FujiFilm FUJI DRI-CHEM NX700 Instruction Manual

FujiFilm FUJI DRI-CHEM NX700 Instruction Manual

Clinical chemistry analyzer
Table of Contents

Advertisement

Quick Links

INSTRUCTION MANUAL
FUJI DRI-CHEM
CLINICAL CHEMISTRY ANALYZER
FUJI DRI-CHEM NX700
FUJI DRI-CHEM NX700i
This Manual describes details on how to operate
the FUJI DRI-CHEM NX700/NX700i and cautions
to be observed when operating it. Please read this
manual thoroughly before actually operating the
FUJI DRI-CHEM NX700/NX700i. After reading
this manual, store it nearby the FUJI DRI-CHEM
NX700/NX700i so that you can see it whenever
necessary.
EN
Safe Usage and
Handling Precautions
Part Names and
Consumables
Measurement
Applicable Usage
Methods
Loading Consumables
Maintenance
Quality Control
Troubleshooting
Function Menu
Other Functions
USM
Specifications
Glossary
7th Edition
897N201058J April 2022
1
2
3
4
5
6
7
8
9
10
11
12
13

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FUJI DRI-CHEM NX700 and is the answer not in the manual?

Questions and answers

Summary of Contents for FujiFilm FUJI DRI-CHEM NX700

  • Page 1 Other Functions Specifications Glossary This Manual describes details on how to operate the FUJI DRI-CHEM NX700/NX700i and cautions to be observed when operating it. Please read this manual thoroughly before actually operating the 7th Edition FUJI DRI-CHEM NX700/NX700i. After reading...
  • Page 3 Installation may only be conducted by authorized service personnel. <Intended Purpose> The FUJI DRI-CHEM NX700 and FUJI DRI-CHEM NX700i (henceforth, the FDC NX700) is an in vitro diagnostic medical device to automatically quantitate the concentration or the activity of the components in blood or urine by colorimetric end-point, rate and ISE tests, using the FUJI DRI-CHEM SLIDEs.
  • Page 4 So for the latter, sections that describe the unit can be skipped). Name PF Unit FUJI DRI-CHEM NX700 FUJI DRI-CHEM NX700i NOTE: 4 : The PF Unit is mounted on the device. - : The PF Unit is not mounted on the device.
  • Page 5: Table Of Contents

    1.14 Handling samples ………………………………………………………………………… 1-7 1.15 Sampler cover …………………………………………………………………………… 1-8 1.16 Diluents and reference fluid ……………………………………………………………… 1-8 1.17 FUJIFILM PF plasma filters ……………………………………………………………… 1-8 1.18 FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes, and FUJI DRI-CHEM MIXING CUPS ……………………………………………………………………………………… 1-9 1.19 Light source lamp ………………………………………………………………………… 1-10 1.20 Recording paper …………………………………………………………………………...
  • Page 6 Table of contents Sample tubes for the FDC NX700 ……………………………………………………… 3-31 3.9.1 Blood collecting tubes …………………………………………………………… 3-31 3.9.2 FUJI Sample Tubes …………………………………………………………… 3-33 3.9.3 Sample racks …………………………………………………………………… 3-34 3.10 QC card system …………………………………………………………………………… 3-35 Applicable usage methods CRP calibration …………………………………………………………………………… 4-1 4.1.1 Calibration measurement ………………………………………………………...
  • Page 7 Table of contents Cleaning and replacing the reference fluid cap packing ……………………………… 6-27 PF unit maintenance ……………………………………………………………………… 6-29 6.9.1 Cleaning the PF unit (sensor, reference plate, and suction pad) ………… 6-29 6.9.2 Inspecting the PF Suction Pad ………………………………………………… 6-31 6.9.3 Replacing the PF Suction Pad …………………………………………………...
  • Page 8 Table of contents 9.3.10 Lot compensation coefficients ………………………………………………… 9-25 9.3.11 Standard curve coefficients …………………………………………………… 9-27 9.3.12 Lot qualification ………………………………………………………………… 9-30 9.3.13 PF calibration coefficients ……………………………………………………… 9-31 9.3.14 Dilution setting …………………………………………………………………… 9-33 9.3.15 Tube setting for diluent & reference fluid ……………………………………… 9-36 9.3.16 Print setting for measured result ………………………………………………...
  • Page 9: Safe Usage And Handling Precautions

    This section contains the safety precautions that must be followed for safe operation of the FUJI DRI-CHEM NX700 (For the remainder of the manual, FUJI DRI-CHEM NX700 is shortened to “FDC NX700”). Before using this analyzer, please read this chapter carefully and follow the precautions given, so that you can operate it correctly.
  • Page 10: Biohazards And Disposal

    1 Safe usage and handling precautions Biohazards and disposal WARNING: Items such as measured slides, tips, mixing cups, and sample tubes coated with samples such as blood and urine, and cotton swabs and cloths used to clean the equipment are infectious industrial waste.
  • Page 11: Warnings Regarding Electric Shock

    1 Safe usage and handling precautions Warnings regarding electric shock WARNING: The voltage supplied to the equipment is AC 100 - 240 V. To avoid electrical shock, observe the fol- lowing precautions. • Avoid installation locations in which water may splash on the equipment. •...
  • Page 12: Caution Regarding Moving Parts

    1 Safe usage and handling precautions Caution regarding moving parts WARNING: Do not place your fingers near moving parts (sampler and sample disk) when operating the equip- ment, including during maintenance. Also be sure not to get your fingers, hair, clothing, or accessories caught in moving parts. WARNING: During sample processing, ensure the sampler cover is closed and locked with the sampler cover lock key to prevent injuries and biohazards.
  • Page 13: Action For Malfunction

    1 Safe usage and handling precautions CAUTION: Hold the handle of the device when lifting or moving it. Handle CAUTION: • When moving the device to another installation location, remove all slides, samples, diluent, refer- ence fluid, mixing cups, tips, and other consumables beforehand. •...
  • Page 14: Measurement Results

    1 Safe usage and handling precautions 1.10 Measurement results IMPORTANT: Make a diagnosis in a comprehensive manner, considering other relevant test results or clinical symptoms. IMPORTANT: When an equipment problem (error) appears before or during measurement, or a warning appears in the measurement results, the measurement results may not be accurate.
  • Page 15: Crp Measurement

    1 Safe usage and handling precautions 1.12 CRP measurement WARNING: As calibrator CP is derived from blood, it should be handled in the same way as a patient sample. When handling calibrator CP, always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility.
  • Page 16: Sampler Cover

    FUJIFILM plasma filter PF is shortened to "PF" in this document. IMPORTANT: Before using a FUJIFILM PF, refer to “Instructions for Use” provided with the PF, paying special at- tention to the precautions concerning the samples measured and the blood collecting tubes used.
  • Page 17: Cups

    1 Safe usage and handling precautions 1.18 FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes, and FUJI DRI-CHEM MIXING CUPS NOTE In this manual, FUJI DRI-CHEM AUTO TIPS may be shortened to “tip”, FUJI Sample Tube to “tube”, and FUJI DRI-CHEM MIXING CUPS to “mixing cup”. FDC NX700 performs sample application automatically via the sampler.
  • Page 18: Light Source Lamp

    1 Safe usage and handling precautions IMPORTANT: FUJI DRI-CHEM AUTO TIPS: This product does not allow autoclave treatment before use. IMPORTANT: Please store at room temperature. 1.19 Light source lamp CAUTION: The light source lamp gets very hot. Before replacing the lamp, turn the power off, wait at least 5 minutes, and then check that the lamp has cooled.
  • Page 19: Labels

    1 Safe usage and handling precautions 1.21 Labels The following safety labels are affixed to the FDC NX700. Power button Sampler cover lock Biohazard label key caution label Refer to Section 1.15  Moving parts caution label Biohazard label High temperature PF tube caution Lamp caution label Biohazard label...
  • Page 20: Symbols

    1 Safe usage and handling precautions 1.22 Symbols Sign Description Warning, Caution, Important, consult documents Biological risks High temperature caution Power button Power switch Fragile item, handle carefully Protect from rain This way up Stacking limitation Temperature limitation Manufacturer Serial number Country of manufacture Date of manufacture Authorized representative in the European Community...
  • Page 21 1 Safe usage and handling precautions Sign Description CE mark of conformity Importer Consult an electronic Instructions for Use (e-IFU) Unique device identifier 897N201058J FDC NX700 Instruction Manual 2022.04 1-13...
  • Page 23: Part Names And Consumables

    Part names and consumables Part names NOTE The following is an illustration of the FDC NX700 (full option model). FDC NX700i has slight difference. Printer Sampler cover LCD (touch panel) Power button Sampler cover lock Front bottom cover Card reader Disposal box Sampler nozzle cover Plasma filter unit (PF unit)
  • Page 24 2 Part names and consumables Air filter (Louver) RS-232C connector USB connector LAN terminal Power switch Power inlet Air filter (Louver) 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 25 2 Part names and consumables Disposal box Light source lamp Sample rack (A - E) Reference fluid holder Slide cartridge (A - E) Calibrator holder Diluent holder (D1 - D3) Mixing cup Sample application Tip rack unit cover 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 26 2 Part names and consumables < Details of PF unit > PF reference plate PF unit PF sensor / suction pad PF sample rack PF pump 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 27: Touch Panel Screen Names And Functions

    2 Part names and consumables Touch panel screen names and functions On this equipment, the touch panel is used to perform configurations, run measurements, con- firm measurement results and messages, and perform other operations. LCD (touch panel) 2.2.1 Measurement-preparation screen If the power is switched on, the Measurement-preparation screen appears after startup.
  • Page 28 2 Part names and consumables < Display of the sample disk > 8 Mixing cups Displays the number of remaining mixing cups. If this display is tapped, mixing cups can be loaded. Refer to Section 5.3  9 Diluent If this display is tapped, the diluent can be loaded.
  • Page 29: Help Screen

    2 Part names and consumables 2.2.2 Help screen During measurement, maintenance, consum- able exchange, and other actions on the FDC NX700, you can tap the button on the Meas- urement-preparation screen to check the proce- dures. 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 30 2 Part names and consumables button The Measurement-preparation screen ap- pears. button Return to the previous screen. 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 31: Names And Functions Of Software Keyboard Screens

    2 Part names and consumables Names and functions of software keyboard screens When data entry for operations and configura- tions is necessary, the software keyboard ap- pears on the LCD automatically. Enter alphanu- meric characters, symbols, and katakana. (LCD touch panel) 1 Alphabetical characters, numbers, symbols, and katakana buttons Enter alphabetical characters, numbers,...
  • Page 32 2 Part names and consumables Change the entry screen by using the [ABC], [abc], [123], or [-#/], button. < When the [ABC] button is selected > < When the [abc] button is selected > When the [123] button is selected When the [-#/] button is selected 8 [ABC] button 0 [123] button...
  • Page 33: Bundled Items

    2 Part names and consumables Bundled items Recording paper 2 rolls Slide cartridge Slide weight Slide cartridge label (ABCDE label) Light source lamp O-ring Tip rack Sampler leak check tool PF leak check tool (For FDC NX700) I-SER label ISE slide set caution label European authorized representative label Key (for sampler cover) Instruction manual...
  • Page 34: Consumables And Optional Items

    2 Part names and consumables Consumables and optional items To purchase the consumables or optional items listed below, please contact the dealer from whom you purchased the FDC NX700. 2.5.1 Consumables Name Package/unit FUJI DRI-CHEM AUTO TIPS 6 cases FUJI Sample Tubes FUJI HEPARIN TUBE (1.5 mL) 500×1 box FUJI HEPARIN TUBE (0.5 mL containing)
  • Page 35: Optional Items

    2 Part names and consumables 2.5.2 Optional items Name Package/unit Sample rack For φ16×100 mm For φ13×100 mm For φ13×75 mm For 1.5 mL For 0.5 mL For φ16×100 mm (when using PF) For φ13×100 mm (when using PF) For φ13×75 mm (when using PF) * Slide cartridge * Slide weight Sample barcode reader...
  • Page 36: Measurement

    Measurement Operation overview On the FDC NX700, first load the consumables and other items to prepare for measurement (Step 1), and then perform the actual measurement (Step 2). Step 1 - Measurement preparation Cleaning the disposal box Refer to Section 3.2(1). Checking the recording paper Refer to Section 3.2(2) Starting the FDC NX700...
  • Page 37 3 Measurement Step 2 - Measurement Basic ISE test measurements  Refer to Section 3.5 Refer to Dilution measurements  Refer to Section 3.6 Section 3.3 measurements CRP measurements  Refer to Section 3.7 Slide Sample Checking Checking the sample the tips type Start...
  • Page 38: Measurement Preparation

    3 Measurement Measurement preparation (1) Cleaning the disposal box Empty the disposal box. IMPORTANT: If the disposal box is full, a slide transfer error or tip disposal error may occur, which may negatively af- fect the measurement values. IMPORTANT: After emptying the disposal box, be sure to load the disposal box.
  • Page 39 3 Measurement (2) Checking the recording paper Press the area shown in the illustration to release Cover the lock ( 1 ), and then lift up the cover ( 2 ). Press the concave section. Check the remaining recording paper (FUJI thermal paper).
  • Page 40 3 Measurement (3) Starting the FDC NX700 Press the power button to start the FDC NX700. NOTE If the equipment is connected to DMS, switch on the equipment before starting DMS. Failure to observe this precaution could interfere with data transfer. IMPORTANT: If the power has been shut off during measurement due to a power outage or other issue, replenish...
  • Page 41 3 Measurement (4) Checking the date and time Check that the date and time displayed on the LCD are correct. If they are incorrect, set the date and time using [Date & Time settings] in Normal mode. (Refer to Section 9.2.1) ...
  • Page 42: Basic Measurements

    3 Measurement Basic measurements NOTE By default, this analyzer switches off the light source lamp to save lamp life when the ana- lyzer has not been used for the time configured under [Lamp management] in Normal mode. When the lamp is off, refer to Section 3.3.1 before operating the equipment. WARNING: When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility.
  • Page 43 3 Measurement (3) Loading samples and slides WARNING: When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If you acci- dentally touch a sample, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area.
  • Page 44 3 Measurement 4. Align the direction and load the slide Slide cartridge cartridge at the sample location to be measured. 5. Make sure the sample rack and sample tube types are correct. NOTE To avoid a suction error or malfunction, use a sample tube that matches the la- bel on the sample rack.
  • Page 45 3 Measurement (4) Registering sample information Tap a sample rack display to display the sample in- formation registration screen. Enter or select the sample information (sample num- ber, sample ID, reference standard, and sample type), and then tap the [OK] button 7 . 1 Enter the sample number using the software key- board (Section 2.3).
  • Page 46 3 Measurement (5) Starting measurement Close the sample cover, and then lock it with the sampler cover lock key. WARNING: Ensure the sampler cover is locked during sample processing to prevent injuries and infection. When the display reads [Ready to test], the sam- pler cover can be opened.
  • Page 47 3 Measurement If sample application is started, the sample rack dis- play blinks during sample application. If sample application ends, the sample rack display switches off. When slides and samples have been loaded in multiple sample racks, the next sample rack display blinks orange, and then sample applica- tion is started.
  • Page 48 3 Measurement (6) Checking the measurement results After measurement of all samples has ended, tap button, and then check the measurement results. The measurement results are printed in the order that the slides were loaded in the cartridge. To retest (Section 4.5), tap the (retest) button.
  • Page 49: Lamp Off And Measurement-Start Scheduling

    3 Measurement 3.3.1 Lamp off and measurement-start scheduling By default, when this analyzer has not been used for some time (a time from 20 through 240 min- utes can be set) while [Ready to test] is displayed on the LCD, it switches off the light source lamp to save lamp life.
  • Page 50 3 Measurement If measurement is possible, it starts automatically. NOTE To cancel a scheduled measurement-start, tap the [STOP] button. 3-15 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 51: Measurement Results Screen

    3 Measurement Measurement results screen After measurement has ended, tap button, and then check the measurement results. NOTE Results of the last 270 samples are backed up in data. Result data display area < Display of measurement results > Green: Within reference interval Red: Outside reference interval Lower limit Upper limit...
  • Page 52 3 Measurement < Buttons of the measurement results screen > button Close the measurement results screen, and then return to the Measurement-preparation screen. 2 Change sample No. button Change the sample number for which meas- urement results are to be displayed. 3 Feed button Feed the recording paper with a fixed amount.
  • Page 53 3 Measurement < Calendar display > 1 Measurement month scroll buttons Change the display of the calendar month. 2 Date buttons Display the measurement results screen of a date. The currently displayed date is dis- played in blue. NOTE Dates with no measurements are displayed in gray and cannot be se- lected.
  • Page 54: Ise Test Measurements

    3 Measurement ISE test measurements IMPORTANT: Analyzer cannot detect expiries of ISE slides if the installed software version is earlier than 2.0. Be sure manually that slides are still valid. Otherwise, with the software version 2.0 or later is installed, analyzer can automatically detect ex- piries of the ISE slides that will expire 2023 or later or whose lot numbers are 130000 to 139999.
  • Page 55 3 Measurement 4. Load samples and slides. For details of the sample and slide loading method, refer to Section 3.3 (3). NOTE Place an ISE slide with the arrows fac- ing up and pointing opposite the sample label of the cartridge (in this case, “A”) as shown in the illustration.
  • Page 56: Dilution Tests

    3 Measurement Dilution tests There are 2 types of dilution measurement. • Dilute and measure automatically by using [Dilution setting] in Mode function for administrator to preset dilution factors for each test and sample type. Refer to Section 9.3.14  •...
  • Page 57 3 Measurement (Example) When diluting Sample A 2x 1. Check the remaining volume of the con- sumables (tips, mixing cups, and dilu- ent). If consumables are insufficient, supplement or replace them. • Loading tips Refer to Section 5.2  • Loading mixing cups Refer to Section 5.3 ...
  • Page 58 3 Measurement 4. In the list, tap [x2]. - : Dilution is performed according to [Dilution setting] in Mode function for administra- tor. None : No dilution NOTE If necessary, enter or choose other A sample information. ..0160503009 ..0160503010 STOP START...
  • Page 59 3 Measurement 6. Close the sample cover, and then lock it with the sampler cover lock key. WARNING: Ensure the sampler cover is locked during sample processing to prevent injuries and infection. When the display reads [Ready to test], the sam- pler cover can be opened.
  • Page 60: Crp Tests

    3 Measurement CRP Tests NOTE Each time a new box of CRP slides is used, a new QC card must be read. Refer to Section 3.10  NOTE Calibration is necessary for CRP measurements under the following conditions. • When the slide lot has changed •...
  • Page 61 3 Measurement 2. Load samples and slides. For details of the sample and slide loading method, refer to Section 3.3 (3). 一般比色 CM slide NOTE When measuring a CRP slide and CM slide at the same time, stack the CRP 図-4.9 slide last or first in the cartridge.
  • Page 62 3 Measurement 5. At the Measurement-preparation screen, tap the [START] button to start measure- ment. 3-27 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 63: Shutting-Down The Fdc Nx700

    3 Measurement Shutting-down the FDC NX700 Diluent display 1. Remove the samples. 2. Tap the diluent display or reference fluid display on the LCD to rotate the sample disk, and then remove the diluent and reference fluid from the analyzer. After removing the items, tap the [Finish] button to rotate the sample disk again. Reference fluid display Reference fluid...
  • Page 64 3 Measurement 5. Press the power button. Power button 6. Tap [OK]. CAUTION: Were you not able to operate on the screen, press the power button at front to turn off the power. NOTE When the analyzer will not be used for a long time, set the power switch to the [○] side.
  • Page 65 3 Measurement 7. Empty the disposal box. WARNING: Do not touch used slides and tips with your bare hands, as this may cause infection. If you accidentally touch an item, immediately rinse the contaminated area thoroughly under run- ning water, and then disinfect the area. Seek medical assistance if necessary.
  • Page 66: Sample Tubes For The Fdc Nx700

    3 Measurement Sample tubes for the FDC NX700 The following sample tubes can be used on the FDC NX700. • Ø16×100 mm blood collecting tube • Ø16×100 mm blood collecting tube (when using PF) • Ø13×100 mm blood collecting tube •...
  • Page 67 3 Measurement (2) Tube contents and inspection items that can be measured IMPORTANT: Refer to the “Instructions for Use” of the slides for usable blood collecting tubes and observe the listed items. IMPORTANT: For contents, inspection items, cap colors, and other details, contact the manufacturer of the blood collecting tube to be used.
  • Page 68: Fuji Sample Tubes

    3 Measurement 3.9.2 FUJI Sample Tubes Use these tubes in the place of blood collecting tubes when the sample volume is low. WARNING: When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If you acci- dentally touch a sample, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area.
  • Page 69: Sample Racks

    3 Measurement 3.9.3 Sample racks A sample rack is specified for each sample tube. Place each sample tube into the specified sample rack. NOTE The sample racks are optional. 0.5ml 1.5ml 0.5ml × × × × × × Sample rack 897N201058J FDC NX700 Instruction Manual 2022.04 3-34...
  • Page 70: Qc Card System

    3 Measurement 3.10 QC card system (1) Overview The QC card system compensates for variations between lots of FUJI DRI-CHEM slides and ensures uniform performance. Calculation method A variety of the reference fluids developed for FUJI 標 DRI-CHEM slides are measured on each lot of 準...
  • Page 71 (4) Using DI cards When you receive DI cards from FUJIFILM due to a change of the slide’s manufacturing pro- cess, read the DI cards according to the instructions. The reading method is the same as for QC cards.
  • Page 73: Applicable Usage Methods

    Applicable usage methods CRP calibration 4.1.1 Calibration measurement Calibrate when you start a new production number (lot) of CRP slides or when calibration has ex- pired. WARNING: As calibrator CP is derived from blood, it should be handled in the same way as a patient sample. When handling calibrator CP, always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility.
  • Page 74 4 Applicable usage methods 1. Read the QC card via the analyzer. Read the QC card of the calibration slide. Refer to Section 3.10  2. Tap the button. 3. Tap the [Calibration] button. 4. Tap the [Calibration] button. The Calibration Lot No. Screen appears. 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 75 4 Applicable usage methods 5. Check the displayed calibrator concen- tration settings (CP1, CP2, and CP3). IMPORTANT: When the calibrator concentration values are in- correct, calibration is not performed normally. NOTE Refer to the “Instructions for Use” of the calibrator. NOTE You can change the calibrator concen- tration settings by tapping the [Settings]...
  • Page 76 4 Applicable usage methods 9. Tap the [START] button to start calibra- tion measurement. If a slide with a different lot number has been loaded, an error screen appears. Tap the [OK] button to return to the previous screen. Check the slide, and then continue measurement.
  • Page 77 4 Applicable usage methods 10. If measurement ends, the calibration measurement results screen appears. NOTE Store printed calibration measurement results during the effective period of the calibration. NOTE The calibration compensated coeffi- cients are saved to the analyzer auto- matically. 11.
  • Page 78: Configuring Calibrator Concentration

    4 Applicable usage methods 4.1.2 Configuring calibrator concentration Correct the calibration concentration when the concentration included with the calibrator CP to be used and the concentration in the calibrator concentration display area at the Calibration Lot No. screen are different. 1. Tap the [Settings] button. The administrator password entry screen ap- pears.
  • Page 79 4 Applicable usage methods 3. Tap the [Input] field of each CP. The software keyboard screen appears. 4. Enter a value, and then tap the [OK] button. For the details of the software keyboard screen, refer to Section 2.3. IMPORTANT: Enter the value using Unit (A) (mg/dL). 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 80 4 Applicable usage methods 5. If the button is tapped, the entered values are registered, and the analyzer returns to the Calibration Lot No. screen. NOTE If the [Reset] button is tapped, the en- tered values are reset. If the button is tapped, the entered values are regis- tered and printed.
  • Page 81: Verifying Calibration

    4 Applicable usage methods 4.1.3 Verifying calibration There is a function to verify calibration. 1. At the Calibration screen, tap the [Verifying calibration] button. 2. Tap the checkbox of a level under Cali- bration level selection to select a level to be measured.
  • Page 82 4 Applicable usage methods 4. Follow the on-screen instructions to load consumables such as slides, calibrator, or diluent. NOTE During verification, measurements are performed one by one for the selected calibrator (CP). 5. Close the sample cover, and then lock it with the sampler cover lock.
  • Page 83: Interrupting Sample Application And Stat Tests

    4 Applicable usage methods Interrupting sample application and STAT tests Before adding consumables during sample application, tap the [PAUSE] button to interrupt sample application. When the analyzer has run out of tips during sample application, the analyzer stops (as when the [PAUSE] button was tapped). In this paused state, you can display the loading screen by touching the display of each consumable.
  • Page 84: Requesting A Stat Test

    4 Applicable usage methods (2) Canceling the remaining sample application after interrupting sample applica- tion If the [STOP] button is tapped while the analyzer is interrupted, the remaining sample application is canceled. If the cancel operation is completed, the LCD changes to [Ready to test]. All sample rack dis- plays get switched off.
  • Page 85 4 Applicable usage methods 2. Tap sample rack display C. 3. Register the measurement information, and then tap the [OK] button. 897N201058J FDC NX700 Instruction Manual 2022.04 4-13...
  • Page 86 4 Applicable usage methods 4. Tap the [RESTART] button to (re)start measurement. NOTE In this example, the analyzer finishes the remaining application of Sample A, and then measures Sample C. Measurement order: A Þ C Þ B 4-14 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 87: Manual Sampling

    4 Applicable usage methods Manual sampling 4.3.1 Measuring by manual sample application WARNING: When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If you acci- dentally touch a sample, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area.
  • Page 88 4 Applicable usage methods 3. Tap the [START Manual pipetting] button. The sampler unit moves above the sample disk, and the sample disk turns so that Sample A comes to the front. 4. Tap sample rack display A. In the Sample information registration screen, register the measurement information, and then tap the [OK] button.
  • Page 89 4 Applicable usage methods 7. Tap the [START] button. A slide is transferred to the sample application position. 8. Prepare so that sample application with the pipette is always possible, and then tap the [START] button. 897N201058J FDC NX700 Instruction Manual 2022.04 4-17...
  • Page 90 4 Applicable usage methods 9. After the progress bar has finished (all □ change to ■), apply the sample to the slide while [Spot now] is displayed. The spotted slide is automatically transferred to the measurement section after the display of [Spot now] stops.
  • Page 91: Using A Pipette

    4 Applicable usage methods 4.3.2 Using a pipette WARNING: When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the sample handling rules of your facility. If you acci- dentally touch a sample, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area.
  • Page 92 4 Applicable usage methods Applying samples to slides 1. Move the tip of the pipette near the slide. 2. Slowly push the push button and make a drop. 液滴を作ってスライ ドに近づけます 液滴を作ってスライ ドに近づけます Make a drop and then move it near the slide 液滴を作ってスライ...
  • Page 93: Tests Using A Plasma Filter (Pf)

    4 Applicable usage methods 4.4 Tests using a plasma filter (PF) By using a PF, you can filter a plasma sample from a whole blood sample without using a centrifuge. NOTE The FDC NX700i cannot measure using a PF. IMPORTANT: Insert one (sample) PF in the sample disk. If measurements are performed on multiple samples simultaneously using PFs, the blood cells may precipitate, making it impossible to obtain sufficient plasma, and the measurement results for GPT (ALT) slides may be affected.
  • Page 94 4 Applicable usage methods PF cards (1) Overview When the measurement values for filtered plasma are compared to the values for plasma sep- arated on a centrifuge, systematic differences may be found. Always read the PF card before using a PF. The PF card contains information to compensate for these differences. (2) Using PF cards One PF card is included in one PF box (with 50 filters).
  • Page 95: Measurement

    4 Applicable usage methods 4.4.1 Measurement 1. Prepare the slides. 2. Prepare the consumables. Prepare the tips, mixing cups, diluent, and refer- ence fluid as required. • Loading tips Refer to Section 5.2  • Loading mixing cups Refer to Section 5.3 ...
  • Page 96 4 Applicable usage methods 4. Load the sample. (1) Place the specified amount of whole blood sample in a blood collecting tube. NOTE The PF has about 3 mL of dead vol- ume (that will be absorbed into the filter). If the sample volume is low, the filtered sample volume will be too low for measurement.
  • Page 97 4 Applicable usage methods 6. Close the sample cover, and then lock it with the sampler cover lock key. WARNING: Ensure the sampler cover is locked during sample processing to prevent injuries and infection. When the display reads [Ready to test], the sam- pler cover can be opened.
  • Page 98: Blood Collecting Tubes For Tests Using A Pf

    4 Applicable usage methods 4.4.2 Blood collecting tubes for tests using a PF NOTE The FDC NX700i cannot measure using a PF. The following blood collecting tubes can be used with PF measurements: φ16×100 mm, φ13×100 mm, and φ13×75 mm. (Outer diameter of Ø16 indicates Ø15.0 to 16.6 mm and Ø13 indicates Ø12.0 to 13.3 mm.) IMPORTANT: Use blood collecting tubes containing Heparin.
  • Page 99: Retesting

    4 Applicable usage methods Retesting You can test again using the previous sample number, sample ID, sample type setting, and refer- ence interval setting by tapping the (retest) button at the measurement results screen. Further- more, when a measurement cannot be performed due to an error, you can skip the re-entry and re- configuration of the setting information by tapping the (retest) button.
  • Page 100 4 Applicable usage methods (Example) Retesting sample A 1. When [Ready to test] appears on the LCD, make sure that the analyzer is not applying samples, and then load the consumables and sample. Refer to Section 3.3  2. Tap the (retest) button on the Meas- urement results screen.
  • Page 101 4 Applicable usage methods NOTE In case of a sample scheduled for re- testing, appears on the sample rack display. 5. Close the sample cover, and then lock it with the sampler cover lock key. WARNING: Ensure the sampler cover is locked during sample processing to prevent injuries and infection.
  • Page 102: Slide Loading Order When Measuring Different Items Simultaneously

    4 Applicable usage methods 4.6 Slide loading order when measuring different items simul- taneously NOTE Any order of slide stacking can be measured. However, please follow the recommended slide stacking rules below. Otherwise, the analyzer may consume excessive amounts of dis- posal tips and cups. (1) Measuring CRP and CM (colorimetric) slides Always put a CRP slide first or last.
  • Page 103: Automatic Acquisition Of Sample Information (Work Lists)

    4 Applicable usage methods 4.7 Automatic acquisition of sample information (work lists) If the analyzer is connected to DMS, you can receive patient IDs and their sample information by tapping the button at the Measurement-preparation screen before measurement. NOTE DMS is sold separately. 1. Tap the button.
  • Page 104 4 Applicable usage methods 2. Tap the sample rack for which you want to register sample information. 3. Tap the sample information you want to register. NOTE If the [Select all] button is tapped, infor- mation is registered in order from the first sample information.
  • Page 105 4 Applicable usage methods The work list measurement items of the regis- tered sample information are displayed. If the button is tapped, registration NOTE is cleared. NOTE The work list measurement items can be printed by tapping the button. NOTE If the [Deselect all] button is tapped, registrations for all racks are cleared.
  • Page 106 4 Applicable usage methods NOTE Work lists can also be obtained from the sample information registration screen. A ..0160503009 ..0160503010 STOP START MYID001 4-34 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 107: Batch Change Of Sample Types

    4 Applicable usage methods Batch change of sample types Sample types can be switched as batch (all of sample rack A through E). 1. Tap the (Function menu) button. 2. Tap the [PS], [W] or [U] button for sam- ple type switching. NOTE This setting is applied to sample rack A through E.
  • Page 108: Measuring Plasma And Urine Samples At The Same Time

    4 Applicable usage methods Measuring plasma and urine samples at the same time (Example) Measuring a plasma sample as Sample A and a urine sample as Sample B 1. Tap sample rack display A. 2. Check the sample type setting. Make sure the sample type is [Plasma/Serum].
  • Page 109 4 Applicable usage methods 4. Tap sample rack display B. 5. Tap the [Sample type] button, and then choose [Urine]. 6. Tap the [OK] button to confirm the meas- urement conditions of Sample B. The sample rack display lights up green. A ..0160503009 ..0160503010 STOP START MYID001 7.
  • Page 110: Editing Sample Numbers And Sample Ids

    4 Applicable usage methods 4.10 Editing sample numbers and sample IDs You can edit the sample number and sample ID of measurement results. You can also send edited data to a host computer. 1. To edit the sample number, tap the [No.] button at the measurement results screen (Refer to Section 3.4).
  • Page 111: Using The Sample Barcode Reader (Sold Separately)

    4 Applicable usage methods 4.11 Using the sample barcode reader (sold separately) You can easily enter the sample ID and operator ID by using the optional sample barcode reader. IMPORTANT: If the sample barcode reader malfunctions or the print quality of the barcode label is poor, incor- rect barcode data may be read.
  • Page 112 4 Applicable usage methods 2017-01-01 10 : 00 Ready to test 2. Tap the [Sample ID] field. The software keyboard appears. NOTE You can also enter the sample ID using the barcode reader on the measurement results screen. A ..0160503009 ..0160503010 STOP...
  • Page 113 4 Applicable usage methods <When inputting the operator ID> 1. Tap the operator ID button. The software keyboard appears. 2. Read the operator ID using the sample barcode reader. After reading is complete, the analyzer makes a beep sound. The operator ID is displayed on the LCD and registered.
  • Page 115: Loading Consumables

    Loading consumables Loading consumables To load consumables, tap the display of each consumable to display the loading screen. When the consumable loading is complete, press the [Finish] button to exit. NOTE If there are not enough consumables (tips, mixing cups, diluent, reference fluid, etc.), sam- ple application may be interrupted or cancelled.
  • Page 116: Setting Tips

    5 Loading consumables Setting tips Tips are required for all measurements. Check the number of remaining tips on the LCD before per- forming measurements. If necessary, add new tips. WARNING: If the analyzer is applying samples, tap the [PAUSE] button to interrupt application (refer to Sec- tion 4.2), and then load the tips.
  • Page 117 5 Loading consumables 4. After the sample disk stops, remove the tip racks from the sample disk. NOTE Always remove the tip racks from the sample disk before loading tips. Tip rack 5. Load tips in all holes of the tip racks. NOTE Always set tips in all holes of the tip racks.
  • Page 118: Loading Mixing Cups

    5 Loading consumables Loading mixing cups Mixing cups are necessary for measurements requiring dilution (refer to Section 3.6) and CRP measurements (refer to Section 3.7). Furthermore, mixing cups are also necessary when urine (U) is selected for the sample type during an ISE measurement because measurement is performed automatically at 2x dilution.
  • Page 119 5 Loading consumables 4. Remove the mixing cups to be replaced Mixing cups from the sample disk. Set one or two new mixing cups. WARNING: Used mixing cups are infectious industrial waste. Process this waste in compliance with any appli- cable regulations in your country, such as by incin- eration, melting, sterilization, or disinfection.
  • Page 120: Loading Diluent

    5 Loading consumables Loading diluent Diluent is necessary for dilution measurements (refer to Section 3.6) and CRP measurements (refer to Section 3.7). Dilution and CRP measurements each require a specified diluent. Before measure- ment, refer to the technical documentation for the diluent to be used. Furthermore, mixing cups are also necessary when urine (U) is selected for the sample type during an ISE measurement because measurement is performed automatically at 2x dilution.
  • Page 121 5 Loading consumables 2. With [Warming up] or [Ready to test] dis- played on the LCD, make sure that the analyzer is not applying samples. 3. Tap the diluent display on the LCD. Diluent display 4. The Diluent loading screen appears. The sample disk turns, and the consumables are moved to the front.
  • Page 122 5 Loading consumables 6. Tap the [Finish] button to exit. The sample disk turns, and the sample rack is moved to the front. 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 123: Loading Reference Fluid

    5 Loading consumables 5.5 Loading reference fluid Reference fluid is necessary for ISE measurements. (Refer to Section 3.5). WARNING: When handling reference fluid, follow biohazard procedures (wear gloves, a lab coat, and safety goggles) in accordance with the medical handling rules of your facility. If you accidentally touch an item, immediately rinse the contaminated body part thoroughly under running water.
  • Page 124 5 Loading consumables 2. With [Warming up] or [Ready to test] dis- played on the LCD, make sure that the analyzer is not applying samples. 3. Tap the reference fluid display on the LCD. Reference fluid display 4. The Reference fluid loading screen ap- pears.
  • Page 125 5 Loading consumables NOTE Make sure the reference fluid cap is closed properly. Sealing 6. Tap the [Finish] button to exit. The sample disk turns, and the sample rack is moved to the front. 897N201058J FDC NX700 Instruction Manual 2022.04 5-11...
  • Page 127: Maintenance

    Maintenance 6 Maintenance To keep the performance of the FDC NX700 at its best, users must perform daily and periodic mainte- nance, and designated repair personnel must perform maintenance. When repairs are necessary, con- tact repair personnel. WARNING: When you clean the analyzer, always wear gloves, a lab coat, and safety goggles. If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area.
  • Page 128: Periodic Maintenance Items

    6 Maintenance Periodic maintenance items 6.1.1 Items for daily and periodic inspection by the user Users must perform the daily and periodic maintenance and replacement of parts in the following table. Inspection Location Cleaning interval Replacement interval Procedure interval Refer to Air filter Once a month Section 6.2.
  • Page 129: Items For Maintenance And Inspection By Repair Personnel

    6 Maintenance 6.1.2 Items for maintenance and inspection by repair personnel Designated repair personnel must perform the maintenance and replacement of parts in the follow- ing table. Parts Replacement interval Interference filter Once every 4 years Syringe Once every 3 years Incubator pressure plate Inspection once every 3 years WARNING:...
  • Page 130: Cleaning The Air Filters

    6 Maintenance 6.2 Cleaning the air filters The air filters should be checked and cleaned once a month. IMPORTANT: If the air filters are not cleaned, an increase in the temperature inside the analyzer may negatively affect the measurement values. NOTE Clean all 4 filters. 1. Switch off the analyzer power.
  • Page 131 6 Maintenance 3. Remove dust adhering to the filter with a Air filter vacuum cleaner or running water. NOTE If you wash the filter with running water, make sure it is dried well before setting in the analyzer. 4. Insert the filter into the louver and put the louver back in the analyzer.
  • Page 132: Inspecting And Cleaning The Transfer Bar, Incubator, Sample Application Unit, And Ise Unit

    6 Maintenance Inspecting and cleaning the transfer bar, incubator, sample application unit, and ISE unit Measuring results are affected by stains (from blood samples, etc.) inside of the incubator and ISE unit. These parts should be checked and cleaned at least once every three months. When a slide transfer error or a tip ejecting error occurred, clean the sample application unit.
  • Page 133 6 Maintenance 3. Tap the [Reference plate level check] button to run a check. If check ends, the Check results screen appears. 4. Check the results of [Reference plate level check]. When has appeared in the check results, cleaning is necessary. Tap the [OK] button. Using the procedure on the following pages, clean the transfer bar, incubator, sample application unit, and ISE unit.
  • Page 134: Cleaning The Transfer Bar, Incubator, Sample Application Unit, And Ise

    6 Maintenance 6.3.2 Cleaning the transfer bar, incubator, sample application unit, and ISE unit (1) Cleaning the transfer bar 1. Tap the [Cleaning mode] button in Normal mode. 2. Tap the [Start Cleaning] button. NOTE Keep your fingers clear as the sample disk moves.
  • Page 135 6 Maintenance 3. The screen to the left appears. 4. Lower the tip rack from the analyzer. 5. Loosen the 2 thumbscrews and pull the sample application unit cover frontward to remove it. Thumbscrew 6. Remove the two slide transfer weights. NOTE There are 2 types of slide transfer weight, one each for CM tests and ISE...
  • Page 136 6 Maintenance 7. Use gauze or a cotton swab that has Transfer bar been soaked in lukewarm or cold water to wipe the upper surface of the transfer bar visible in the sample application unit. 8. Next, clean the incubator, sample appli- cation unit, and ISE test unit.
  • Page 137 6 Maintenance WARNING: Always unplug the power cord before the following procedure. (2) Incubator 1. Open the sampler cover and panel unit. Move the sampler unit to the right, and then remove the panel unit pedestal. NOTE Be careful not to open the reference fluid cap.
  • Page 138 6 Maintenance 4. Loosen the 5 screws on the upper sur- face of the incubator, and then remove the incubator. Thumbscrew 5. Moisten gauze or a similar material in Incubator pressure plate sterilizing ethyl alcohol, strongly wring out the gauze, and then wipe the 13 pressure plates on the backside of the incubator.
  • Page 139 6 Maintenance 7. Turn over the reference plate and clean Reference Reference white plate black plate the white plate and black plate with a dry cotton swab. Photometer 8. Clean the photometer head with a dry cotton swab. 9. Load the reference plate. Tighten the 2 thumbscrews evenly.
  • Page 140 6 Maintenance (3) Sample application unit 1. Make sure that the sample application unit cover and two slide transfer weights were removed for cleaning of the trans- fer bar. 2. Wipe the sample application unit with Sample application unit gauze or a cotton swab moistened with lukewarm or cold water.
  • Page 141: Reassembling Parts And Ending Cleaning

    6 Maintenance 6.3.3 Reassembling parts and ending cleaning 1. Reassemble the incubator. Align the arrow marks on the incubator and the reference plate, and then tighten the 5 thumb- screws evenly by tightening opposing thumb- screws in numerical order ( 1 to 5 ). IMPORTANT: Tighten the five thumbscrews securely.
  • Page 142: Replacing Recording Paper

    6 Maintenance Replacing recording paper When red lines appear along the sides of the recording paper, the printer is nearly out of paper, so replace the recording paper roll with a new one. NOTE Use the recording paper specified for the FDC NX700. 1.
  • Page 143: Replacing And Cleaning The Light Source Lamp

    6 Maintenance Replacing and cleaning the light source lamp The average life of the lamp is about 1,000 hours. Replace the light source lamp if a photometer gain error has occurred or the lamp’s cumulative illumination time is near 1,000 hours. CAUTION: The light source lamp gets very hot.
  • Page 144 6 Maintenance 4. Wipe the light source unit lens using a dry cotton swab. Lens 5. Turn the thumbscrew of the lamp unit to the left, remove the lamp, and then re- place it with a new one. Lamp base IMPORTANT: Tighten the thumbscrews securely.
  • Page 145 6 Maintenance 6. Insert the connector of the light source Thumbscrew B lamp. 7. Insert the lamp unit into the analyzer with thumbscrew B facing up. Tighten thumbscrew A securely. IMPORTANT: Tighten the thumbscrew securely. Otherwise, the measurement results may be negatively affected. Thumbscrew A Connector 8.
  • Page 146: Inspecting And Replacing The Sampler O-Ring

    6 Maintenance Inspecting and Replacing the Sampler O-ring The sampler nozzle O-ring wears over time. Inspect it once a month and replace it once a year. WARNING: When inspecting and replacing O-rings, always wear gloves, a lab coat, and safety goggles. If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly un- der running water, and then disinfect the area.
  • Page 147 6 Maintenance 3. Tap [Sampler leak check]. 4. Follow the on-screen instructions to pre- pare for the leak check. 5. Tap the nozzle to be checked to select it, and then run the leak check. Once check ends, the Check results screen ap- pears.
  • Page 148 6 Maintenance 6. Check the results of the [Sampler leak check]. If the check results are [OK], it is not necessary to replace the O-ring. NOTE The check results can be printed by tapping the button. If the check results are [NG], tap the [Replace O-ring] button, and then replace the O-ring according to the following procedure.
  • Page 149: Exchanging The Sampler O-Ring

    6 Maintenance 6.6.2 Exchanging the sampler O-ring The sampler nozzle O-ring wears over time. Replace it once a year. WARNING: Process used O-rings as infectious waste. IMPORTANT: If you do not replace the O-ring periodically, the sample application volume will be inaccurate, which may negatively affect the measurement values.
  • Page 150 6 Maintenance 4. Tap [Replace O-ring]. 5. Tap the nozzle on which you want to replace the O-ring to select it. The sampler unit moves, and the sampler noz- zle comes down. CAUTION: Keep your fingers away as the sampler unit moves.
  • Page 151 6 Maintenance 7. Tap [Leak check] to check for leaks. Refer to Section 6.6.1  897N201058J FDC NX700 Instruction Manual 2022.04 6-25...
  • Page 152: Cleaning The Slide Reading Section

    6 Maintenance Cleaning the slide reading section When a slide-detection error has occurred, dust and other soiling may be the cause. Clean the slide reading section to avoid these errors. 1. Switch off the analyzer power. 2. Open the sampler cover and panel unit. 3.
  • Page 153: Cleaning And Replacing The Reference Fluid Cap Packing

    6 Maintenance 6.8 Cleaning and replacing the reference fluid cap packing Clean the reference fluid cap packing each time the reference fluid is replaced. Replace the reference fluid cap packing once every three months. IMPORTANT: If you do not clean and replace the reference fluid cap packing, the reference fluid will be concen- trated, which may negatively affect the measurement values.
  • Page 154 6 Maintenance 4. Set the reference fluid cap packing prop- erly. NOTE Do not touch the bottom side (reference fluid side) of the reference fluid cap packing. Reference fluid cap packing 6-28 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 155: Pf Unit Maintenance

    6 Maintenance PF unit maintenance 6.9.1 Cleaning the PF unit (sensor, reference plate, and suction pad) Clean when errors occurred during PF filtering. 1. Remove the sample rack and slide car- tridge. 2. Tap the button. 3. Tap the [Normal mode] button. 4.
  • Page 156 6 Maintenance 5. Tap the [Cleaning] button. 6. Tap the [OK] button. CAUTION: Keep your fingers away as the sample disk and PF section move. 7. Clean according to the on-screen in- structions. Surface of PF sensor PF Suction Pad (glass surface) (black) 8.
  • Page 157 6 Maintenance 6.9.2 Inspecting the PF Suction Pad The PF Suction Pad must be inspected monthly and replaced yearly. 1. Tap the button. 2. Tap the [Normal mode] button. 3. Tap [PF maintenance]. 897N201058J FDC NX700 Instruction Manual 2022.04 6-31...
  • Page 158 6 Maintenance 4. Tap the [Leak check] button. 5. Follow the on-screen instructions to load the PF sample rack and PF leak check tool on Sample rack A, and then tap the [START] button to run the leak check. If check ends, the Check results screen appears. CAUTION: Keep your fingers away as the sample disk and PF section move.
  • Page 159 6 Maintenance If the check results are [NG], tap the [Cleaning] button, clean the PF unit, and then run the leak test again. Refer to Section 6.9.1  If this does not resolve the problem, tap the [Procedure] button, and then check the attach- ment or replace the PF pump tube to using the procedure on the following page.
  • Page 160: Replacing The Pf Suction Pad

    6 Maintenance 6.9.3 Replacing the PF Suction Pad Replace the PF Suction Pad once a year. WARNING: Process used PF Suction Pad as infectious waste in compliance with any applicable regulations in your country. 1. With [Warming up], [Lamp warming up] or [Ready to test] displayed on the LCD, make sure that the analyzer is not apply- ing samples.
  • Page 161 6 Maintenance 4. After the sample disk stops, remove the tip racks from the sample disk. Tip rack 5. Tap the [Finish] button to exit. The sample disk turns, and the sample rack is moved to the front. 6. Press the power button. Power button 897N201058J FDC NX700 Instruction Manual 2022.04 6-35...
  • Page 162 6 Maintenance 7. Tap [OK]. 8. Pull the part of PF Suction Pad down- ward as shown left. 9. Turn the PF Suction Pad in the direction shown and remove it while loosening. image 10. Install the new PF Suction Pad by turn- ing it in the opposite direction.
  • Page 163: Replacing The Pf Pump Tube

    6 Maintenance 6.9.4 Replacing the PF pump tube WARNING: Before the following procedure, always switch off the power of the analyzer, and unplug the power cord. Replace the PF Tube once a year. Replacement is also necessary if the PF has not been used for one month or longer.
  • Page 164 6 Maintenance 4. While pushing the head of joint (G), pull PF pump the joint (G) to the right to remove the Joint (G) tube (D). Red tube (D) Pull off joint (G) to the right Joint (G) while pushing this part to the right.
  • Page 165 6 Maintenance 8. Attach the front bottom cover. 9. Attach the disposal box. 10. After the replacement, run the leak check. If the check results are [NG], perform the re- placement again. Refer to Section 6.9.2  897N201058J FDC NX700 Instruction Manual 2022.04 6-39...
  • Page 167: Quality Control

    Quality control IMPORTANT: To maintain the accuracy of your test results, it is recommended to perform quality control on routine bases by use of control fluids. Perform quality control in compliance with local regulation and proce- dure which you have to obey. Also, it is also recommended to do after major repairs are performed on the analyzer.
  • Page 168: Control Chart

    7 Quality control Control chart 1. Tap the button. 2. Tap the [Control] button. 3. The measurement-preparation screen of control mode appears. 4. Tap the (graph display) button. 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 169 7 Quality control 5. Tap the button of the measurement item to be viewed. 6. The control chart appears. NOTE If the control liquid button is tapped, you can change the control liquid to be dis- played. NOTE Measurement results of outside of measurement range (@), disposal box open (*), and expired slide (#) are not reflected in the control chart.
  • Page 170 7 Quality control 7. If a measurement date is selected and then the [Detail] button is tapped, the de- tail screen appears. < Display of measurement results > Green: when within the acceptable range of control Red: when outside of the acceptable range of control Acceptable Acceptable range...
  • Page 171: Troubleshooting

    Troubleshooting WARNING: When you troubleshoot, always wear gloves, a lab coat, and safety goggles. If you accidentally touch a contaminant, immediately rinse the contaminated area thoroughly under running water, and then disinfect the area. Seek medical assistance if necessary. IMPORTANT: If an analyzer malfunction (error) appears before or during measurement or warnings appear in the measurement results, the measurement results may be inaccurate.
  • Page 172: Error Log

    8 Troubleshooting 8.1.1 Error log You can view the history of errors that occurred by using [Error log] in the function menu. 1. Tap the button. 2. Tap the [Error log] button. An error log list is displayed. 3. Tap an error. The error details are displayed.
  • Page 173 8 Troubleshooting 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 174: Error Code And Reference Section Table

    8 Troubleshooting 8.1.2 Error code and reference section table Error code Error description Reference Reference section page E0021 The disposal box was open during measurement 8.2.15 8-43 E0035 Photometric control system error 8.2.8 8-26 E0050 Light source lamp is burned out 8.2.5 (2) 8-21 E0080, E0081...
  • Page 175 8 Troubleshooting Error code Error description Reference Reference section page E1502, E1503, Temperature control error 8.2.7 8-25 E1504 E1510, E1511, ISE temperature control error 8.2.7 8-24 E1512, E1514, E1515 E1520, E1521 Installation environment temperature error 8.2.7 8-24 E1530, E1531, Signal system error inside circuit board 8.2.8 8-27 E1532...
  • Page 176 8 Troubleshooting Error code Error description Reference Reference section page W0020 The disposal box is open 8.2.15 8-43 W0022 The disposal box is full 8.2.15 8-43 W0030 Insufficient light intensity 8.2.5 (1) 8-20 W0036 Reference black plate soiling 8.2.5 (1) 8-20 W0040, W0041 The sample type configuration and sample type...
  • Page 177 8 Troubleshooting Error code Error description Reference Reference section page W4011 Panel card error 8.2.9 (5) 8-33 W4012 DI card error 8.2.9 (4) 8-32 W4013, W4014 QC card error 8.2.9 (5) 8-33 W4220 Errors related to calibration verification 8.2.14 (3) 8-41 W4460-W4474, Signal system error inside circuit board...
  • Page 178: Printout Reference Table

    8 Troubleshooting 8.1.3 Printout reference table Printout Meaning and conditions Troubleshooting Measurement value is higher than the upper limit of the preset reference interval. Measurement value is lower than the lower limit of the preset reference interval. > Measurement value is higher the upper limit of the meas- urement range.
  • Page 179 8 Troubleshooting The valid term of the slide has expired. For immuno-chemical tests, the validity term of the slide or the validity term of calibration has expired. IMPORTANT: The measurement value might be incorrect. IMPORTANT: Expired slides cannot be used. Always use valid slides.
  • Page 180: Troubleshooting

    8 Troubleshooting Troubleshooting 8.2.1 Startup errors (1) The analyzer does not start after the power button is pressed. Check that the power cable is properly connected. And, turn the power switch OFF and ON again, and press the power button. If the analyzer still does not start, please contact your dealer for assistance.
  • Page 181: Slide Reading Errors

    8 Troubleshooting 8.2.3 Slide reading errors Error Error message Error condition Troubleshooting code W0173 Slide identification The slide information Clean the slide reader, and then rerun on backside of the slide the measurement. could not be read cor- No slide information. Refer to Section 6.7 ...
  • Page 182 8 Troubleshooting (2) When tip cannot be ejected Error Error message Error condition Troubleshooting code E0086 Sample tip eject ERR Tips cannot be dis- After analyzer shutdown, empty dis- carded. posal box, and then turn the power ON. After analyzer shut- down, empty disposal box, and then turn the power ON.
  • Page 183 8 Troubleshooting (3) When there are insufficient number of or no mixing cups Error Error message Error condition Troubleshooting code W0151 Lack of mixing cup There are not enough After you replace the mixing cups, re- mixing cups. run from the measurement request. Lack of mixing cups. Refer to Section 5.3 ...
  • Page 184 8 Troubleshooting (5) During a sample aspiration or application error Error Error message Error condition Troubleshooting code E0111 Aspiration clogging Clogging was detected The sample may contain fibrin deposits. ERR (sample) during sample aspira- After checking the sample, rerun from tion.
  • Page 185 8 Troubleshooting (6) When application is poor Error Error message Error condition Troubleshooting code Sample application problem Because the sample is Check the following points. not applied to the slide, (a) Make sure that FUJI DRI-CHEM the error mark [<] or [@] AUTO TIPS are being used.
  • Page 186 8 Troubleshooting (7) During a diluent/reference fluid volume error Error Error message Error condition Troubleshooting code W0141 Surface detection ERR There is too much dilu- Follow the instructions of the error code (diluent) ent volume. to replace the diluent or reference fluid, and then rerun from measurement re- Diluent volume is too quest.
  • Page 187 8 Troubleshooting (8) During a diluent / reference fluid error Error Error message Error condition Troubleshooting code E0121 Aspiration clogging Clogging was detected Check whether the diluent / reference ERR (diluent) during diluent aspira- fluid are loaded in a proper blood col- tion. lecting tube or tube. If there was a load- Clogging was detected ing problem, replace the diluent and during sampling.
  • Page 188 8 Troubleshooting (9) During a calibrator volume error Error Error message Error condition Troubleshooting code W0061 Calibrator shortage There is not enough Check the calibrator liquid volume, and calibrator. then measure again. Calibrator is insuffi- cient. Check the calibrator volume and perform calibration or verify calibration again.
  • Page 189 8 Troubleshooting (10) During a sampler drive system error Error Error message Error condition Troubleshooting code E1000, S a m p l e r u p / d o w n Movement error of Switch off the analyzer when the meas- E1001, motor ERR sample/dilution nozzle...
  • Page 190: Photometric System Errors

    8 Troubleshooting 8.2.5 Photometric system errors IMPORTANT: If a photometric system error has occurred, perform each troubleshooting step listed below and rerun the measurement. (1) During a fluctuation of light source intensity Error Error message Error condition Troubleshooting code Soiled black plate W0036 The reference plate is Switch off the analyzer and clean each WARNING dirty.
  • Page 191 8 Troubleshooting (2) When lamp is burned out Error Error message Error condition Troubleshooting code E0050 Lamp replacement The light source lamp Switch off the power and replace the is burned out. light source lamp. Switch the analyzer Low light intensity. on again after replacement.
  • Page 192: Transfer Errors

    8 Troubleshooting 8.2.6 Transfer Errors Error Error message Error condition Troubleshooting code E0200- Feed motor ERR Movement error of the The sample application unit or the incu- E0204, slide supply motor bator may be dirty or the disposal box Problem with the slide E0220- may be full of slides and tips.
  • Page 193 8 Troubleshooting Error Error message Error condition Troubleshooting code E1600- Sample disc motor Movement error of the The sample application unit or the incu- E1603, disk motor of the sam- bator may be dirty or the disposal box W1604 ple disk section may be full of slides and tips.
  • Page 194: Temperature Controller Errors

    8 Troubleshooting 8.2.7 Temperature Controller Errors Error Error message Error condition Troubleshooting code E0565 ISE temperature con- When the measure- Please contact your dealer for assis- trol ERR ment results are ap- tance. pended with a “+” or “-” Problem with ISE. mark or an error listed Contact your dealer.
  • Page 195 8 Troubleshooting Error Error message Error condition Troubleshooting code E1502 Temperature control When the measure- Switch off the power and connect the ment results are ap- incubator cable again. pended with a “$” mark Incubator temperature NOTE After inserting the connector of or an error listed to the sensor disconnected.
  • Page 196: Circuit Board Signal Errors

    8 Troubleshooting 8.2.8 Circuit board signal errors Error Error message Error condition Troubleshooting code E0035 Light too intense A circuit board signal If the error still occurs after switching the error has occurred. analyzer off and on, please contact your Light too intense.
  • Page 197 8 Troubleshooting Error Error message Error condition Troubleshooting code E0540 ISE AD conversion A circuit board signal When an error occurs after turning the initialization ERR error has occurred. power on and off, please contact your dealer for assistance. Tests other than For ISE measurement, ISE do not require a restart.
  • Page 198 8 Troubleshooting Error Error message Error condition Troubleshooting code E4454 Memory loading ERR A circuit board signal Please contact your dealer for assistance. error has occurred. Contact your dealer. W4460 Memory recovery WARNING (setting W4461 info.) W4464 Recovery performed W4469 from backup memory.
  • Page 199 8 Troubleshooting Error Error message Error condition Troubleshooting code E4475 Memory initial-value A circuit board signal Please contact your dealer for assistance. recovery ERR error has occurred. Recovery performed from backup memory. Contact your dealer. W4477 Memory recovery Check the operator ID setting. WARNING (Operator ID setting) Recovery performed...
  • Page 200 8 Troubleshooting Error Error message Error condition Troubleshooting code W4721 Clock control failure A circuit board signal Set the date and time using [Date & Time WARNING error has occurred. settings] in Normal mode. Set date & time. Refer to Section 9.2.1 ...
  • Page 201: Qc Card, Pf Card, And Di Card Errors

    8 Troubleshooting Error Error message Error condition Troubleshooting code W5050 No USB memory The USB memory is Insert USB memory. not inserted. USB memory can not If the error occurs again, please contact be recognized. your dealer for assistance. E5500 A b n o r m a l s a m p l e Signal system error When an error occurs after turning the...
  • Page 202 8 Troubleshooting Error Error message Error condition Troubleshooting code W4010 No QC type info. There is no type infor- Read the corresponding QC card. mation for the corre- No type information sponding item. corresponding to the slide. Read QC card corre- sponding to the slide.
  • Page 203: Slide Loading Errors

    8 Troubleshooting Error Error message Error condition Troubleshooting code W4011 QC card ERR Panel card read error Please contact your dealer for assis- tance. Cannot save data. W4013 QC card ERR A QC card data error Check whether the QR code of the QC has occurred.
  • Page 204 8 Troubleshooting Error Error message Error condition Troubleshooting code W0042 Incorrect sample rack A re-inspection was Load a PF sample rack in which a WARNING attempted for a meas- sample extracted by PF has been urement that uses the loaded. Correct sample rack PF, but a rack other not set in place.
  • Page 205: Ise Measurement Errors

    8 Troubleshooting 8.2.11 ISE measurement errors IMPORTANT: ISE measurement values to which error codes have been appended (E1 - E8) may not be correct. Perform the following troubleshooting and rerun the measurement. (1) When a self-test error occurred Error Error message Error condition Troubleshooting code...
  • Page 206 8 Troubleshooting (2) Errors are appended to measurement results During an ISE measurement, the checks listed Check menu Description in the table to the left are performed. If there is a problem with a measurement value, Checking abnormal time course Drift check an error code from E1 through E8 is appended (voltage) during testing...
  • Page 207: Pf Errors

    8 Troubleshooting (3) Other Errors Error Error message Error condition Troubleshooting code W0091 U n a c c e p t a b l e I S E An ISE measurement Check ISE measurement errors that oc- test error has occurred, curred before this error, and then carry and ISE measurement out the corresponding troubleshooting.
  • Page 208 8 Troubleshooting Error Error message Error condition Troubleshooting code E1300, PF motor ERR PF motor error move- Turn the power on and off. If the error E1301, ment error occurs again, please contact your dealer Problem with the plas- E1303, for assistance.
  • Page 209: Errors Related To Data Communication, Sample Barcode Reader

    8 Troubleshooting 8.2.13 Errors Related to Data Communication, Sample Barcode Reader (1) When a communication error occurs Error Error message Error condition Troubleshooting code W4500, Serial communica- Serial communication Check the following: communication W4501, tion ERR error cable, host computer, and communica- W4502 tion destination settings.
  • Page 210: Calibration Errors

    8 Troubleshooting (2) Sample barcode reader error Error Error message Error condition Troubleshooting code W5330 Barcode reader not The sample barcode Connect the sample barcode reader to connected reader is not connected. the USB connector, and then switch the analyzer off and on. Check the barcode reader connection.
  • Page 211 8 Troubleshooting Error Error message Error condition Troubleshooting code W0191 Wrong slide lot for A slide different from Load a slide of the specified lot, and calibration the lot specified during then calibrate again. calibration has been A different lot of slide loaded.
  • Page 212 8 Troubleshooting (5) Standard curve configuration error Error Error message Error condition Troubleshooting code E4202 Calibration curve set- The wrong calibrator Check the expiration date and the set- ting ERR (span error) may have been used. ting position of the calibrator, and then calibrate again from the beginning.
  • Page 213: Other Errors

    8 Troubleshooting 8.2.15 Other Errors Error Error message Error condition Troubleshooting code W0020 No disposal box Measurement was Insert the disposal box completely, and started with the dis- then start measurement. Put Disposal box to posal box pulled out. the end and then press START key.
  • Page 214 8 Troubleshooting Error Error message Error condition Troubleshooting code I5920 Turn the power ON A long time has passed Switch the analyzer off and on. since start-up. 30 days or more have passed from when the analyzer was started. For stable system performance, turn the power ON.
  • Page 215: Slide Code Table

    8 Troubleshooting Slide code table Slide code Slide code Test Test Test code Sample code Test code Sample code GOT/AST GPT/ALT TCHO CKMB TBIL AMYL DBIL HDLC NOTE Slide codes are subject to change without notice. Check the slide code printed on the slide box. 897N201058J FDC NX700 Instruction Manual 2022.04 8-45...
  • Page 217: Function Menu

    Function menu If the (Function menu) button is tapped, the Function menu screen appears. Function menu list Function menu items Settings Advanced settings Reference section Calibration — — 4.1.1 Control — — START — — 4.3.1 Manual pipetting Date & Time settings —...
  • Page 218 9 Function menu Function menu items Settings Advanced settings Reference section Language setting — 9.3.1 Host connection setting None 9.3.2 Serial connection LAN connection Calculated Parameter — 9.3.3 Unit conversion — 9.3.4 Switch measurement range dis- 9.3.5 — play Reference interval range setting —...
  • Page 219: Control

    9 Function menu Control This mode is used for measurement of control serum (FUJI DRI-CHEM CONTROL QP) and for per- forming control surveys. When control mode is set, measurements are performed under the condi- tions listed below. • Correlation coefficients: follow [Control measurement setting] (Initial values disabled (a=1, b=0)) of the function menu.
  • Page 220 9 Function menu 5. Tap the sample rack display to display the sample information registration screen in control mode. Enter or select the sample information, and then tap the [OK] button ( 7 ). 1 Enter the sample number using the software key- board (Section 2.3).
  • Page 221 9 Function menu 6. Close the sample cover, and then lock it with the sampler cover lock key. WARNING: Ensure the sampler cover is locked during sample processing to prevent injuries and infection. When the display reads [Ready to test], the sam- pler cover can be opened.
  • Page 222 9 Function menu 8. After measurement, tap the button to check the measurement results. IMPORTANT: After using control mode, always tap the [HOME] button to clear control mode. <Display of measurement results> Green: when within the acceptable range of control Red: when outside of the acceptable range of control Acceptable lower Acceptable upper...
  • Page 223: Normal Mode

    9 Function menu Normal mode General operators use this function menu to, for example, set the time and configure the brightness of the LCD. Settings Advanced settings Reference section Date & Time settings — 9.2.1 Brightness • Volume Display brightness 9.2.2 Error sound Other general sound...
  • Page 224: Date & Time Settings

    9 Function menu 9.2.1 Date & Time settings This mode is used to set the date and time settings stored in the memory. • Tap “ ” or “ ” to change the date and time values. • Configure the order of the year, month, and date by using [Date format setting].
  • Page 225: Lamp Management

    9 Function menu 9.2.3 Lamp management Display and reset the cumulative illumination time of the lamp currently installed in the analyzer. Fur- thermore, choose whether to turn off the lamp automatically when the analyzer has not been used for some time. •...
  • Page 226: Mode Function For Administrator

    9 Function menu Mode function for administrator NOTE To display Mode function for administrator, open the Function menu page, tap the [Mode function for administrator] button, and then enter the password. NOTE In order to prevent other people from oper- ating the analyzer, the input dialog for the password will appear when no operation is performed for a fixed period of time during...
  • Page 227: Language Setting

    9 Function menu 9.3.1 Language setting Configure the language to be used. Tap the button of language to be configured. NOTE Rebooting the device is necessary for the setting to take effect. The confirmation di- alog for the power-off will appear when is tapped to exit Mode function for adminis- trator.
  • Page 228 9 Function menu • If [Serial connection] is set, configure the com- munication type and whether to output from the printer. You can choose from 4 communication types. - Type 1: Bidirectional connection - Type 2: Unidirectional connection (bidirectional format) - Type 3: Unidirectional connection (standard format) NOTE...
  • Page 229 9 Function menu If the button is also tapped when Type 4 was selected, the settings screen of Type 4 of page 3 becomes active. • Tap [Examination code] to select the standard for examination item classification code to be used in communication.
  • Page 230: Calculated Parameter

    9 Function menu 9.3.3 Calculated Parameter You can configure the settings to calculate the following items from simultaneously measured values and then display the results. Items registered for calculation have their measurement results dis- played and printed. Test name Calculation Unit Calculation formula test name...
  • Page 231 9 Function menu NOTE Even when registering one sample information to multiple sample racks (A to E) for testing, be sure to load the slides used for calculation items in the same slide cartridge to perform the measurements. If the [Calculated Parameter] button is tapped in Mode function for administrator, precautions for use of calculation items are displayed.
  • Page 232: Unit Conversion

    9 Function menu 9.3.4 Unit conversion Configure the main unit and secondary unit for each measurement item. The following table shows the units that can be used for each configuration item and the conversion coefficient Test Test Unit (A) Unit (B) Conversion Test Test Unit (A) Unit (B) Conversion...
  • Page 233: Switch Measurement Range Display

    9 Function menu 9.3.5 Switch measurement range display Switch the display method for data outside the measurement range (passing the upper/lower limit of the determination range) for all tests. • If [@] is configured, an [@] symbol is added to indicate the displayed measurement value is outside the measurement range.
  • Page 234: Reference Interval Range Setting

    9 Function menu 9.3.6 Reference interval range setting You can add, edit, and change the order of the reference interval names that can be selected when registering sample information. (Refer to Section 3.3 (4)) Furthermore, you can set a reference inter- val for each sample type ([P/S, W] or [U]) of a reference interval name.
  • Page 235 9 Function menu Changing the display order Tap the button of the test for which you want to change the display order. Next-page buttons Tap the field of the test for which you want to con- figure the transfer destination, and then tap the [OK] button.
  • Page 236: Sample No. /Id Settings

    9 Function menu 9.3.7 Sample No. /ID settings • Configure whether to display the sample number in the measurement results under [Edit and print No.]. • Configure whether to record the sequence num- ber of the analyzer under [Store No.]. 9.3.8 Work list selection setting Configure whether to display the sample ID or pa- tient ID at the Work list index screen.
  • Page 237: Correlation Coefficients

    9 Function menu 9.3.9 Correlation coefficients Enter, reset, or print the correlation coefficients for each sample type ([P/S,W] or [U]). For details, refer to “Description of the correlation function” at the end of the document. IMPORTANT: Incorrect entry of “a” and “b” causes incorrect measurement results. Always make sure that the entries (a, b) are correct.
  • Page 238 9 Function menu Tap the button of the measurement item you want to edit. NOTE Only measurement items for which QC cards have been read are displayed. If the [Input] field of an item is tapped, the software keyboard (refer to Section 2.3) appears so you can enter the value.
  • Page 239 9 Function menu IMPORTANT: Description of the correlation function This function is designed to determine the correlation between the measured data obtained using the FDC NX700 and the data obtained using the conventional measuring method with your own instruments. Y=aX+b The X-axis shows the measured data obtained us- ing your instruments, and the Y- axis shows the measured data obtained using the FDC NX700.
  • Page 240 9 Function menu 4. Depending on the values input for (a, b), the apparent limits of the measurement range will shift as follows. (Example) GLU measurements If a=1 and b=0 (Y=X), the measure- ment range is 10 to 600 mg/dl. But by inputting regression coefficients as shown, the lower limit changes to A and the upper limit to B.
  • Page 241: Lot Compensation Coefficients

    9 Function menu 9.3.10 Lot compensation coefficients Enter the lot compensation coefficients (c, d, and e) printed on the QC cards of slides. This mode is needed if it is not possible to read the data directly because the QC card was lost or spoiled. IMPORTANT: Incorrect inputs for (c, d, and e) will cause incorrect measurement results.
  • Page 242 9 Function menu Tap the button of the Lot No. for which you want to enter a lot compensation coefficient. NOTE If the button is tapped, all lot compen- sation coefficient settings are printed. If the [Input] field of an item is tapped, the software keyboard (refer to Section 2.3) appears so you can enter the value.
  • Page 243: Standard Curve Coefficients

    9 Function menu 9.3.11 Standard curve coefficients The calibration curve coefficients (A, B, C, D, E, and F) of the QC card information can be entered. IMPORTANT: Incorrect entry of calibration curve coefficients causes incorrect measurement results. Always make sure that the entered coefficients are correct. NOTE The calibration curve coefficients of this setting are entered automatically by QC card read- ing and calibration.
  • Page 244 9 Function menu 3. If the [Input] field of an item is tapped, the software keyboard appears so you can enter the value. NOTE If the button is tapped, the calibra- tion curve coefficients are printed. 4. First tap the [Input] field of [Lot]. The Lot number entry screen (software key- board screen) appears.
  • Page 245 9 Function menu <Configuration of calibration curve coefficients (A, B, C, D, E, and F)> 1. Enter the calibration curve coefficient (A:), and then tap the [OK] button. The Calibration curve coefficient (B:) entry screen (software keyboard screen) appears. 2. Enter B through F using the same proce- dure. <Configuring the expiration date for use> For the expiration date for use (EXP), enter xxxx- xx (four digits for the year and two digits for the month).
  • Page 246: Lot Qualification

    9 Function menu 9.3.12 Lot qualification This function button appears (enabled) when USM function is enabled. It is possible to (dis-)qualify slide LOTs which are registered in analyzer. 1. Select a slide name which needs to change the setting. 2. (Dis-)qualification for selected slide LOT can be changed by selecting [ON] or [OFF] button.
  • Page 247: Pf Calibration Coefficients

    9 Function menu 9.3.13 PF calibration coefficients Configure the PF calculation correction coefficients. Next-page buttons Tap the [Data input] button. NOTE I f t h e b u t t o n i s t a p p e d , a l l P F calculation correction coefficients are printed.
  • Page 248 9 Function menu If the [Input] field of an item is tapped, the software keyboard (refer to Section 2.3) appears so you can enter the value. NOTE If the [Reset] button is tapped, a reset con- firmation screen appears. To reset the PF calculation correction coefficients of select- ed items, tap the [OK] button.
  • Page 249: Dilution Setting

    9 Function menu 9.3.14 Dilution setting You can configure the dilution factor and position of the diluent to be used for each measurement item and sample type (PS and U). When a dilution factor is not configured for each sample on the sample information registration screen, measurement is performed using the dilution factor config- ured here.
  • Page 250 9 Function menu 3. For the sample type (liquid type), tap a button, and then select from [D1], [D2], or [D3]. 4. For the dilution factor, tap a button, and then select from [None], [2], [3], [4], [5], or [10]. 9-34 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 251 9 Function menu NOTE In the case of CRP, the selection menu to the left appears. (Supplement) Default diluent positions Diluent position for sample type [PS] Diluent position for sample type [U] Purified Saline Purified Saline Diluent water solution Diluent water solution TCHO...
  • Page 252: Tube Setting For Diluent & Reference Fluid

    9 Function menu 9.3.15 Tube setting for diluent & reference fluid Configure the test tube type to be used for diluent and reference fluid. NOTE For the types of containers that can be used for diluent and reference fluid, refer to Section 5.4 and Section 5.5. Tap the button of the container you want to use. NOTE [D1] through [D3] under [Diluent] indicate position of the diluent holder.
  • Page 253: Control Fluid Setting

    9 Function menu 9.3.17 Control fluid setting Configure the control liquid name / lot number and acceptable range. 1. Tap the button of the control liquid to be edited. NOTE [-] (control liquid not registered) of Step 1 cannot be edited. 2. Tap a [Lot] button to choose a lot. 3. If you touch a display section (lower limit or upper limit) of the acceptable range to be edited, a software keyboard (Refer to Section 2.3) appears so you can edit the...
  • Page 254: Control Measurement Setting

    9 Function menu 9.3.18 Control measurement setting Configure whether to enable the correlation coefficient when performing a control measurement. To enable the correlation coefficient for a control measurement, tap the checkbox to select it. NOTE By default, the checkbox is cleared (not enabled).
  • Page 255: Test Item Name Setting

    9 Function menu 9.3.20 Test item name setting You can configure the display of item names. The names appear in the screen display and printed paper of measurement results. 1 Choose the item names to be used for display and printing. 2 Choose the names to be used for external com- munication output to the host computer.
  • Page 256: Password Setting

    9 Function menu 9.3.21 Password setting When tapping the [Mode function for administrator] button on the Function menu screen, the pass- word input may be required. The password can be changed here. If the input field is tapped, the software keyboard (refer to Section 2.3) appears so you can edit the password.
  • Page 257: Other Functions

    Other functions 10.1 Data communication The analyzer can transmit measurement results to a host computer or PC that has been approved by IEC/UL60950-1 or IEC/UL62368-1. To communicate, it is necessary to prepare the software and a cable to receive data. Before you use this function for the first time, please contact your dealer for assistance.
  • Page 259: Usm (User Safety Mode)

    USM (User Safety Mode) 11.1 Operator ID input when measurement start and in mode function When measurement start button or function menu button is tapped, an Operator ID input display is shown. Then, input Operator ID. If input operator ID is registered and also within expiry, the meas- urement starts or the function menu becomes active.
  • Page 260: Difference Of Operable Functions Among Operator Id Level (Ope,Adm,Emg)

    11 USM (User Safety Mode) 11.2 Difference of operable functions among operator ID level (OPE,ADM,EMG) Operable functions are different depending on operator ID level. NOTE OPE: Operator ADM: Administrator EMG: Emergency Measurement Result OPE/ADM: When expired slide is used, EXP is indi- cated instead of measurement result. EXP is used not only on screen indica- tion, but also on print and in communi- cation.
  • Page 261 11 USM (User Safety Mode) Mode function OPE/EMG: Operable function is limited to [Control], [Calibration], [START Manual pipetting] and some maintenance functions. 897N201058J FDC NX700 Instruction Manual 2022.04 11-3...
  • Page 262 11 USM (User Safety Mode) ADM: All functions are operable. 11-4 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 263: Usm Menu

    Rename the file which is created by the tool to [dcn_uid.dat], Store it in the USB, WData └ GUI └ dcn_uid.dat *Please use FUJIFILM recommended USB only. 3. Select [Operator ID setting] in ADM mode. 4. Select [Operator ID import]. 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 264 11 USM (User Safety Mode) 5. The left pop-up is shown. Then set USB, and after a few seconds, select [OK]. 6. The left pop-up will be shown when im- port is complete. Take off the USB and then tap [Shutdown]. 11-6 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 265: Operator Id -Export

    11 USM (User Safety Mode) 11.3.2 Operator ID -Export 1. Select [Operator ID setting] in ADM mode. 2. Select [Operator ID export]. 3. The left pop-up is shown. Then set USB, and after a few seconds, select [OK]. 897N201058J FDC NX700 Instruction Manual 2022.04 11-7...
  • Page 266 11 USM (User Safety Mode) 4. The left pop-up will be shown when ex- port is complete. Then tap [OK], and take off the USB. 5. Operator ID data, [dcn_uid.dat], is stored in below folder structure below in the USB. WData └ GUI └ dcn_uid.dat * Browsing and editing of dcn_uid.dat is possible in Operator ID Editor.
  • Page 267: Editing Registered Operator Id

    11 USM (User Safety Mode) 11.3.3 Editing registered operator ID Registered Operator ID can be edited as the following procedure. NOTE Editable items are the expiry of ID and the level. 1. Select [Operator ID setting] in Mode function for administrator screen. 2.
  • Page 268 11 USM (User Safety Mode) 4. To change the expiry, touch the Expira- tion field and edit it using the software keyboard. 5. To change the level, select either [ADM/ OPE]. 11-10 897N201058J FDC NX700 Instruction Manual 2022.04...
  • Page 269: Editing Emergency Id

    11 USM (User Safety Mode) 11.3.4 Editing emergency ID (Dis-)qualification of registered Emergency ID can be set. Also measurement limit count by use of Emergency ID can be set. 1. Select [Operator ID setting] in Mode function for administrator screen. 2.
  • Page 270 11 USM (User Safety Mode) 4. To edit the total limit count, select [Edit emergency ID count]. 5. To reset the current count, tap [Reset] button. To set a new limit count, select either [0], [100] or [200]. In case of 0, emergency ID does not function.
  • Page 271: Specifications

    Specifications Processing speed 190 tests per hour (simultaneous CM and ISE measurements) Number of incubator cells CM: 13; ISE: 1 Incubation temperature 37°C (CM) / 30°C (ISE) Incubation time 1 to 6 minutes Capacity for simultaneously loaded slides Slide lot information Each item: maximum 5 lots Sampler (1) Number of samples...
  • Page 273: Glossary

    Glossary Glossary of the display & printout messages NOTE Because the display spaces and printouts are limited, some abbreviations are used. NOTE Periods will not be used after abbreviations on the display and printout messages. Abbreviation Meaning Analog/Digital administrator average Bef comp Before compress cali...
  • Page 276 26-30, NISHIAZABU 2-CHOME, MINATO-KU, TOKYO 106-8620, JAPAN https://www.fujifilm.com/ Balcke-Duerr-Allee 6, 40882 Ratingen, Germany Oudenstaart 1, 5047 TK Tilburg, The Netherlands...

This manual is also suitable for:

Fuji dri-chem nx700i

Table of Contents

Save PDF