Title Page MMMOPERATING INSTRUCTIONS MMMI O P E R A T I N G I N S T R U C T I O N S VISIC100SF Gas Adjustment Kit...
Page 2
Original document This document is an original document of SICK AG. O P E R A T I N G I N S T R U C T I O N S | VISIC100SF 8027092/AE00/V1-0/2022-10 | SICK Subject to change without notice...
4.3.12 Training ..................12 4.3.13 Disposal..................12 Operation ....................13 Operating concept..................13 8027092/AE00/V1-0/2022-10 | SICK O P E R A T I N G I N S T R U C T I O N S | VISIC100SF Subject to change without notice...
Page 4
Final steps ..................27 Technical data..................28 Test gases ..................... 28 O P E R A T I N G I N S T R U C T I O N S | VISIC100SF 8027092/AE00/V1-0/2022-10 | SICK Subject to change without notice...
Risk or hazardous situation which could result in severe personal injury or death 8027092/AE00/V1-0/2022-10| SICK O P E R A T I N G I N S T R U C T I O N S | VISIC100SF Subject to change without notice...
Fig. 1: Type plate, Gas Adjustment Kit Designation Part number Serial number 8027092/AE00/V1-0/2022-10| SICK O P E R A T I N G I N S T R U C T I O N S | VISIC100SF Subject to change without notice...
It is assumed that the test gases are provided by the customer. Calibration gas data, see “Test gases”, page O P E R A T I N G I N S T R U C T I O N S | VISIC100SF 8027092/AE00/V1-0/2022-10| SICK Subject to change without notice...
No special ventilation is required when the limit values (see 4.3.1) are observed. 8027092/AE00/V1-0/2022-10| SICK O P E R A T I N G I N S T R U C T I O N S | VISIC100SF Subject to change without notice...
Test gas cylinders that are not emptied must be returned to the supplier and the supplier is responsible for their disposal. O P E R A T I N G I N S T R U C T I O N S | VISIC100SF 8027092/AE00/V1-0/2022-10| SICK...
The status LEDs indicate the operating status. Green: Ready for operation ● Yellow: Malfunction ● 8027092/AE00/V1-0/2022-10| SICK O P E R A T I N G I N S T R U C T I O N S | VISIC100SF Subject to change without notice...
The number following is always 1 because the adjustment tool can only address one sensor. O P E R A T I N G I N S T R U C T I O N S | VISIC100SF 8027092/AE00/V1-0/2022-10| SICK...
Carbon monoxide 0–300/0–200 1129 Nitrogen monoxide 0–100 1130 Nitrogen dioxide 0–5 8027092/AE00/V1-0/2022-10| SICK O P E R A T I N G I N S T R U C T I O N S | VISIC100SF Subject to change without notice...
– Backlight is yellow. – All points are displayed on the LED. O P E R A T I N G I N S T R U C T I O N S | VISIC100SF 8027092/AE00/V1-0/2022-10| SICK Subject to change without notice...
The factory-set power-on time is 5 min. and can only be read out. 8027092/AE00/V1-0/2022-10| SICK O P E R A T I N G I N S T R U C T I O N S | VISIC100SF Subject to change without notice...
>>> Read-in successful: First subordinate menu item is displayed. 2 Press “Down arrow”to display the following menu items. O P E R A T I N G I N S T R U C T I O N S | VISIC100SF 8027092/AE00/V1-0/2022-10| SICK...
Exchange gas sensor. Procedure see “Perform adjustment”, page 23 8027092/AE00/V1-0/2022-10| SICK O P E R A T I N G I N S T R U C T I O N S | VISIC100SF Subject to change without notice...
Internal error that cannot be corrected. Exchange gas sensor. Procedure see “Perform adjustment”, page 23 O P E R A T I N G I N S T R U C T I O N S | VISIC100SF 8027092/AE00/V1-0/2022-10| SICK Subject to change without notice...
Allen key, SW4 ● Special pliers/universal tool for cable glands ● 8027092/AE00/V1-0/2022-10| SICK O P E R A T I N G I N S T R U C T I O N S | VISIC100SF Subject to change without notice...
2 Set the VISIC100SF to Maintenance mode. 3 Observe the running-in and warm-up time for the gas sensors. O P E R A T I N G I N S T R U C T I O N S | VISIC100SF 8027092/AE00/V1-0/2022-10| SICK...
NOTICE: Only use undamaged sealing rings on the tapping valves. 8027092/AE00/V1-0/2022-10| SICK O P E R A T I N G I N S T R U C T I O N S | VISIC100SF Subject to change without notice...
5 End the input with “Confirm”. 6 Press “ESC”. >>> Value is saved. O P E R A T I N G I N S T R U C T I O N S | VISIC100SF 8027092/AE00/V1-0/2022-10| SICK Subject to change without notice...
14 Remove the hose adapter. Possible malfunction messages see “Adjust zero point”, page 19 8027092/AE00/V1-0/2022-10| SICK O P E R A T I N G I N S T R U C T I O N S | VISIC100SF Subject to change without notice...
17 Remove the hose adapter. Possible malfunction messages see “Adjust sensitivity”, page 20 O P E R A T I N G I N S T R U C T I O N S | VISIC100SF 8027092/AE00/V1-0/2022-10| SICK Subject to change without notice...
4 Replace the housing cover and tighten it with the two screws. 8027092/AE00/V1-0/2022-10| SICK O P E R A T I N G I N S T R U C T I O N S | VISIC100SF Subject to change without notice...
The flow volumes for CO, NO and NO are set fixed using the valves. O P E R A T I N G I N S T R U C T I O N S | VISIC100SF 8027092/AE00/V1-0/2022-10| SICK Subject to change without notice...
Page 29
5/8" 5/8" Valve outlet 18 UNF 18 UNF 18 UNF 8027092/AE00/V1-0/2022-10| SICK O P E R A T I N G I N S T R U C T I O N S | VISIC100SF Subject to change without notice...
Page 30
Phone +7 495 283 09 90 E-Mail info@sick.com E-Mail office@sick.com.gr E-Mail info@sick.ru Vietnam Hong Kong Singapore Phone +65 6744 3732 Phone +852 2153 6300 Phone +65 6744 3732 E-Mail sales.gsg@sick.com E-Mail ghk@sick.com.hk E-Mail sales.gsg@sick.com Detailed addresses and further locations at www.sick.com...