10. Follow the directions on the screen
and push the 3-Liter syringe plunger
in over a period of approximately
6 seconds for the last SSD verifica-
tion.
11. The final "SSD CAL" screen (Figure
13) will display the volume of the
calibration syringe, the corrected
measured volume and associated
percentage error for each of the three
flow rates indicated on the illustra-
tion.
12. Press DONE to save the calibration
results, DISCARD to delete, or
PRINT to print out a record of the
results. See Printing Results on page
38 for more information on printing.
NOTE: When performing either the
CAL check or SSD CAL maneuver, if
the measured flow or volume is not
within allowable range, the screen
will indicate "Unsuccessful CAL Try
Again?" Refer to the Calibration Error
section of the Troubleshooting Guide
on page 52 to resolve the problem.
P-495220-00 Rev. D