Rossmax Pencil Thermometer TG100
Quick Guide
How to Use
Push the On/Off bu�on by the display window. The display will read for 2 seconds and beep sound will be
heard.
1. Releasing the on/off bu�on, last measured temperature will be shown for about 2 seconds. Then shows
the self-test temperature 37.0°C (98.6°F) .And then Lo°C (Lo°F) displays.
2. Wipe or dip the probe with alcohol for disinfec�on. Posi�on the probe in oral, rectal, axillary etc. For oral
use, please put the sensor into the posi�on marked by "√" according to the picture. To insure the accuracy,
please keep the mouth dosed for 5 minutes before tes�ng.
3. Once the degree sign °C (°F) on the display has stopped flashing (usually within 30 to 60 seconds), an alarm
signal will sound for approx. 10 seconds, the measured temperature is shown on LCD window. The
temperature reading will not change a�er the unit is removed from tes�ng posi�on.
If the temperature <37.8°C (100.0°F), the buzzer will alarm "Bi---Bi---Bi---Bi---".
If the temperature ≥ 37.8°C (100.0°F), the buzzer whose frequency is higher will alarm "Bi-Bi-Bi----Bi-
Bi-Bi--" about 10 �mes rather than "Bi---Bi---Bi---Bi---"
4. The unit will automa�cally turn off in 10 minutes for energy-saving. However, to prolong the life of the
ba�ery, it is suggested to press the On/Off bu�on to turn the unit off once the measurement is done.
5. Please dispose of any used ba�ery and the thermometer according to the law.
Note: A�er step 2, if the room temperature is greater than 32.0°C (89.6°F ), it will be displayed the room
temperature.
Note: Usage of the probe cover may result in a 0.1°C (0.2°F) difference from actual temperature.
Note: To clean the probe before and a�er using the thermometer to ensure an accurate reading and avoid
cross contamina�on approximately.
For a�er sales service, please contact Cypress Medic Sdn Bhd (www.cypressmedic.com.my)
Need help?
Do you have a question about the TG100 and is the answer not in the manual?
Questions and answers