6.3.5. Setting zero point for low-flow cut-off function
Setup → Sensor Setup → ZP Adjust
Setup → Sensor Setup → ZP Adjust → ZeroPnt
Setup → Sensor Setup → ZP Adjust → CutOff
Setup → Sensor Setup → ZP Adjust → Reset
29
Enter the zero point and the low-flow cut-off point.
If the installed sensor shows a flow rate of > 0 m³/h even if there
is no flow, you can enter a zero point for the characteristic.
To delete the cut-off point, press the "CLR" button.
To return to the previous screen, press the "Back" button.
The low-flow cut-off function is used to set consumption rates
below the entered low-flow cut-off point to 0 m³/h so that they
do not cause the consumption counter value to increase.
To delete the cut-off point, press the "CLR" button.
To return to the previous screen, press the "Back" button.
To reset the entered zero point or the low-flow cut-off point,
press the "Reset" button.
To return to the previous screen, press the "Back" button.