FAQs
Can the 2595KL be calibrated and if so, how much weight is required?
Yes, the scale can be calibrated following the procedure below and 200 pounds of certified weights
are required. Note: Only weights certified and traceable to national standards are to be used
for calibration procedures.
Press and hold the REWEIGH and ZERO buttons at the same time,
while pressing the ON/OFF button to turn on the scale. "LOAD" will
briefly appear on the display, followed by "200.0 LB" indicates you
can calibrate. Release all three buttons.
Ensure there is no weight on the scale and press the REWEIGH but-
ton. "Clear" will flash on the display, followed by "--------". At this
time the scale is calculating zero calibration. After zero calibration,
the display will flash "PUT" followed by "200.0 LB".
Place 200 lb on the scale. Then press REWEIGH. "CAL" will appear
on the display, followed by "FAC" followed by 5 digits. The calibra-
tion factor should be in the range between 8000 and 10000. If the
number is out of that range, contact Customer Service at 1-800-815-
6615.
If the calibration factor is within range, press the REWEIGH button.
"DONE" will appear on the display, followed by "START", followed by
"0.0". Remove the weight from the scale.
Calibration procedure is complete.
What parts are replaceable on the 2595KL?
Power adapter
Battery Holder/Door
Power adapter holder
Foot for chair leg
Pipe plug for chair leg
Pipe plug for foot rest
Foot for foot rest
Locking hardware for wheel
Health o meter
Professional Scales
®
9500 West 55th Street USA
McCook, IL 60525-7110
Phone: 1-800-815-6615
Fax: 1-800-654-7330
ACTION
Description
Keypad
Main board
Wheels
Foot mat
Seat
Arm rest
www.homscales.com
"START", followed by "0.0".
Part #
ADPT30
2595KLKEYPAD
2595BATTHOLD
B3817001-0
B2138803-0
B2266801-0. Note: the hardware does not come
with it.
B3810002-0
B401280
B401289
B3819701-0
B3438901-2
B401296
B401294
2595KLCASTER
DISPLAY SHOWS
"LOAD" followed by
"200.0 LB"
"Clear" followed by
"-------". Then "PUT"
followed by "200.0 LB"
"CAL" followed by "FAC"
followed by "XXXXX".
"DONE" followed by
2595KL_FAQs_20210114