How To Use By Remote I/O; How To Use In Serial Mode - Varian Turbo-V 81-AG Instruction Manual

Rack controller
Hide thumbs Also See for Turbo-V 81-AG:
Table of Contents

Advertisement

HOW TO USE BY REMOTE I/O

1.
Power ON the controller
2.
Press the "COUNTERS" and "MEASURES" button
toghether.
3.
Select MODE = REMOTE from the
"INPUT/OUTPUT" menu
4.
Exit from configuration menu
5.
Connection example:
Controller
Imput
Connector
Output
Connector
Fault (+24 VDC)
Start (+24 VDC)
Low Speed (+24 VDC)
Low Speed
R2, R1 (+ 24 VDC)
R2, R1
Pump Power
Monitor (10V=150W)
Monitor (Com)
Pump Frequency
Monitor (10V=81 KRPM)
Open = Stop
Cloed = Start
Start/Stop
1
Low Speed
2
Interlock
3
System Override
4
Soft Start
5
6
7
8
9
4 K7
15
8
Fault
4 K7
13
6
Start
4 K7
12
5
4 K7
14, 11
7, 4
2
1
9

HOW TO USE IN SERIAL MODE

1.
Power on the controller
2.
Connect the serial cable
3.
Select by front panel if you want to work in 232 or
485, and if necessary, the serial address.
4.
Select SERIAL from the INPUT/OUTPUT menu
5.
Open the T+ software (see orderable parts table)
6.
Ready to operate by SERIAL connections.
Open = High Speed
Cloed = Low Speed
Open = Normal Start
Cloed = Soft Start
OFF = OK
ON = Fail
OFF = Stop, Normal Operation
ON = Ramp
OFF = High speed
ON = Low speed setted
OFF = Pump under speed threshold
ON = Pump over speed threshold
Power value
v
v
Speed value
96
TECHNICAL INFORMATION
Closed
= Stop
Open
= Allow Start
87-900-986-01 (C)

Advertisement

Table of Contents
loading

This manual is also suitable for:

969-8988969-8989969-8990

Table of Contents