Page 1
Operating Manual TM 01 Temperature Measuring Device BINDER GmbH Address Post office box 102 D-78502 Tuttlingen Tel. +49 7462 2005 0 +49 7462 2005 100 Internet http://www.binder-world.com E-mail info@binder-world.com Service Hotline +49 7462 2005 555 Service Fax +49 7462 2005 93 555 Service E-Mail service@binder-world.com...
Page 2
électromagnétique et abrogeant le directive 98/336/CEE. Das oben beschriebene Produkt trägt entsprechend die Kennzeichnung CE. The product described above, corresponding to this, bears the CE-mark. Le produit décrit ci-dessus, en correspondance, porte l’indication CE. 1 / 2 TM 01 01/2009 page 2/27...
Page 3
D-78532 Tuttlingen, 12.01.2009 BINDER GmbH P. M. Binder V. Siegle Geschäftsführender Gesellschafter Leiter F & E Managing Director Head of R & D Directeur général Chef de service R&D 2 / 2 TM 01 01/2009 page 3/27...
Connection of TM 01 to the APT-COM™ 3 Network ................13 7.1.1 Connection of TM 01 via the W&T interface converter RS 232 / RS 422 .......14 7.1.2 Connection of TM 01 via the W&T interface converter Ethernet / RS 422 ......14 7.1.3...
Dear customer, For the correct operation of the temperature measuring devices TM 01, it is imperative that you read this operating manual completely and carefully and observe the given instructions. Safety This operating manual is part of the scope of delivery. Always keep it at hand.
Instruction how to avoid the hazard: mandatory action Observe all other notes and information not necessarily emphasized in the same way, in order to avoid disruptions which could result in direct or indirect injury or property damage. TM 01 01/2009 page 6/27...
5405194 / 5601143 / 5773287 / 6079403 D 78532 Tuttlingen / Germany TM 01 Serial No. 00-00000 Tel. + 49 (0) 7462/ 2005-0 Made in Germany Internet: www.binder-world.com Figure 1: Type plate TM 01 Indications of the type plate Information Nominal temperature 60°C Nominal temperature 140°F...
General safety instructions on installing and operating the temperature measuring device TM 01 With regard to operating the temperature measuring device TM 01 and to its installation location, please observe regulations BGR 120 issued by the German professional association for the chemical industry (formerly ZH 1/119 laboratory guidelines issued by the employers’...
For monitoring, the nominal value set in the temperature chamber can be entered into the temperature measuring device directly at the keys of the TM 01 or via APT-COM™ 3. If these nominal values exceed or fall below the permanently set tolerances, this is reported by APT-COM™ 3 via its alarm system.
Equipment overview Figure 3: Front view TM 01 Flexible Pt 100 temperature sensor, 2.8 m LED display for the temperature set-point of the supervised device (only of relevance while using the alarm functions of APT-COM™ 3) LED display for the measured temperature Keyboard to enter the device address (for connection with APT-COM™)
Wait at least one hour until the measuring device has attained ambient temperature and is completely dry. Location of installation and ambient conditions Set up the temperature measuring device TM 01 on an even surface, free from vibration and in a well- ventilated, dry location. CAUTION Danger of overheating.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the unit. Installation of the equipment Electrical connection The temperature measuring device TM 01 has a fixed power connection cable of 1800 mm in length (G) and a shockproof plug. • Power supply fuse: 10 Amp •...
The measured values can be transmitted via the RS 422 interface (I) online to a PC with the APT-COM™ communication software (option) from BINDER. In addition, APT-COM™ enables convenient data storage and alarming through warning messages on the PC as well as remote alarming via telephone line.
(5023-0115) COM1 or COM2 Figure 5: Connection of TM 01 via the W&T interface converter RS 232 / RS 422 For instruction to configure this type of connection, see installation manual 7001-0101 for the W&T interface converter RS 232 / RS 422. It is also included with the manuals on your APT-COM™...
7.1.3 Connection of TM 01 and several BINDER chambers with interface RS422 S eria l i nte rface RS 232 (COM Port) Measuri ng Ca ble 5m o f measur ing computer compu te r ( 5023 -009 5) A PT-COM™ 3 Up to 4 p lug distribu to r u nits p lus evt.
The addresses No. 1 to 30 are available. In case the Alarm Box AB01 is connected, the address no. 30 is assigned to the Alarm Box. 7.2.1 Setting the address at the TM 01 TM 01 functional keys Address setting at the TM01:...
COM Port in window “Add New Chamber“ you can enter the address that had been set at the TM 01 under “RS422 MOD Bus Address“. Figure 8: Address entry in APT-COM™ 3 Following confirmation with “Apply“, the TM 01 is now displayed with the selected settings in the...
Setting the temperature set-point and the tolerance limits in APT-COM™ 3 First you need to create and start a measurement for TM 01 in APT-COM™. Following measurement start the measurement window for the TM 01 measurement opens up. Figure 10: Setting the temperature set-point in the measurement window of APT-COM™ 3 Figure 11: Setting the tolerance limits in the measurement window of APT-COM™...
Never try to open the device in order to repair it. In case of problems please contact BINDER service at one of the indicated service addresses (chap. 8.3). CAUTION Danger of corrosion. Damage to the unit. ∅ Do NOT use acidic or chlorine cleaning detergents.
Sending back the unit to BINDER GmbH If you send a BINDER product to us for repair or any other reason, we will only accept the product upon presentation of an authorization number that has previously been issued to you. We will issue an authorization number after receiving your complaint either in writing or by telephone prior to your sending the BINDER product back to us.
(Elektro- und Elektronikgerätegesetz, ElektroG) from 23 March 2005, BGBl. I p. 762 or contact BINDER service who will organize taking back and disposal of the unit according to the German national law for electrical and electronic equipment (Elektro- und Elektronikgerätegesetz, ElektroG) from 23 March 2005, BGBl.
(category 9) only intended for professional use“. They must not be disposed of at public collecting points. The temperature measuring device TM 01 bears the symbol for the marking of electrical and electronic equipment manufactured / placed on the market in the EC after 13 August 2005 and be disposed of in separate collection according to the directive 2002/96/EC on waste electrical and electronic equipment (WEEE).
For final decommissioning and disposal of the measuring device, please contact BINDER service. Observe the statutory regulations for appropriate, environmentally friendly disposal. The main board of the measuring device includes a lithium cell. Please dispose of it according to national regulations. TM 01 01/2009 page 23/27...
(certified since December 1996 by TÜV CERT). All test equipment used is subject to the administration of measurement and test equipment that is also constituent part of the BINDER QM DIN EN ISO 9001systems. They are controlled and calibrated to a DKD-Standard at regular intervals.
Details about utilized substances / biological substances / Einzelheiten über die eingesetzten Substanzen/biologische Materialien: Designations / Bezeichnungen: ____________________________________________________________________________ ____________________________________________________________________________ ____________________________________________________________________________ Safety measures required for handling these substances / Vorsichtsmaßnahmen beim Umgang mit diesen Stoffen: ____________________________________________________________________________ ____________________________________________________________________________ ____________________________________________________________________________ TM 01 01/2009 page 25/27...
Page 26
Kind of transport / transporter / Transportweg/Spediteur: Transport by (means and name of transport company, etc.) Versendung durch (Name Spediteur o.ä.) ___________________________________________________________________________________ Date of dispatch to BINDER GmbH / Tag der Absendung an BINDER GmbH ___________________________________________________________________________________ TM 01 01/2009 page 26/27...
Page 27
We are aware that, in accordance with Article 823 of the German Civil Code (BGB), we are directly liable with regard to third parties, in this instance especially the employees of BINDER GmbH, who have been entrusted with the handling / repair of the unit / component. / Es ist uns bekannt, dass wir gegenüber Dritten –...