Flow mode (ml/hr)

Advertisement

Short Programming Guide - Flow Mode (mL/hr)
Caution:
Users must familiarise themselves with the full user instructions.
These notes provide only a reminder of main functions.

Pump activation and mode selection

The default delivery setting for the S-PID 30 is timed mode. This allows
programming of the duration of infusion. To change the mode of operation
remove and then replace the battery.
The pump will perform a short self test procedure during which a full screen
display of all elements should appear.
When the symbol [-t-] is displayed, press and hold the [P] button until the
display flashes. Press the [-] button until the display changes to [-F-]. Press
[P] to confirm and the pump will switch to [OFF] mode.
.

Locking/unlocking the keypad

B
y default keyboard lock is not activated on this device.
With the pump displaying [OFF] press the (-) key for approximately 7
seconds. A number representing the number of completed infusions will
appear on the screen.
While still holding the [-] key depress the [+] key.The screen will display
flashing [L.0] and the keyboard is unlocked.
Use the (+) or (-) button to select required value. (L1 is on, L0 is off)
This symbol will display when the keyboard is locked.
Crono S-PID 30

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Crono S-PID 30 and is the answer not in the manual?

Questions and answers

Summary of Contents for AMT Crono S-PID 30

  • Page 1 Crono S-PID 30 Short Programming Guide - Flow Mode (mL/hr) Caution: Users must familiarise themselves with the full user instructions. These notes provide only a reminder of main functions. Pump activation and mode selection The default delivery setting for the S-PID 30 is timed mode. This allows programming of the duration of infusion.
  • Page 2 Programming the pump With the keyboard unlocked, and the pump displaying [OFF] pressing the [P] button will allow you to step through a menu to set the following parameters in turn. End of infusion alarm On/Off Syringe Partial Volume 1ml steps Setting procedure Press the [P] button to select alarm setting.

Table of Contents