• One (1) Operator’s Manual See the next page for a description of optional accessories. WARRANTY: Drucker Diagnostics warranties that this centrifuge is free from defects in workmanship and parts for 2 years. 200 Shady Lane, Suite 170 – Philipsburg, PA 16866, USA +1-877-231-3115 (U.S.
Specifications General Specifications for the HORIZON 24 Flex Centrifuge Overall Dimensions (H x W x D): ....9˝ x 15˝ x 17˝ Centrifuge Motor: ........1/2 H.P. Brushless DC Timer: ............Electronic, with hold or 0.5 to 99 minutes, +/- 1% Weight: ............
Control Panel / Parts of the Centrifuge Lid Knob Lid Safety Interlock System Control Panel (see below) Cabinet Running Latched Unlocked Up/Down RPM Indicator Indicator Indicator Change Buttons Light Light Light RPM Display Start Button Open / Stop Button Time Display Up/Down Time Change Buttons...
Setup Location 1. Unpack the centrifuge and verify that all of the supplied equipment is present. 2. Choose a setup location which meets the following criteria: a) A bench top clearance height of 21” is required in order to open the lid. b) The clearance envelope is the space around the centrifuge which is required for safety.
Initial Setup Procedure (continued) 10. The ‘RUNNING’ indicator light will illuminate. 11. The buckets will slide up into the horizontal position and the unit will accelerate to the current set speed. 12. Listen to the sound of the centrifuge. A smooth whirring sound should be heard. If there are any loud or unusual sounds, stop the centrifuge by pressing the ‘OPEN / STOP’...
Guide to Balanced Centrifuge Loads One to four devices can be processed in the same cycle. To ensure that the load is balanced, op- posing buckets must be loaded with equally filled devices or a device and a counterbalance. Please refer to device I.F.U.
Control Panel Functions Time Adjustment and Timer Operation: The run time may be set from 30 seconds to 99 minutes and 30 seconds. Press the up and down arrow buttons next to the time display to adjust the run time. Adjustments may be made prior to a run. A quick tap will adjust the time by 30 seconds. Hold down the button for 1 minute adjustments.
Care and Preventative Maintenance With proper care and maintenance your HORIZON centrifuge will provide years of laboratory service. For proper care, the following steps should be taken: 1. Provide Adequate Ventilation: For cooling purposes, the HORIZON draws in ambient air through the air intake cover on the top of the lid and exhausts this air in the back of the base.
Troubleshooting: 1. Problem: The rotor does not spin freely. Solutions: • Make sure that nothing has fallen into the rotor chamber. • If there is nothing obstructing the rotor, the motor may be damaged. Contact your authorized dealer for further assistance. 2.
5. Problem: The time or speed can not be adjusted. Solutions: • With the lid open, press and hold the 'OPEN / STOP' button for 5 seconds. Two beeps and a flash of the lid light will confirm that cycle selection is locked. •...
Contact your authorized dealer to order replacement parts or accessories. INSTRUCTIONS FOR DISPOSAL OF WEEE BY USERS IN THE EUROPEAN UNION This product must not be disposed of with other waste. Instead, it is the user’s responsibility to dispose of their waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment.