Getinge 86 Series Installation Manual
Getinge 86 Series Installation Manual

Getinge 86 Series Installation Manual

Washer disinfectors
Hide thumbs Also See for 86 Series:

Advertisement

GETINGE 86-SERIES
INSTALLATION MANUAL
502606400
SEV0725001-

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 86 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Getinge 86 Series

  • Page 1 GETINGE 86-SERIES INSTALLATION MANUAL 502606400 SEV0725001-...
  • Page 2: Table Of Contents

    Contents FOREWORD ____________________________________________3 SAFETY REGULATIONS _________________________________4 General safety regulations ___________________________________ 4 The machine must be assembled and installed: ________________ 4 When the work is complete, check that _______________________ 5 Product liability ____________________________________________ 5 Attention symbols __________________________________________ 5 INSTALLATION __________________________________________6 Wall-mounted model ________________________________________ 7 Connecting electric power ___________________________________ 8 Alternative connection arrangements _________________________ 9 Connecting water, steam, waste and dryer ____________________ 10...
  • Page 3: Foreword

    • Installation manual (this book) • Technical manual • Spare parts list Getinge reserves the right to change the specification and design without prior notice. The information in this manual was up to date on the date of issue of the manual. © Copyright The content of this manual must not be copied, in whole or in part, without the written consent of Getinge.
  • Page 4: Safety Regulations

    Observe safety measures for steam and hot water. • The electrical cabinet may only be opened by authorized and trained personnel. • Spare parts must be obtained only via Getinge EDC. The machine must be assembled and installed: • by qualified personnel.
  • Page 5: When The Work Is Complete, Check That

    Qualification according to ISO 15883 before putting the machine into service. Incorrect use may result in damage and injury. Product liability Any modification or incorrect use of the equipment without the approval of Getinge Dis- infection AB invalidates Getinge Disinfection AB´s product liability. This product was manufactured by:...
  • Page 6: Installation

    INSTALLATION • The floor where the machine is to stand must be flat and level within ±5 mm. • Pull the machine (on a transport pallet) to the location where it will be installed. • Lift the machine off the pallet. •...
  • Page 7: Wall-Mounted Model

    Wall-mounted model • If the machine has double doors and is to be mounted in a wall, the distance be- tween machine and wall must be 15 mm. • If machines are to be installed next to each other they must be adjusted to the same height on installation.
  • Page 8: Connecting Electric Power

    Connecting electric power Installation may only be done by authorized personnel. Connect electric power to the machine as follows: • Connect the power supply to the electrical connections of the machine; see the dia- gram with alternative connection arrangements. • For easier servicing and maintenance, the supply to the machine should be fitted with a lockable isolator with a 3 mm break gap.
  • Page 9: Alternative Connection Arrangements

    Alternative connection arrangements MAIN SWITCH MAIN SWITCH -Q01 -Q01 HUVUDBRYTARE HUVUDBRYTARE HAUPTSCHALTER HAUPTSCHALTER INTERRUPTEUR ELECTRIQUE INTERRUPTEUR ELECTRIQUE JB01 JB01 ELECTRICAL CONNECTION ELECTRICAL CONNECTION EL-ANSLUTNING EL-ANSLUTNING ELEKTROANSCHLUSS ELEKTROANSCHLUSS CONNEXION ELECTRIQUE CONNEXION ELECTRIQUE Electrically heated 40/415V 3N+PE, 50Hz 1.1kW 31A FUSE 3x3A 40V 3+PE, 0Hz 1.3kW 54A FUSE 3x0A 30/400V 3N+PE, 50/0Hz 19.7kW 30A FUSE 3x3A 30V 3+PE, 50Hz 19.7kW 54A FUSE 3x3A...
  • Page 10: Connecting Water, Steam, Waste And Dryer

    Connecting water, steam, waste and dryer Installation may only be done by authorized personnel. • Provide water and, if required, steam connections with separate stopcocks. Flush out the water and steam pipes that are to be connected to the machine, to prevent clogging of filters and valves.
  • Page 11 Chamber depth 720 mm Clean side Soiled side Chamber depth 800 mm Clean side Soiled side Figure 5. Connecting water, steam, waste and dryer 1. Cold water 350 mm above the floor 7. Drain connection at floor 2. Hot water 350 mm above the floor 8.
  • Page 12 • PTFE tape is recommended for sealing the connections. • Connect the disinfector to a waste outlet with a capacity of at least 50 litres/min. Connect the waste as shown in Figure 5. Diameter of waste: 50 mm. V1474 Figure 6. Connecting a dryer •...
  • Page 13: Function Check

    Function check • Check that the machine is connected to the correct voltage and the waste, water, steam and condensate connections are correctly connected. Open the valves for water and steam. • Fill the detergent containers and place alarm sensors in the containers. When starting a new machine, the door can be unlocked and opened with U.
  • Page 14 • On a machine with manual double doors, check the door locking function when the program is complete as follows: - If the soiled-side door cannot be opened, check that the clean-side door is closed. - Open and close the clean side door. - Press W.
  • Page 15: Technical Data

    TECHNICAL DATA Dimensions Outside dimensions Width 1110 mm Depth 8666, CM303: 910 mm 8668, CM304: 990 mm Height 1870 mm Chamber, effective dimension Chamber depth 8666, CM303 720 mm 8668, CM304 800 mm Chamber width 665 mm Chamber height 667 mm Chamber volume 8666, CM303 300 litres...
  • Page 16 Distilled/de-ionized water (option) Recommended water quality: The maximum conductivity is determined by the desired washing result. Max temperature 60°C (140°F) Note 2. Connection ISO G-3/4 Pressure 100-800 kPa (where the pressure is lower than 100 kPa, a separate pressure booster pump must be connected. This is available as an option.) Flow rate 30 liters/min...
  • Page 17 Other points Circulation system Design pressure: 200 kPa Working pressure: max 130 kPa Design temperature 160 °C (320 °F) Working temperature 93°C (199°F) Electrical connection See the relevant connection option. Environmental requirements Air humidity max 80% at 31 °C (87 °F) Room temperature 5-40 °C (41-104 °F ) Heat dissipation to the room...
  • Page 18: Water Quality - Washer Disinfectors

    High concentrations of and high exposure to sanitary chemicals may cause corrosion and pitting on stainless steel. Getinge Disinfection AB therefore recommends that water used in the pre-rinsing, wash- ing and final rinsing phases should be of drinkable quality in accordance with the guide- lines.
  • Page 19 Page 19 of 0...
  • Page 20 Australia France South Africa Getinge Australia Pty Ltd Getinge France SAS Getinge South Africa (Pty) Ltd 154 Lytton Road BP 49, avenue du Canada P O Box 4849 Bulimba Qld 4171 ZA de Courtaboeuf Hercules info@getinge.com.au Les Ulis, FR-9194 Pretoria SA 000 getinge.france@getinge.fr...

This manual is also suitable for:

502606400

Table of Contents