HORIZON
Model 24 Flex-AH
Operator's Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HORIZON 24 Flex-AH and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Drucker Diagnostics HORIZON 24 Flex-AH

  • Page 1 HORIZON Model 24 Flex-AH Operator’s Manual...
  • Page 2: Table Of Contents

    ABLE OF ONTENTS Symbols ................................1 Model Description .............................. 2 Features ..................................2 Intended Use ................................2 Warranty ..................................2 Caution and Warning Statements ........................3 Initial Setup ................................. 4 Operation ................................4 Quick Start ................................5 Display Cycle Count ............................. 6 Add a New Program Cycle ...........................
  • Page 3: Symbols

    Denotes conformity to specific UK directives and regulations. FDA LISTED FDA Listed Denotes that the product has been properly listed with the FDA. Denotes conformity to quality standards and quality ISO Certification management systems. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 1...
  • Page 4: Model Description

    General purpose laboratory centrifuge intended for the density-based separation of fluids through centripetal acceleration. ARRANTY Drucker Diagnostics warrants that this centrifuge is free from defects in workmanship and parts for two (2) years. Drucker Diagnostics – Customer Service: +1-814-692-7661 –...
  • Page 5: Caution And Warning Statements

    Refer to General Specifications section for maximum carrier load specification. WARNING: Maximum sample weight per bucket is 60 g. WARNING: Only use Drucker Diagnostics components in this centrifuge. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com...
  • Page 6: Initial Setup

    60 seconds have elapsed, press the UNLOCK button. The lid will unlock for another 15 seconds. o Turn the lid knob counterclockwise and open the lid. The lid light will turn off. OU MAY NOW SAFELY REMOVE THE SAMPLES Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 4...
  • Page 7: Quick Start

    Pressing the UNLOCK button during operation will terminate the Stop run and unlock the lid after the rotor has come to a stop. Begins running the cycle displayed on the screen. The lid must be Start closed. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 5...
  • Page 8: Display Cycle Count

    To change the value shown on the BRAKE display, use the up and Values down buttons next to the BRAKE screen. Press the GEAR button again to save. Save the Save the program by pressing the START button. Setting Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 6...
  • Page 9: Naming A Newly Added Program Cycle

    Next Character Exit Cursor Press the GEAR button twice to exit cursor mode. Mode Save Setting Press the START button to save your cycle name. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 7...
  • Page 10: Deleting A Program Cycle

    UNLOCK button for 5 seconds. Two beeps and a flash of the lid Cycle Lock light will confirm that cycle selection is now unlocked. ETTINGS Time G-Force 1,500 5 minutes 420 xg 2,500 4 minutes 1,170 xg ADIPOS Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 8...
  • Page 11: Loading

    Opposing buckets must be equally loaded or empty or loaded with equally weighted samples. All buckets in rotor must be from the same manufacturing lot. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 9...
  • Page 12: Care And Preventative Maintenance

    ONLY isopropyl alcohol or a 10% (5500 PPM) bleach solution should be used to disinfect the centrifuge and its accessories. o All surfaces must be dried immediately after cleaning and disinfecting. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 10...
  • Page 13: Calibration Testing

    “Balancing continuously Loads” section of this manual. The centrifuge is stuck on one of the Cycle selection is locked. Press the UNLOCK button for 5 seconds. settings Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 11...
  • Page 14 If the lid remains locked after this and will not unlock, the electronics may have been damaged. Contact customer service for assistance. Lid does not stay up o Tighten the center screw on the lid hinge. Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 12...
  • Page 15: General Specifications

    In Inches: RCF or G-force = 0.00001118 x RCF or G-force = 0.0000284 x Rotor Radius (cm) x (RPM) Rotor Radius (in) x (RPM) 6.6 in (16.7 cm) Radius Drucker Diagnostics – Customer Service: +1-814-692-7661 – CustomerService@DruckerDiagnostics.com | Page 13...
  • Page 16 Product Family: HORIZON Flex Series (HORIZON 24 Flex-AH, 24 Flex-AV) Complies with UL61010-1/CSA C22.2 No. 61010-1 and IEC61010-2-020 Protected by U.S. Patents #6,811,531, # 7,422,554, #D718,463, & #D734,489. Other Patents Pending FDA LISTED INSTRUCTIONS FOR DISPOSAL OF WEEE BY USERS IN THE EUROPEAN UNION This product must not be disposed of with other waste.