Thank you for choosing NIVELCO instruments.
We are convinced that you will be satisfied with our product.
1. APPLICATION
NIVOPRESS N submersible hydrostatic level transmitters are applicable for the continuous level measurement of clean
liquids or liquids with slight chemical contamination in wells, open reservoirs or tanks. The NC is recommended for level
detection of contaminated water. NIVOPRESS N is easy to install in tanks and deep wells, and especially recommended
for controlling submersible pumps. The use of supplemental accessories is recommended. Using a sewage adapter direct
contact between the sewage and the diaphragm of the built-in pressure sensor can be avoided. 2-wire types are available
with built-in 4-wire Pt100 temperature sensor or separate 2-wire temperature transmitter. 2-wire types have HART
communication interface. The Ex-types are available for explosion hazardous environments.
2. TECHNICAL DATA
SUBMERSIBLE PROBE
TYPE
NP, NF, NZ, NR
0...200 mH
Measurement range
(1)
Overload allowed
(as a function of range)
Output
4...20 mA + HART
Power supply
Load
(U
= power supply.;
t
U
= min. power supply 12 V)
min
HART
secondary value,
®
Built-in temperature sensor
accuracy: ±3°C (±5.4°F)
Thermal compensation error
(2)
Linearity error (level transmitter)
Temperature error
Operating temperature
(3)
Process connection
Ingress protection
Electrical protection
Electrical connection
Wire cross section
Cable length
NP, NF: Ø22x173 mm
(Ø0.87 x 6.8")
Dimensions
NZ, NR: Ø38x174 mm
(Ø1.5 x 6.85")
NP, NF:
probe: 0.2 kg (0.44 lb)
NZ, NR:
Mass
probe: 0.3kg (0.66 lb)
Additional temperature transmitter
ND types
Additional temperature transmitter
NP types
1.4404 or
Sensor
(1.4571 and 1.4435)
Material of
Housing
wetted
Cable coating
parts
Sealing
Protecting cap
m H
O means: 1 metre of water column,1 mH
O ~ 0.1 bar
(1)
2
2
valid if temperature compensation is active, can be switched in parameter HART - P10b
(2)
special order max. +75 °C (+167 °F)
(3)
A
D
E
C
DDITIONAL
ATA FOR
X
ERTIFIED
Type
Power supply
Ex marking
Reference document
Temperature range
Intrinsically safe data
A
CCESSORIES
Cable terminal box
Dimensions
Ingress protection
Operating temperature
Material
Cable gland
Electrical connection
Cable terminal box with
overvoltage protection*
Data
Electrical data
Cable mounting wedge clamp
Max. mech. load
Operating temperature
2-wire
NK, NN,
NC, NT
ND, NH
O
0...20 mH
O
2
2
As per order code
20x ( ≤3 mH
O)
2
3x
10x ( >3 mH
O)
2
4...20 mA
®
12...30 V DC
U
U
t
min
R
max
0,02
A
HART
®
–
±0.2%
Not applicable
±0.25%
≤ ±0.1% / 10 K
−30...+60 °C (–22...+140 °F)
NAA-209 cable mounting wedge clamp, threaded types with ¾" BSP thread
IP68
Class III
Shielded cable with breathing capillary Ø7 mm (Ø0.275 inch)
0.34 mm
(AWG 22)
2
0...300 m (0...985 ft) as per order code; (Ex ia IIC: max. 100 m [328 ft])
NK, NN: Ø22x173 mm
Ø40x146 mm
(Ø0.87 x 6.8")
(Ø1.55 x 5.75")
ND, NH: Ø38x174 mm
(Ø1.5 x 6.85")
Probe:
0.4 kg (0.88 lb)
Cable: PUR cable: 0.065 kg/m, FEP cable: 0.072 kg/m
Output: 4...20 mA; Power supply: 14...30 V DC;
Temperature range: 0...+60 ºC (+32...+140 ºF);
Accuracy: ±3 °C (±5.4 °F)
Pt100B sensor, 4-wires
Al
O
ceramic
2
3
1.4571
Polyurethane / FEP
Viton
(FKM)
®
1.4571
–
M
– ATEX A
ODELS
PPROVAL
Up to 100 m (328 ft) cable length:
for IIC gas group: C
≤ 52 nF, L
i
NAA–101
93 x 93 x 55 mm (3.66 x 3.66 x 2.16 inch)
IP65
−40...+70 °C (−40...+158 °F)
Polystyrol
M20x1.5 (cable outer diameter 5...10 mm
[0.2...0.4 inch])
Terminal block for cable with max. cross section
of 2.5 mm
(AWG 13)
2
NAA–102
See: NAA-101
See: OVP
NAA–209
300 m (985 ft) cable
−20...+60 °C (−4...+140 °F)
3-wire
NB, NG
NH
0...200 mH
O
0...200 mH
2
2
3x ( ≤20 mH
O)
2
3x
2x ( >20 mH
O)
2
4...20 mA + HART
0...+10 V (0 V ≤ 80 mV)
®
18...30 V DC / 6 mA
≥ 5 kΩ
secondary value, accuracy:
–
±3°C (±5.4°F)
±0.2%
Not applicable
±0.45% ( ≤20 m H
O)
2
±0.25%
±0.25% ( >20 m H
O)
2
≤ ± 0.1% / 10 K
≤ ±0.2% / 10 K
NPH: Ø22 x 173 mm
Ø24 x 212 mm
(NPH: Ø0.87 x 6.8")
(Ø1 x 8.3")
NZH: Ø38 x 174 mm
(NZH: Ø1.5 x 6.85")
Probe: 0.2 kg (0.44 lb)
-
-
1.4404 or
(1.4571 and 1.4435)
POM-C
1.4571
1.4571
POM-C
1.4571
N
.: BKI16ATEX0009
O
N–5– Ex
14...30 V DC
II 1G Ex ia IIC T6 Ga, between 100 m (328 ft) and 300 m (985 ft) cable length:
npk4110m0600h_04
−30...+60 °C (−22...+140 °F)
U
= 30 V, I
= 100 mA, P
i
i
≤ 1.4 mH (calculated with 100 m [328 ft] integrated cable), for IIB gas group: C
i
Overvoltage protection
Mounting
Dimensions
Ingress protection
Breakdown voltage
Absorbed energy
Internal resistance
Leakage current
*only for 2-wire 4...20 mA equipments!
NAZ-101-0
Adapters
NAZ-102-0
NAZ-105-0
NAZ-106-0
®
O
2.1 A
CCESSORIES
User's Manual
Warranty Card
EU declaration of Conformity
= 0.8 W
i
OVP–12/33*, OVP–22/33*
outdoor
72 x 42 x 19 mm
(2.8 x 1.65 x 0.75 inch)
IP54
33 V
600 W / 1 ms
13 Ω
10 A
Threaded spacer
¾" BSP – ½" BSP
¾" BSP – M20x1.5
¾" BSP – 1" NPT
¾" BSP – 1" BSP
II 1G Ex ia IIB T6 Ga
≤ 132 nF, L
≤ 1.6 mH
i
i
OVP–32/33*
EN 60715 - 35 mm (1⅜ inch) rail
62 x 65 x 18 mm
(2.44 x 2.56 x 0.7 inch)
IP20
Material
KO35 (1.4571)
npk411en2006h
1/4