Section 3: Calibration
3.1
Calibration
3.1.2
Using The IntelliPoint with Auto-Calibration
The Intellipoint™ model RLL (Standard Sensitivity) and
RPL (High Sensitivity) feature Auto-Cal calibration. Auto-
Calibration is suitable for liquid and slurry applications. For
applications such as granular materials, interface measurement
of two liquids, and plugged chute installations, a Manual
Calibration IntelliPoint is recommended.
If you purchased an "Auto-Calibration" IntelliPoint and have
determined you require a "Manual Calibration" IntelliPoint
based on the Application Guide, please contact the AMETEK
Drexelbrook Service Department.
After the IntelliPoint is installed in the vessel, simply apply
power. The electronic unit will auto calibrate.
Caution – The material being measured must be below the
sensing element when power is applied (Sensing element
uncovered).
Calibration is complete.
For sensing elements with active lengths over 6 feet (2
meters) a manual calibration mode (#8 standard sensitivity)
is recommended.