THIS MANUAL MUST BE CAREFULLY READ BY ALL PERSONS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR USING OR SERVICING THE PRODUCT. Like any piece of equipment, the Flow-Check Meter from MSA will perform as designed only if used and serviced according to the instructions. OTHERWISE,...
Page 2
Introduction and Use INTRODUCTION Note: If an adjustable valve-connector is used, fully close it. The Flow-Check Meter from MSA is used to verify the minimum required air flow rates on the following respira- 2. Turn the air source on. tors from MSA: 3.
Page 3
CONSTANT-FLOW HOODS FROM MSA WHICH USE A MASK MOUNTED POWERED AIR-PURIFYING VALVE-CONNECTOR RESPIRATOR (PAPR) WITH FULL FACEPIECE Testing PAPR with Hood must use Adapter P/N 489999 1. Remove the motor blower module from the mask. 1. Remove the breathing tube from the valve-connector 2.
Page 4
Dirt on the inside of the tube or on the indicating ball must be removed before the meter is used. The meter may be cleaned using ConfidencePlus™ Cleaning Solution (P/N 10009971) from MSA or a mild cleaning solution. Follow the instructions supplied with the cleaning solution.