ENBIO S User Manual
Hide thumbs Also See for S:

Advertisement

Quick Links

User manual
ver. 21.01.2022

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the S and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ENBIO S

  • Page 1 User manual ver. 21.01.2022...
  • Page 2: Table Of Contents

    5.4 Information menu ........................ 44 FRONT 5.4.1 Counters ..........................45 5.5 Restarting ..........................45 6. Maintenance and care ................47 The latest version of the manual is available at LOCKING KNOB 6.1 Replacement parts....................... 48 www.enbio.com 6.2 Periodic inspections ......................48...
  • Page 3: Introduction

    The Enbio S has not been designed to sterilize liquid loads, bio-medical This symbol is located on the device’s rating plate and - Portable RF communications equipment (including peripherals such as antenna –...
  • Page 4: Scope Of Delivery And Unpacking Of The Device

    Remove it from the bag g. If using wastewater containers, we recommend using containers of the for use in ENBIO S sterilizers. Therefore tap water cannot be used as feed min. When moved from a cold room to a warm one, the device may placed in the carton and tighten it yourself in a specially designated same volume as those used for the deionised water.
  • Page 5: Tool Preparation And Loading

    • For non-packaged tools – place the tools on the tray in such a manner • Do not exceed a maximum weight of 500 g (1.10 lb) for ENBIO S. so they do not contact each other directly. This will accelerate the drying •...
  • Page 6 If you decide to continue working or the . field has been ceed a maximum weight of 500 g (1.10 lb) for ENBIO S. selected, the screen will display a chart of pressure during the entire pro- By pressing the...
  • Page 7: Test Programs

    The vacuum leak test enables testing the autoclave for the This test assesses the device’s performance in sterilizing charges composed field on the display. presence of leaks. The following are checked during the test:...
  • Page 8: Information Menu

    The ENBIO S sterilizer counts the number of performed processes and uses inspection is recommended, the val- it to notify you about the recommended dates of replacing elements sub- ues will be highlighted in yellow.
  • Page 9 6. MAINTENANCE AND CARE If the field is selected, the following messages will be alter- However, in order to do so, you must enter the 4-digit security code 0000. To clean the tray well it must be removed from the front of the device. To nately displayed, notifying you that the process has been aborted by the do this, lift the tray gently up and pull it away from the front.
  • Page 10: Replacement Parts

    Minimum RAM – at least 512 MB Screen resolution – at least 1200 x 720 or more In order to ensure correct functioning of the ENBIO S sterilizer, it is recom- Using up-to-date antivirus software – strongly recommended. mended to perform periodic service inspections and replace parts subject...
  • Page 11: Software Installation

    8.1 Software installation Next, the information about placing the software shortcut on the comput- After making your selection, click „Next”. By clicking the Install button you After installation, the following message is displayed. er desktop will be displayed. will install the EnbioDataViewer software. To install the software, double click on the software installation file.
  • Page 12: Program Construction And Main Functionalities

    • Closing the program EnbioDataViewer updating procedure 1. Run Enbio Data Viewer software. 2. At main screen of Enbio Data Viewer click “Help” button and then “About the program”. 3. At new window choose „Check the current software version”. 4. The Enbio website will open automatically, from which you should download the latest version of the EnbioDataViewer program –...
  • Page 13: Warning Messages And Error Codes

    The program allows you to generate a report from every process • Range of dates performed by the autoclave. It contains all necessary process data and If any irregularities in the device’s operation occur, the screen will display • About program • Sterilization number the result of sterilization.
  • Page 14: Error Codes

    9.3 Error codes 10. WARRANTY CLAIM HANDLING “Power failure” Temporary power loss Confirm error. during operation “Pressure during Pressure exceeded Confirm error. The following table includes the error codes that may be displayed during In order to report a problem with the device, call to +1 878 670 1641. If standby”...
  • Page 15: Warranty Terms And Conditions

    11. WARRANTY TERMS AND CONDITIONS Rating plate located on the bottom of the device. ENBIO sterilizers are covered by a standard 12-month warranty. Detailed warranty terms and conditions are available from the supplier of this device. 12. TECHNICAL INFORMATION ENBIO S...
  • Page 17 Manufacturer: Enbio Group AG Eichengasse 3, 4702 Oensingen, Switzerland www.enbio.com...