INSTRUCTION MANUAL FOR HOT AIR OVEN DSO-D & DSO-DF SERIES

Warnings:

* Make sure that the door is closed well to prevent heat loss and power drain.
* The surface is hot. Do not touch.
* Do not put any objects on the top of the instrument.
* Please keep the environment ventilated.
1. Connect with power supply. Please make sure the power voltage is the same as the voltage shown on the
label stuck on the back of machine.
2. Put your sample in and close the door.
3. Press "ON/OFF Button", then you will see the "Power indicator' becomes green. Wait for about 5 seconds;
you can see that on the display screen, the top row is the present temperature value (PV) and the bottom
row is the setting temperature value (SV).
4. Set temperature: you will see
enter the value.
5. Set time: Select time mode: Continouos mode
5.1 Press
to set Continuous mode. OUT2 indicator on display screen will disappear. The
setting is finished. Go to Step 6.
5.2 Press
to set Timer mode. OUT2 indicator on screen will appear.
5.2.1 You will see
5.2.2 Use
to enter the value. Press
DIGISYSTEM LABORATORY INSTRUMENTS INC.
or
.
or
key to select time unit
to next setting.
Air vent
. Press
or
to adjust SV value and then press
or Timer mode
(hour) or
www.digisystemlab.com
Controller
.
(minute). Press
to
REV.1-2017

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DSO-D Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for DIGISYSTEM DSO-D Series

  • Page 1 5.2 Press to set Timer mode. OUT2 indicator on screen will appear. 5.2.1 You will see 5.2.2 Use key to select time unit (hour) or (minute). Press to enter the value. Press to next setting. DIGISYSTEM LABORATORY INSTRUMENTS INC. www.digisystemlab.com REV.1-2017...
  • Page 2 Keep the machine clean all the time, so no routine maintenance is required. Cleaning can be done with a damp cloth. Avoid the use of solvents as they may damage the product housing and inner chamber. Circuit Diagram: DIGISYSTEM LABORATORY INSTRUMENTS INC. www.digisystemlab.com REV.1-2017...
  • Page 3 Auto-tuning function. located User’s calibration thermometer is different Refer to the other functions, execute the PV value is not correct from the factory’s calibration thermometer function for PV shift (offset) value. DIGISYSTEM LABORATORY INSTRUMENTS INC. www.digisystemlab.com REV.1-2017...

This manual is also suitable for:

Dso-df series

Table of Contents