Durr Dental VistaScan Nano Easy Installation And Operating Instructions Manual

Durr Dental VistaScan Nano Easy Installation And Operating Instructions Manual

Hide thumbs Also See for VistaScan Nano Easy:

Advertisement

Quick Links

VistaScan Nano Easy
EN
Installation and operating instructions
2160100227L02
*2160100227L02*

Advertisement

loading
Need help?

Need help?

Do you have a question about the VistaScan Nano Easy and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Durr Dental VistaScan Nano Easy

  • Page 1 VistaScan Nano Easy Installation and operating instructions 2160100227L02 *2160100227L02*...
  • Page 3 Contents Contents Light protection cover ..Important information About this document ....Assembly Warnings and symbols ..Requirements .
  • Page 4: Table Of Contents

    Contents Troubleshooting 13 Tips for operators and service techni- cians ......13.1 Poor X-ray image ... . 13.2 Software error .
  • Page 5 Important information Note, e.g. specific instructions regarding Important information efficient and cost-effective use of the unit. About this document Comply with the Operating Instructions. These installation and operating instructions rep- resent part of the unit. Wear protective gloves. If the instructions and information in these installation and operating instructions are not followed, Dürr Dental will not be able Switch off and de-energise the unit...
  • Page 6 Copyright information Intended purpose All circuits, processes, names, software pro- grams and units mentioned in this document are VistaScan Nano Easy protected by copyright. The unit is intended exclusively for use in dental The Installation and Operating Instructions must applications for the scanning and processing of not be copied or reprinted, neither in full nor in image data on an image plate.
  • Page 7 Never touch the patient and unshielded plug ❯ The VistaScan Nano Easy unit does not possess connections or metallic parts of the device at any significant performance characteristics as set the same time.
  • Page 8 Important information 2.11 Protection from threats from Dürr Dental accepts no liability for dam- the Internet ages or injury resulting from the use of non-approved accessories or optional The unit is to be connected to a computer that accessories, or from the use of non-origi- can be connected to the Internet.
  • Page 9 Product description Product description Overview b ac k VistaScan Nano Easy image plate scanner Cartridge for image plates (S0 up to S2) IQ image plate Light protection cover Plus VistaSoft imaging software DVD Network cable (3 m) Power supply unit with country-specific adapter...
  • Page 10 VistaScan Nano Easy image plate scanner ... 2160110013 Optional accessories – VistaScan Nano Easy basic unit The following optional articles can be used with – Power supply unit the unit: –...
  • Page 11 Product description Light protection covers see "3.2 Accessories" Wear parts and replacement parts Image plates see "3.2 Accessories" Information about replacement parts is available from the portal for authorised specialist dealers at: www.duerrdental.net. 2160100227L02 1902V005...
  • Page 12 Product description Technical data Image plate scanner Electrical data for the unit Rated voltage V DC Max. current consumption Max. power consumption < 12 Protection class Electrical data – power supply unit Nominal input voltage V AC 100 - 240 Frequency 50/60 Nominal output voltage...
  • Page 13 Product description Ambient conditions during storage and transport Humidity 10 - 95, non-condensing Air pressure 500 - 1060 Height above sea level < 16000 Classification Medical Devices Directive (93/42/EEC) Class I Laser class (unit) in accordance with EN 60825-1:2014 Electromagnetic compatibility (EMC) Interference emission measurements High-frequency emissions in accordance with CISPR 11 Group 1...
  • Page 14 Product description Immunity to interference table, near fields of wireless HF communication devices Radio service Frequency band Test level GSM 800/900 TETRA 800 iDEN 820 800 - 960 CDMA 850 LTE band 5 GSM 1800 CDMA 1900 GSM 1900 1700 - 1990 DECT LTE band 1, 3, 4, 25 UMTS...
  • Page 15 Product description Electromagnetic compatibility (EMC) Interference immunity measurements SIP/SOP Immunity to interference, rapid transient bursts – I/O, SIP/SOP ports IEC 61000-4-4:2012 Compliant ± 1 kV 100 kHz repetition frequency Immunity to interference, line-conducted disturbances induced by high-frequency fields – SIP/SOP ports IEC 61000-4-6:2013 0.15 - 80 MHz Compliant...
  • Page 16 Product description Operation Type plate The type plate is located on the rear of the Image plate scanner device. Input unit Cover (open) Operating elements Collection tray REF Order number The image plate scanner is used to read image data stored on an image plate and to transfer the Serial number data to the imaging software (e.g.
  • Page 17 Product description Operating elements Cartridge for S1 image plate is in the device Cartridge for S2 image plate is in the device Status LED flashing Connections The connections are located on the rear of the unit, underneath the cover. 4 3 2 On / off switch Error display yellow Read display blue...
  • Page 18 Product description Image plate The image plate stores X-ray energy, which is re- emitted in the form of light after excitation via the laser. This light is then converted to image infor- mation in the image plate scanner. The image plate has an active side and an inac- tive side.
  • Page 19 Assembly Installation Assembly Setting up the unit Only qualified specialists or employees NOTICE trained by Dürr Dental are permitted to Risk of damage to sensitive compo- install, connect and start using the unit. nents in the unit as a result of shocks or vibrations Requirements Do not expose the unit to any strong...
  • Page 20 Assembly Remove the screw covers and screws from the ❯ For details of the installation refer to the bottom of the device. installation instructions for the wall mount- ing (order number 2160100277). Electrical connections Safety when making electrical connections The device must only be connected to a cor- ❯...
  • Page 21 Assembly Plug in the connecting plug of the power sup- Combining devices safely ❯ ply unit into the socket connection of the – The overall safety of the unit and its main per- device. formance features are independent of the net- work.
  • Page 22 Assembly Commissioning Connecting the unit via the network cable Remove the cover from the rear of the device. ❯ Connect the supplied network cable to the net- ❯ NOTICE work connection of the device. Short circuit due to the build up of condensation Do not switch on the unit until it has ❯...
  • Page 23 Assembly Mark the connected unit in the list. Configuring the appliance in ❯ DBSWIN Configuration is carried out using VistaNetConfig, which is automatically installed during installation of DBSWIN or VistaEasy. Select Start > All Programs > Dürr Dental > ❯ VistaConfig >...
  • Page 24 Assembly Click Apply to save the configuration. ❯ Entering a fixed IP address (recommended) To reset the network settings, keep the unit reset key pressed for 15 - 20 sec- onds while switching on. Deactivate DHCP. ❯ Enter the IP address, subnet mask and gate- ❯...
  • Page 25 Assembly X-ray unit settings If 60 kV can be set on the X-ray unit, this setting is preferred. The standard exposure values for F-speed film (e. g. Kodak Insight) can be used. Intraoral X-ray units for an adult patient The following table shows the standard values for the exposure time and the dose area product of an image plate for an adult patient.
  • Page 26 Assembly DC emitter, 6 mA Tube length 30 cm Without X-ray field X-ray field limitation X-ray field limitation limitation 60 kV 60 kV 60 kV mGycm mGycm mGycm Incisors 0.10 s 0.10 s 0.10 s Premolars 0.13 s 11.8 0.13 s 0.13 s Molars 0.16 s...
  • Page 27 Usage Usage CAUTION Image plates are toxic Image plates that are not packed in a Correct use of image light protection cover can lead to poi- plates soning when placed in the mouth or swallowed. WARNING Only place image plates in the ❯...
  • Page 28 Usage Protect the image plates against sunlight and 10 Operation ❯ ultraviolet light. Store image plates in a light protection cover or CAUTION foil cassette of the correct size. The image data on the image plate is Image plates will be pre-exposed on exposure ❯...
  • Page 29 Usage Check that the correct cartridge has been Remove cartridge. ❯ ❯ inserted with the display (S0, S1, S2). If the incorrect cartridge has been inserted, it must be removed and the correct cartridge inserted. Open the cover. ❯ The green status LED flashes. The green car- tridge display extinguishes.
  • Page 30 Usage Pull off the adhesive strip, fold down the flap ❯ WARNING and close the light protection cover tightly by pressing together firmly. Risk of cross contamination when not using the light protection cover or when using the light protection cover more than once Do not use an image plate without a ❯...
  • Page 31 Usage Place the image plate in the light protection Disinfect the light protection cover and protec- ❯ ❯ cover into the patient's mouth. tive gloves with a suitable disinfection wipe; When doing this, make sure that the active side see "11.2 Light protection cover". of the image plate points towards the X-ray tube.
  • Page 32 Usage 10.3 Scanning the image data Place the light protection cover with the image ❯ plate centrally and straight onto the input unit. Starting the image plate scanner and soft- The torn-off side of the light protection cover ware faces down; the inactive (back) side of the image plate faces the operator.
  • Page 33 Usage Slide the image plate out of the light protection Once the image plate has been inserted into ❯ ❯ cover downwards into the unit. The image the unit, close the cover and leave it closed plate must be inserted fully into the input unit. throughout the entire scanning process.
  • Page 34 Usage 10.4 Erasing the image plate 11 Cleaning and disinfection The image data is automatically erased after When cleaning and disinfecting the unit and its scanning. accessories, observe country-specific directives, The special ERASE mode only activates the era- standards and specifications for medical prod- sure unit of the image plate scanner.
  • Page 35 Usage product has been subjected to material compati- Cartridges (S0-S2) bility testing by Dürr Dental. The cartridges can be cleaned and disinfected with a wipe disinfection or in an immersion disin- NOTICE fection system. Heat or humidity will damage the NOTICE image plate.
  • Page 36 Usage 12 Maintenance 12.1 Recommended maintenance schedule Prior to working on the device or in case of danger, disconnect it from the mains (e. g. pull the mains plug). The recommended maintenance intervals are based on using the device for 25 intraoral images per day on 220 working days per year.
  • Page 37: Troubleshooting

    Troubleshooting Troubleshooting 13 Tips for operators and service technicians Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our service. Prior to working on the device or in case of danger, disconnect it from the mains (e. g. pull the mains plug).
  • Page 38 Troubleshooting Error Possible cause Remedy X-ray image only shadowy The X-ray dose on the image Increase X-ray dose. ❯ plate was insufficient Amplification (HV value) is set Increase amplification (HV ❯ too low in the software value). Unsuitable scanning mode Select a suitable scanning ❯...
  • Page 39 Troubleshooting Error Possible cause Remedy Software unable to combine The X-ray dose on the image Increase X-ray dose. ❯ plate was insufficient the data to make a complete image Amplification (HV value) is set Increase amplification (HV ❯ too low in the software value).
  • Page 40: Software Error

    Troubleshooting Error Possible cause Remedy The X-ray image shows a pre- After the light protection cover Do not push out the image ❯ has been torn off and before plate until the torn-off light erasure at one end pushing into the input unit, the protection cover has been image plate is pushed out of the placed on the input unit.
  • Page 41: Fault On The Unit

    Troubleshooting Error Possible cause Remedy Error message "E2490" The connection to the unit was Restore the connection to the ❯ interrupted while the software unit. was still attempting to communi- Repeat the process. ❯ cate with the unit Error during data transmission Connecting cable used is incor- Only use original cables.
  • Page 42 Troubleshooting Error Possible cause Remedy Unit not responding The unit has not yet completed After switching on, wait 20 - ❯ the startup procedure 30 seconds until the startup procedure has finished. Unit is blocked by the firewall Enable the ports for the unit in ❯...
  • Page 43: Appendix

    Appendix Appendix 14 Scanning times The scanning time corresponds to the time taken for complete scanning of image data and depends on image plate format and pixel size. The time to image will depend largely on the computer system used and its work load. Times stated are approximate.
  • Page 44: File Sizes (Uncompressed)

    Appendix 15 File sizes (uncompressed) The actual file size will depend on the image plate format and the pixel size. File sizes stated are approx- imate and have been rounded upwards. Suitable compression methods can considerably reduce the file size without loss of data. Max.
  • Page 48 Hersteller/Manufacturer: DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com...