Page 1
® settings, then the Levitronix Service Software version V2.0.5.0 (with according manual Levitronix Doc.# PL-4047-00) is needed. Familiarize yourself with the contents of the manual to ensure the safe and effective use of this product. After reading this manual, please store the manual where the personnel responsible for operating the flow control system can readily refer to it at any time.
® User Manual for PuraLev iF30SU www.levitronix.com Table of Contents SAFETY PRECAUTIONS ................................3 SPECIFICATIONS ..................................4 Specification of Components ..............................4 System Overview and General Specification ......................... 6 General Environmental Conditions ............................8 Pressure-Flow Curves ................................8 Maximum Static Pressure for Pump Heads ........................... 9 Flow Control Specifications ..............................
Installation shall be done by qualified personnel only. Following safety precautions and all “CAUTION”, “WARNING” and “DANGER” indications in the relevant sections shall be followed. CAUTION ® Do not under any circumstances open the driver. Levitronix does not assume responsibility for any damage, which occurs under such circumstances. CAUTION High magnetic field strength of pump impeller.
® User Manual for PuraLev iF30SU www.levitronix.com 2 Specifications 2.1 Specification of Components Figure 1: Flow control system with standard main components Figure 2: General standard accessories Figure 3: Standard cables PL-4078-00, Rev00, DCO# 20-228...
Page 5
Note 2: Gamma irradiated and sterile fittings version available (see product literature for PuraLev-i30SU). Pos. Component Article Name Article # Fitting Wet Material Note LFS-03SU-Z-F1 100-30423 See Levitronix ® technical brochure of Polypropylene (FDA, USP Class VI, ® LFS-03SU-Z-F1-SC1 100-30470 LEVIFLOW Single- LFS-SU single-use sensor series for Triclamp 3/8”...
® User Manual for PuraLev iF30SU www.levitronix.com 2.2 System Overview and General Specification 2.2.1 System Configuration for Stand-Alone Model Figure 4 illustrates the system configuration of the Stand-Alone model with integrated user panel and buttons to set the flow manually. The driver also contains a PLC interface for remote control by analog/digital signals.
Page 7
A RS485 (Modbus) enables communication with a user panel or a ® PC with the Levitronix Service Software. Hence parameterization, firmware updates and failure analysis are possible. The RS485 can also be used as a fieldbus to implement more intelligent concepts of pump control.
® User Manual for PuraLev iF30SU www.levitronix.com 2.3 General Environmental Conditions Usage Indoor Altitude Up to 2000 m Operating ambient temperature 0 to 40 °C -20 to 80 °C Storage ambient temperature -20 to 70 °C for single-use pump head (Extremes for Transportation) 15 –...
® User Manual for PuraLev iF30SU www.levitronix.com 2.5 Maximum Static Pressure for Pump Heads [°F] Safety Factor = 1.3 (according to EN12162) Note: Burst pressure is significantly higher. Ambient Temp = 25 C Note: Pressure limit only valid for pumphead mounted on motor. Values only valid after Gamma radiation doses of <...
® User Manual for PuraLev iF30SU www.levitronix.com 2.7 Basic Dimensions of Main Components Pump Head Inlet and Outlet Fitting: Barb 3/8” for typical tubing ID = 1/4” Figure 11: Basic dimensions (in mm and [inch]) of Stand-Alone model Note 1: Drawings with DCP-30.1 (Triclamp) and DCP-30.2-SF1 (sterile fittings) are available on request.
Page 11
® User Manual for PuraLev iF30SU www.levitronix.com Multi-purpose screw holes 6 pcs M3 x 4 mm Multi-purpose screw holes 2 pcs M3 x 4 mm screws with FKM O-ring for protection Figure 13: Basic dimensions (in mm and [inch]) of OEM model Note 1: Drawings with DCP-30.1 (Triclamp) and DCP-30.2-SF1 (sterile fittings) are available on request...
Page 12
® User Manual for PuraLev iF30SU www.levitronix.com IPS Cable Power ICP-1.1 (for EasyConnect and Stand-Alone drivers) IPS Cable Signal ICS-1.1 (for EasyConnect drivers) Wire size: 0.08 mm Wire size: 0.5 mm to driver fieldbus IN to driver fieldbus OUT to AC/DC Supply...
(see explanation in Section 3.3). For the EasyConnect and OEM models the warning signal ® can be monitored with the Levitronix Service Software or configured on one of the digital outputs. For the Stand-Alone model the temperature can be monitored on the display.
® User Manual for PuraLev iF30SU www.levitronix.com 3.3 Thermal Management The driver temperature depends on the ambient and liquid temperature, as well as on the hydraulic operation point. Figure 17 and Figure 18 illustrate the temperature characteristics of the motor depending on these parameters.
Page 15
® User Manual for PuraLev iF30SU www.levitronix.com The above curves are measurements of the motor temperature at certain liquid and ambient temperatures. Equation (Eq. 1) shows how to calculate the motor temperature for other liquid and ambient temperatures based on these curves.
9. At higher liquid temperature rules mentioned above become more important due to higher cavitation tendency of the liquid. ® Contact the Levitronix Technical Service department (see Section 8) for more detailed considerations and support on the design of the hydraulic circuit.
® User Manual for PuraLev iF30SU www.levitronix.com 4 Installation 4.1 Electrical Installation of Stand-Alone Model 4.1.1 Overview of Connections and Designations Flow Sensor Connector User Button Display Power Supply Connector PLC 6 Connector Figure 19: Electrical connections to Stand-Alone driver IPD-30.5...
Page 18
® User Manual for PuraLev iF30SU www.levitronix.com 4.1.2 Overview Electrical Schematics of Driver Interface Levitronix Driver Example of User Interface Circuit Interface Circuit Pin 2: Digital Input 2.2 k 5..24 V Pin 4: Common Digital Input Pin 6: Analog Input 450 ...
Page 19
Consult Figure 16 to get the power consumption depending ® on the flow. If power supplies are used other than the one defined by Levitronix , it is highly recommended to test these under dynamic conditions (acceleration and braking of the pump rotor speed).
Not connected pink-blue For remote control with other devices. RS485+ white RS485+ Fieldbus Modbus RTU protocol. Usage of Levitronix Service Software with a RS485- RS485- yellow RS485 to USB adaptor cable (see Table 4). Internal Internal grey Do not Internal bus needed to connect pumps for serial pumping.
Page 21
(Current Input) Speed limit = 13000 rpm Speed) The designation of the analog inputs can be Cutt-off speed = 0 rpm ® changed with Levitronix Service Software. Analog voltage input: 0 – 10V (7.9 kOhm, no galvanic isolation). grey- Configurable as Analog Input Note: Max.
Page 22
® User Manual for PuraLev iF30SU www.levitronix.com 4.2.2 Overview Electrical Schematics of Driver Interface Levitronix Driver Interface Circuit Example of User Interface Circuit Pin 5: Digital Input 1 2.2 k Pin 4: Digital Input 2 5..24 V 2.2 k Pin 3: Common Digital Input Pin 7: Analog Input 1 450 ...
Page 23
Consult Figure 16 to get the power consumption depending ® on the flow. If other power supplies are used than the one recommended by Levitronix , it is highly recommended to test these under dynamic conditions (acceleration and braking of the pump rotor speed).
Page 24
Technical Service department (see Section 8) for details and support. ® ➔ Address setting of the pump units can be done with the Levitronix Service Software and a PC. A USB/RS485 converter cable with integrated initialization network can be ordered according to Table 4 ➔...
Table 15: Signals of the driver cable with designation for standard firmware Note 1: For other configurations of PLC Inputs and Outputs refer to relevant firmware documentation ® Note 2: Configurations can be done with Levitronix Service Software Note 3: Power wires (P+, P-) have cross section 1.5 mm and all others 0.14 mm...
Page 26
® User Manual for PuraLev iF30SU www.levitronix.com 4.3.2 Overview Electrical Schematics of OEM Driver Interface Cable Wires (UL Cable) Example of User Levitronix Driver Colours/Designation Interface Circuit Interface Circuit Pink: Digital Input 1 2.2 k 5..24 V Grey: Digital Input 2 2.2 k...
Page 27
Consult Figure 16 to get the power consumption depending on the flow. If other power supplies are used, than the one recommended by Levitronix , it is highly recommended to test these under dynamic conditions (acceleration and braking of the pump rotor speed).
Technical Service department (see Section 8) for details and support. ➔ ® Address setting of the pump units can be done with the Levitronix Service Software and a PC. A USB/RS485 converter cable with integrated initialization network can be ordered according to Table 4.
® User Manual for PuraLev iF30SU www.levitronix.com 4.5 Mechanical Installation – General Instructions WARNING The driver is not rated to be placed next to flammable gases. Do not use the driver next to flammable gases without adequate safety precautions to fulfill the relevant regulatory requirements.
1. For mounting and exchanging the sensors and assure proper operation it is recommended to use the LMK mounting kits, ® which can be ordered according to Table 4. Contact Levitronix for detailed drawing specifications. 2. Alternatively, the sensor body can be mounted with the fixation holes as shown in Figure 14. Assure that the fixation of the flow sensor and that the weight of the inlet and outlet circuit on the fittings are not causing excessive forces resulting in bending the flow measurement path, which might influence the flow measurement.
® User Manual for PuraLev iF30SU www.levitronix.com 5 Operation In this section the very basic operation of OEM, EasyConnect and Stand-Alone models is explained. For more extended operation modes and configurations consult the relevant Firmware and Service Software specifications. 5.1 Operation of Stand-Alone Model 5.1.1 Manual Operation...
® User Manual for PuraLev iF30SU www.levitronix.com 5.2 Operation of EasyConnect and OEM Model 5.2.1 Operation with PLC Figure 30 and Figure 31 show the state diagram of the integrated pump and the integrated flowmeter. Power On Status pump: not...
Page 33
® User Manual for PuraLev iF30SU www.levitronix.com Power On No valid Zero Adjustment data Valid Zero Adjustment data Normal operation Zero Adjustment state Error Status flow sensor: Status flow sensor: Slow blinking (2Hz) active Zero Adjustment was successful Measurement Error:...
Technical Department (see Section 8) for detailed protocol specifications. The ® RS485 interface can be used in connection with the USB interface of a PC and the Levitronix Service Software. Various accessory cables are available to connect to the RS485 of the drivers as illustrated in Figure 5 for the EasyConnect model and Figure 7 for the OEM model.
Inspection Prior to Use The pump head should be inspected prior to use for CAUTION any damage. Do not use the pump head if any damage is found. Contact Levitronix ® regarding return of any suspected pump head. Gamma Sterilization The pump heads DCP-30 have been tested to be 6.1.3...
® User Manual for PuraLev iF30SU www.levitronix.com 6.4 Removal of Pump Head 6.3 Mounting of Pump Head 6.4.1 Preparation 6.3.1 Preparation Set the speed to 0 rpm and disable the system. After After removal of the pump head from its packaging...
Table 4 can be used to setup communication. The software can be used for performing detailed troubleshooting. For usage of the Service Software refer to the Service Software User Manual, which is available in the download section on the Levitronix web-page or ...
® User Manual for PuraLev iF30SU www.levitronix.com 9 Appendix 9.1 Regulatory Status 9.1.1 CE Marking ® We herewith declare that the Flow Control System Family PuraLev iF30SU, in its various configurations, is in conformity with the below mentioned European Directives.
® User Manual for PuraLev iF30SU www.levitronix.com 9.2 Symbols and Signal Words Symbol / Description Type Source Signal Word Indication of an imminently hazardous situation that, if not DANGER Signal word SEMI S1-0701 avoided, will result in death or severe injury. Limited to the...