Refer to the following if your RapidVap Vertex Evaporator fails to operate
properly. If the suggested corrective actions do not solve your problem,
contact Labconco for additional assistance.
PROBLEM
CAUSE
Unit will not operate
Unit not connected to
electrical power
Circuit breaker blown
Lid open
Vent hose exhausting into lab
Sample odor in lab
area
Inoperable blower
Gas pressure too high
Solvent in bottom of
chamber
Poor recovery rates
Sample going to dryness
Sample splashing out of tube
Heat setting is too high
22
6
6
R
R
E
S
H
O
O
T
I
E
S
H
O
O
T
I
Product Service 1-800-522-7658
N
G
N
G
CORRECTIVE ACTION
Connect unit to proper electrical
receptacle.
Correct electrical problem and reset
circuit breaker by pressing button.
Close lid.
Redirect hose to fume hood.
Replace blower.
Reduce pressure.
Program time to prevent dryness.
Decrease Nitrogen pressure.
Decrease heat.