Dwyer Instruments Magnehelic 605-1 Installation And Operating Instructions

Dwyer Instruments Magnehelic 605-1 Installation And Operating Instructions

Differential pressure indicating transmitter

Advertisement

Series 605 Magnehelic
Specifications - Installation and Operating Instructions
The Series 605 Magnehelic
Differential Pressure Indicating Transmitter
®
simultaneously provides local indication on a large, easily read analog scale while
also converting that pressure into a standard two-wire, 4-20 mA signal for ranges from
0-0.5 to 0-50 in w.c. Positive, negative or differential air and compatible gas pressure
can be measured with accuracies as low as ±.5% of full scale. The basic mechanical
components of the Series 605 Magnehelic
used in the popular, time-proven Magnehelic
STANDARD ACCESSORIES
Mounting ring
Snap ring
(4) 6-32 x 1-1/4 screws (panel mtg.)
(3) 6-32 x 5/16 screws (surface mtg.)
(2) Tubing to 1/8˝ NPT adapters
(2) 1/8˝ NPT plugs
Adjustment key

MODEL CHART

Range
Maximum
Model
in w.c.
Pressure
605-00N
.05-0-.20
10 psi (68.95 kPa)
605-11
.25-0-.25
10 psi (68.95 kPa)
605-0
0-.50
10 psi (68.95 kPa)
605-1
0-1.0
10 psi (68.95 kPa)
605-2
0-2.0
2 psi (13.79 kPa)
605-3
0-3.0
2 psi (13.79 kPa)
605-6
0-6.0
2 psi (13.79 kPa)
605-10
0-10
2 psi (13.79 kPa)
605-20
0-20.0
11 psi (75.8 kPa)
605-30
0-30
11 psi (75.8 kPa)
605-50
0-50
11 psi (75.8 kPa)
Range
Maximum
Model
in Pa
Pressure
605-12
60-0-60
10 psi (68.95 kPa)
605-13
100-0-100
10 psi (68.95 kPa)
605-60PA
0-60
10 psi (68.95 kPa)
605-125PA
0-125
10 psi (68.95 kPa)
605-250PA
0-250
10 psi (68.95 kPa)
605-500PA
0-500
2 psi (13.79 kPa)
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
®
(3) #6-32 x 3/16 [4.76]
DP HOLES EQUALLY SPACED
ON A Ø4-1/8 [104.78] B.C.
FOR PANEL MOUNTING
indicating transmitter are similar to those
®
differential pressure gage.
®
Electrical
Mechanical
Accuracy ±%
Accuracy ±%
4
4
2
3
2
3
2
2
0.5
2
0.5
2
0.5
2
0.5
2
0.5
2
0.5
2
0.5
2
Electrical
Mechanical
Accuracy ±%
Accuracy ±%
4
4
2
2
2
4
2
3
2
2
0.5
2
Differential Pressure Indicating Transmitter
23/32
1/8 FEMALE
[18.26]
NPT HIGH
PRESSURE
1
CONNECTION
[25.40]
1-3/4
30°
[44.45]
1/2 [12.70]
1-1/8
1/8 FEMALE NPT
[28.58]
LOW PRESSURE
11/16
CONNECTION
[17.46]

SPECIFICATIONS

Gage Specifications
Service: Air and non-combustible, compatible gases.
Wetted Materials: Consult factory.
Accuracy: See chart.
Stability: ±1% FS/yr.
Pressure Limits: See chart.
Temperature Limits: 20 to 120°F (-6.67 to 48.9°C).
Process Connections: 1/8˝ female NPT.
Size: 4˝ (101.6 mm) dial face, 5˝ (127 mm) O.D. x 2-11/16˝ (68.3 mm); For -SS
Bezel: 4-3/4˝ (120.7 mm) O.D. x 2-21/32˝ (67.5mm).
Weight: 1 lb, 12.6 oz (811 g).
Agency Approvals: CE.
TRANSMITTER SPECIFICATIONS
Accuracy: See chart (includes linearity, hysteresis, repeatability).
Temperature Limits: 20 to 120°F (-6.67 to 48.9°C).
Compensated Temperature Range: 32 to 120°F (0 to 48.9°C).
Thermal Effect: ±0.025% FS/°F (0.045% FS/°C).
Power Requirements: 10-35 VDC (2-wire).
Output Signal: 4-20 mA.
Zero and Span Adjustments: Protected potentiometers.
Loop Resistance: DC; 0-1250 Ω maximum.
Current Consumption: DC; 38 mA maximum.
Electrical Connections: Screw terminal block.
Mounting Orientation: Diaphragm in vertical position. Consult factory for other
position orientations.
Phone: 219/879-8000
Fax: 219/872-9057
Bulletin E-68
(4) 6-32 HOLES
EQUALLY SPACED ON
2-1/16
A 5-1/8 [130.18] B.C.
[52.39]
2
FOR FLUSH MOUNTING
[50.80]
Ø4-47/64
1-1/4
[ 120.25]
[31.75]
Ø5
[127.00]
5/8 [15.88]
Ø4 [101.60]
PANEL MAX
FACE
2-1/2
3/16
5-1/2 [139.70]
[63.50]
[4.76]
O.D. MOUNTING
RING
www.dwyer-inst.com
e-mail: info@dwyermail.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Magnehelic 605-1 and is the answer not in the manual?

Questions and answers

Summary of Contents for Dwyer Instruments Magnehelic 605-1

  • Page 1 10 psi (68.95 kPa) 605-125PA 0-125 10 psi (68.95 kPa) 605-250PA 0-250 10 psi (68.95 kPa) 605-500PA 0-500 2 psi (13.79 kPa) DWYER INSTRUMENTS, INC. Phone: 219/879-8000 www.dwyer-inst.com P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: info@dwyermail.com...
  • Page 2 INSTALLATION 1. LOCATION: Select a location where the temperature of the unit will be between 20°F and 120°F. Distance from the receiver is limited only by total loop resistance. See “Electrical Connections”. The tubing feeding pressure to the instrument can 1-1/8˝...
  • Page 3 ELECTRICAL CONNECTIONS 1400 1300 MAXIMUM VALUE (1250) CAUTION: DO EXCEED SPECIFIED SUPPLY VOLTAGE RATINGS. RL MAX = VPS - 10.0V 1200 PERMANENT DAMAGE NOT COVERED BY WARRANTY WILL RESULT. THIS UNIT 20 mA DC IS NOT DESIGNED FOR AC VOLTAGE OPERATION. 1100 1000 Electrical connections to the Series 605 Transmitter are made to the two-screw...
  • Page 4 The minimum current requirement at a given voltage can be calculated by multiplying the number of units x 20 mA. In the example shown this would be 4 x 20 or 80 mA minimum. ©Copyright 2020 Dwyer Instruments, Inc. Printed in U.S.A. 2/20 FR# 440559-00 Rev. 11 DWYER INSTRUMENTS, INC.

This manual is also suitable for:

Magnehelic 605-00nMagnehelic 605-11Magnehelic 605-0Magnehelic 605 seriesMagnehelic 605-2Magnehelic 605-3 ... Show all

Table of Contents