................116 [SOURce:]POWer:PROTection[:STATe]......................116 [SOURce:]VOLTage:PROTection:CLEar......................114 [SOURce:]VOLTage:PROTection:LEVel......................114 [SOURce:]VOLTage:PROTection:MODE....................... 113 [SOURce:]VOLTage:PROTection:TRIPped?....................113 [SOURce:]VOLTage:PROTection[:STATe]......................113 [SOURce:]VOLTage:RAMP:DURation......................122 [SOURce:]VOLTage:RAMP[:STATe]......................122 [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude].................. 101 *CLS.................................92 *ESE................................92 *ESR?................................93 *IDN?................................93 *OPC................................93 *OPT?................................93 *RCL................................95 *RST................................94 *SAV.................................95 *SRE................................
  • Page 148 ® List of commands R&S NGE100B MEASure[:SCALar]:CURRent[:DC]?......................117 MEASure[:SCALar]:POWer?......................... 117 MEASure[:SCALar][:VOLTage][:DC]?......................117 OTP:TRIPped?.............................. 106 OUTPut:GENeral............................105 OUTPut:MODE.............................. 105 OUTPut:SELect............................. 106 OUTPut[:STATe].............................104 SYSTem:BEEPer[:IMMediate]......................... 95 SYSTem:COMMunicate:SOCKet:DHCP[:STATe].................... 97 SYSTem:COMMunicate:SOCKet:IPADdress?....................97 SYSTem:COMMunicate:WLAN:CONNect....................... 97 SYSTem:COMMunicate:WLAN:DISConnect....................98 SYSTem:COMMunicate:WLAN:IPADdress?....................98 SYSTem:COMMunicate:WLAN:PASSword..................... 98 SYSTem:COMMunicate:WLAN:SSID......................99 SYSTem:COMMunicate:WLAN[:STATe]......................98 SYSTem:LOCal..............................96 SYSTem:OPTion?............................96 SYSTem:REMote.............................96 SYSTem:RWLock............................96 SYSTem:TEST:BEEPer...........................
  • Page 149: Index

    ® Index R&S NGE100B Index Symbols *OPC ................144 Data entry ................40 *OPC? ................144 Data sheet ................. 15 *WAI ................144 Default gateway ..............74 Default values Remote ............... 94 Device-specific commands Access SCPI syntax .............. 139 Instrument ..............39 Digital IO ................
  • Page 150 ® Index R&S NGE100B Ground terminal: Maintenance, storage and disposal ........ 133 How to ground terminal ..........25 Changing fuses ............133 Cleaning ..............133 Disposal ..............133 Storage ..............133 Help ................15, 91 Manual interaction ............. 34 How to: Manual operation ..............
  • Page 151 ® Index R&S NGE100B Preparing for use ............... 19 SCPI parameters ............. 141 Protection ................58 Block data ..............142 Overpower protection (OPP) ........58 Boolean ..............141 Overvoltage protection (OVP) ........58 Character strings ............142 Protocol Numeric values ............141 LXI ................
  • Page 152 ® Index R&S NGE100B USB connection USB TMC ..............78 USB VCP ..............78 User manual ..............15 Using the front panel keys ..........34 Wait Remote ............... 94 Ways of operation ............. 34 User Manual 5601.1343.02 ─ 10...