3. Installation
3.1 Delivery contents
Check contents of carton using this list
1x VACUETTE
1x 12-volt power supply
1x Power cable with Schuko plug
1x Certificate of Quality
Optional – not included in the packaging
836578
Rack for VACUETTE
836579
Rack for VACUETTE
3.2 Positioning of the analyzer
®
The VACUETTE
other vibrating instruments which might cause movement of the bench. The workbench must
be flat and leveled.
Direct light on the instrument and sudden changes in temperature should be avoided.
Do not use this device in close proximity to sources of strong electromagnetic
radiation (e.g. unshielded intentional RF sources), as these may interfere with proper
operation.
This equipment has been designed and tested to CISPR 11 Class A.
In a domestic environment it may cause radio interference, in which case, you may
need to take measures to mitigate the interference.
Therefore the electromagnetic environment should be evaluated prior to operation of
the device.
The instrument is not intended to be integrated into a system.
Do not replace the detachable power cord by an inadequately rated power cord.
Connect the VACUETTE
supply into the
intended power inlet on the VACUETTE
MultiMixer
®
MultiMixer
®
20-Pos. MultiMixer for 9 mm VACUETTE® Tubes
®
20-Pos. MultiMixer for 13 mm VACUETTE® Tubes
MultiMixer must not be placed near centrifuges, oscillating agitators or
®
MultiMixer to the electricity supply by plugging the 12V power
®
MultiMixer. Only use the power supply provided.