2.16.2 Additional Accessories..........................27 TECHNICAL PARAMETERS ..........................28 Technical parameters BTL-2000......................28 Technical parameters ADAPTER BTL-282 ....................30 USER’S GUIDE ..............................31 Introduction .............................. 31 Light; the principle of laser light and its specifics ..................31 Types of lasers............................32 4.3.1...
Page 3
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E Contradictions of non-invasive laser therapy ................... 35 LIST OF THERAPIES............................36 Dentistry (Medical Field 03) ........................
BTL-2000 is a portable therapeutic laser that employs red and infrared BTL laser probes of various powers. BTL-2000 is designed for easy use and is understandable even for those who are new to laser technology. The laser beam irradiation strength is easy to regulate. Parameters can easily be changed as desired and your own programmes can be created and saved.
Basic Operation Procedures BTL 2000 is controlled by means of an interactive menu. Press the [ + ] or [ - ] push buttons to move between menu settings. The selected setting is indicated by an arrow in the display’s left margin. The presence of an additional arrow indicates that the menu exceeds four lines when selecting the last setting on the display.
(∼ 0.5 sec.). If the instrument finds an internal problem, you will see the error notice “Error“ and the instrument will then be locked in Safe Mode. In this case, it is necessary to contact your BTL service representative. If the error notification “Error 100 –...
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E Using the [ + ] and [ - ] arrows, insert the safety code. Use the [TAB] push button to move cursor to the next position. Once the code has been entered, press the [ENTER] push button.
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E Enter the first letter of desired diagnosis name and press the [ENTER] push button to enter the list of diagnoses. Use the [ + ] and [ - ] arrows to move between diagnoses names and press the [ENTER] push button to select desired diagnosis.
2.5.4 Frequency When the frequency is adjusted, the instrument automatically recalculates therapy time and power parameters. BTL-2000 offers pre-programmed EAV and Nogier frequencies. 2.5.5 Nogier (Frequencies according to Dr. Nogier) After entering Nogier freqencies, the list of frequencies will be displayed. Use [ + ] and [ - ] arrows to move between the frequencies.
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E 2.5.6.1 Table of EAV frequencies EAV (according to Dr.Voll) Frequency [Hz] Use and Application Area 1.20...
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E User menu (push button [MENU]) When pressing the [MENU] push button, the instrument will enter the System Menu and it is then possible to select from the following list.
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E 2.8.2 Patient List When entering Patient List, the following list will be displayed: >...
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E D e n s . _ 0 . 9 J / c m F r e q u e n c y C o n t i n u a l A r e a...
Page 16
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E 2.8.3.1 New Diagnosis When creating a new diagnosis, enter New Diagnosis by using the [ENTER] push button. After entering New Diagnosis the following menu will be displayed: T h e r a p y n a m e :...
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E 2.8.3.3 Delete Diagnosis If there are no diagnoses in memory, “Memory free” will be shown on display. If there are less than 6 diagnoses in memory, a list of these patient files in alphabetical order will appear on the display.
Page 18
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E 2.8.4.1 Entry Code Change The manufacture code is preset as 0000. It is possible to change the code by entering Entry Code Change. After entering Entry Code Change, the following menu will be displayed: B T L - 2 0 0 0 T E R A P E U T I C...
Page 19
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E 2.8.4.3.1 Comparison of Two Different Duty Factors After entering the DF change setting, the following menu will be displayed: D u t y f a c t o r Use the [ + ] and [ - ] arrows to adjust Duty Factor.
Page 20
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E Main menu Diagnoses – Selection according to branches and diagnoses Program –...
2.8.5 Acupuncture The BTL-2000 instrument offers two methods for locating acupuncture points: Locating acupuncture points using probe with acupuncture attachment (not included in standard accesories): Skin resistance is measured by an appropriate course of tension 0.5 V (at 600 kΩ). An acupuncture point can be recognized by a local decrease in skin resistance and can be seen in % on the bar-graph and is accompanied by an accoustic signal.
2.8.6 PROBE TEST: Optical Power of Probe This function measures probe output and the status of the probe diode. BTL-2000 automatically tests the probe performance and corrects the power for optimal output (only for the BTL probes). The warning messages ATTENTION and LASER IRRADIATION will be displayed followed by the message PROBE TEST.
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E Therapy 2.9.1 Display during therapy The instrument continuously checks for laser probe connection. When a probe is connected the parameters of laser therapy will be displayed.
The instrument can be charged overnight without any danger. Use only the original adapter, BTL-282, supplied by the manufacturer. Accumulators used are NiCd and it is necessary discharge them occasionally (once in 14 days or once in month, depending on how often the instrument is used).
The producer of the instrument is not responsible for any damages caused by incorrect voltage conversion. Probe durability During use period there is aging of laser diode. BTL-2000 measures and displays the laser power on the display. The instrument automatically adjusts to maximum power according to the state of the diode. When the power decreases, due to probe malfunction or diode aging, to less than 40% total power, we recommend replacement of the probe.
ISO 10993-1, ISO 10993-3, ISO 10993-5. • The user is responsible for the use of materials and accessories that are not supplied by BTL. This includes the cleaning and disinfecting liquids. •...
This warranty may differ from the warranty terms and conditions provided by your supplier and by applicable laws in your country. 2.16 Accessories Do not connect non BTL accessories to the instrument. 2.16.1 Standard Accessories Adapter with mains power cable (BTL-282) BTL-2000 –...
- Atmospheric pressure 650 hPa to 1100 hPa - Position - Maximum Duration 1 year Power supply - external Adapter BTL-282 (see 3.2 Technical parameters ADAPTER BTL-282) Battery Supply - battery type NiCd accumulator - number - total voltage 9.6 V...
Page 29
- In Compliance with ČSN EN 60825-1 ČSN EN 60601-2-22 Laser Modes - Continual - Pulse BTL is Designed to Comply with ČSN EN 60601-1 Appendix ČSN EN 60601-1 A2 Appendix ČSN EN 60601-1 A12 ČSN EN 33 4200 ČSN EN 60825-1 Appendix ČSN EN 60825-1 A11...
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E Technical parameters ADAPTER BTL-282 General Parameters 0,3 kg ± 10 % (0,2 kg without supply cable) - Weight –...
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E USER’S GUIDE Introduction This guide is intended to make you familiar with performing non-invasive laser beam therapy. At the same time, it will give you general instructions on how to perform non-invasive laser beam therapy in clinical practice.
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E Coherent Non-coherent A laser beam of a high intensity therefore represents the “death ray” of science fiction authors; a laser beam can destroy material or matter.
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E Visible spectrum 500nm 700nm 400nm 600nm Ultraviolet radiation, Infra-red radiation, X-rays, gamma...
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E All of these effects are based on objective studies that have been validated and verified. •...
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E Programme Density Freq. Comment number [J/cm [Hz] Lichen planus Rosacea Sanatio postoperativa Striae...
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E Programme Density Freq. Comment number [J/cm [Hz] Rhinitis vasomotorica Sanatio postoperativa Sanatio Posttrepanationis...
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E Programme Density Freq. Comment number [J/cm [Hz] Sy. Tietz (1) Cont part of therapy Sy.
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E Programme Density Freq. Comment number [J/cm [Hz] Decubitus Dermatitis Eczema Furunculus...
Page 42
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E APPENDIX A Direction of Laser Beam Direction of Laser Beam Power Cord Laser Control Probe Laser Aperature...
Page 43
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E APPENDIX B Maximum outputs: 60mW/685nm, 400mW/830nm Laser classified by: IEC 601-2-22:1995 IEC 60825-1:1993/A2:2001 VISIBLE AND INVISIBLE...
Page 44
B T L - 2 0 0 0 U S E R ' S M A N U A L & U S E R ’ S G U I D E LASER RADIATION CLASS III. B Label 3 – Laser Radiation Class III warning label: use in laser therapy room. Page 44 z 45...
Page 45
The contents of this document is provided "as is". Except as required by applicable law, no warranties of any kind, either expressed or implied, are made in relation to the accuracy, reliability or contents of this document. BTL Industries Limited reserves the right to revise this document or withdraw it at any time without prior notice.