INSTRUCTION MANUAL
NO
/ NO Converter
2
Model: ZDL03
Preface
We thank you very much for purchasing Fuji's NO
Converter.
● Please read this instruction manual thoroughly
before installing, using and applying maintenance
this device.
Damage or accident may be caused when
mishandled.
● Specification of this NO
without notice for modification.
● Do not remodel or modify this NO
without the manufacturer's permission. Fuji Electric
will not accept any liability whatsoever for any
trouble or accident caused by such modification.
● Operator of the NO
/NO converter should keep this
2
instruction manual.
● Operator should keep this instruction manual near at
hand at all times, after thoroughly reading it.
Safety precautions
Please read the safety precautions written as bellow
before use, for correct use of the converter.
Please observe cautions stated bellow, for it contains
important information on safety.
Precautions under this mark is stated
when wrong handling may cause
hazardous situation. Possibility of
CAUTION
Medium level damage or injury, and
physical damage is predicted.
● Installation, wiring and piping should be carried
out by professionals or suppliers. Incomplete
installation may cause fall of the device, electric
shock or fire.
● Gas analyzer should be turned OFF when wiring,
maintenance or inspection is carried out. This is to
prevent electric shock and injury.
INZ-TN1ZDL03-E
/NO converter may alter
2
/NO converter
2
CAUTION
Head Office
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fesys.co.jp/eng
Instrumentation Div.
International Sales Div.
No. 1, Fuji-machi, Hino-city, Tokyo 191-8502, Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http://www.fic-net.jp/eng
/NO
2
● Use wires with proper wire rods and diameters
that meets this device. Wrong ones may cause
electric shock or fire.
● Do not insert metal rod or fingers to the power
supply terminal. It may cause electric shock.
● Remove any metal objects such as wristwatch
while operating maintenance or inspection. It may
cause electric shock.
● Use pipes with proper material and joints stated in
the instruction manual. It may cause gas leakage.
● Use replacement parts that are specified by the
manufacturer. Otherwise, it may cause
malfunction, electric shock or gas leakage.
● Do not touch the converter section, due to high
temperature. Wear protective gloves while
exchanging catalyst. It may cause burn injury.
1. Overview
The NO
/NO converter is to be coupled with a NOx gas
2
analyzer or NH
converter using a special catalyst which efficiently
converts NO
2. Major Specifications
Catalyst
Gas flow rate
Set temperature
Thermocouple
Power supply
Power consumption : About 85VA
Material of parts that contact gas
Gas inlet/outlet connection method
- 1 -
CAUTION
gas analyzer for flue exhaust. It is
3
in sample gas to NO.
2
: Carbon, replacement required every
8 months (when NO
is 10 ppm or lower)
: About 0.5L/min.
: 220±10°C
: K
: 100V AC, 50/60Hz
: Ceramic, Viton, Glass wool,
SUS316
: Insert Teflon tube φ6 mm/φ4 mm
into Viton connection port of inner
diameter φ5.5 mm. (Withstand
pressure: 10 kPa)
concentration
2

Advertisement

loading
Need help?

Need help?

Do you have a question about the ZDL03 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Fuji Electric ZDL03

  • Page 1 ● Do not remodel or modify this NO /NO converter the instruction manual. It may cause gas leakage. without the manufacturer’s permission. Fuji Electric ● Use replacement parts that are specified by the will not accept any liability whatsoever for any manufacturer.
  • Page 2 (b) Under SV (Set Value) display, set 220°C by pressing 3. Outline Drawing key and determine it by the This gas converter is composed of the converter block and ENT key. (200°C need be set for NO/CO the temperature regulator block. analyzer.) [150] (c) Run automatic tuning when temperature is unstable.
  • Page 3 6. Scope of Delivery (1) Main unit (2) NO /NO catalyst (already set into the main unit) (3) Glass wool (already set into the main unit) 7. Spare Parts for One Year 1. ZBN1DL72 5ppm or lower (0.5L/min flow late) 10ppm or lower (0.3L/min flow late) 2.