Sterilmatic - Digital Electric Sterilizer

Service Connections

DIMENSIONS ARE IN INCHES [MM]
DIMENSIONS ARE IN INCHES [MM]
18.69 [475]
9.34 [237]
B
3.5
[89]
D
B1
15.5
SEE
[394]
NOTE 2
A
C
1.5 [38]
SYMBOL
A
Drain - 1/2" (13mm) N.P.T. of 3/8" (10mm) OD copper (see note 1)
B
Steam Exhaust Connection - 5/8" (16mm) (see note 2)
C
Electrical Connection - (*see electrical specifications)
D
Power Supply
The manufacturer reserves the right to modify materials and specifications without notice.
Unit must be grounded. Main supply voltage fluctuations are not to exceel ± 10% nominal supply voltage.
Clearance from combustible and non-combustible construction should be a minimum of 4" in the back and 2" on the sides.
Good quality water feed is the responsibility of the owner. Water quality must be within the following general guidelines.
TDS: 40-125 ppm
Hardness: 35-100 ppm
The best defense against poor water quality is a water treatment system designed to meet your water quality conditions.
Appliance to be installed with backflow protection according to federal, state or local codes.
IMPORTANT: Exhaust line must be vented to the outside to eliminate the exhaust steam and the accompanying noise from entering the room. Use
1/2" (13mm) copper tubing or suitable alternate. The overall height and length of the line should not rise more then 4' (1.2 meters) above the unit and
exceed 15' (4.5 meters) with a minimum of bends. The line should slope downward after leaving the sterilizer in order to ensure condensate drainage.
IMPORTANT: Failure to comply with this outline will affect the sterilization process.
When an exhaust condenser is supplied; the following services must be provided: 1/2" (13mm) cold water: 1" (25mm) waste: 115V electrical line.
Terry System Cartridge Changes / Installation – "2-3 gallons of water MUST be purged at each cartridge change or new installation prior to
water supply being fed to the steamer. Failure to do so can result in component damage within the steamer which is not covered under warranty.
For additional guidance on proper installation, refer to install documentation provided with each Terry System and Replacement Cartridge Set."
As continued product improvement is a policy of Market Forge, specifications are subject to changge without notice.
INSTALLATION & OPERATION MANUAL 14-00413 REV 0 (08/21)
STEAM SHIELD
3 [76]
5.63
[143]
4.75
[121]
C
1 [25]
6 [152]
Customer provided Drain Catch Pan
Approx. 4-6 quarts (3.8-5.7 liters)
SERVICE CONNECTIONS
DESCRIPTION
ELECTRICAL CHARACTERISTICS
Model
kW
Hz
208V
9.3
60
45A
STM-ED
12.4
60
-
10.4
50
-
STM-EDX
12.4
50
-
WATER SUPPLY AND DRAIN SPECIFICATIONS
pH: 7.0 - 8.5
B
STEAM SHIELD
SEE NOTE 2
DOOR IN
OPEN POSITION
B1
A
4 [102]
ADJ. LEGS
1. An air break must be provided if a unit drain line is run.
2. Vent exhaust to atmosphere. B1 is actual connection,
but must exit casing at B.
3
Heat of Rejection (Heat Loss Into the Room) 6,000 BTU'S.
4. For use at altitudes up to 6500ft (200m).
1 Phase
3 Phase
220V
240V
208V
240V
220/380V
-
-
26A
-
-
52A
-
30A
48A
-
-
-
-
52A
-
-
Silica: <13 ppm
Chlorides: <25 ppm
DISCLAIMER
Service Connections
21 [533] CHAMBER
9 [230]
12 [305]
11 [280]
CLEARANCE
9 [228]
CLEARANCE
CONDUIT
Locate electric box as close to
C
sterilizer as possible and per
local electrical codes
NOTES
240/415V
-
-
-
-
16A
-
-
18A
Chlorine: <0.2 ppm
Chloramine: <0.2 ppm
3 [76]
D
5
PAGE
OF 24