The measured values are transmitted to a monitoring software (ELPRO Cloud), which stores and analyzes the data, provides alerts if alarm limits are violated, and generates reports. The system provides superior visibility and transparency in meeting GxP requirements.
1, 2, 3, 5, 10, 15, 20, 30, 60 minutes; user programmable via software (ELPRO Cloud) NOTE: Logging interval will impact battery life Channels 2 independent temperature channels with separately connected sensors Probe type External 4-wire Pt100 probe 1000465V01 DOC - ECOLOG-PRO 2PTR and 1THR - Operation Manual_EN.docx | Page 5 / 15...
Page 6
Green flashing light: Measurement on one or both channels has started. Everything is OK. The interval of the flashing light is 10s. 1000465V01 DOC - ECOLOG-PRO 2PTR and 1THR - Operation Manual_EN.docx | Page 6 / 15...
Page 7
How to connect Make sure that you have defined the sensor's settings in the ELPRO Cloud software and that the "Add sensor" wizard shows the image displayed above. This gives you a 10 minutes time frame to connect your module via the following steps: 1.
Battery life 14 months (provided that the module's radio connection is stable and that the defined temperature range (see below) is maintained) 1000465V01 DOC - ECOLOG-PRO 2PTR and 1THR - Operation Manual_EN.docx | Page 8 / 15...
Page 9
Green flashing light: Measurement on one or both channels has started. Everything is OK. The interval of the flashing light is 10s. 1000465V01 DOC - ECOLOG-PRO 2PTR and 1THR - Operation Manual_EN.docx | Page 9 / 15...
Page 10
How to connect Make sure that you have defined the sensor's settings in the ELPRO Cloud software and that the "Add sensor" wizard shows the image displayed above. This gives you a 10 minutes time frame to connect your module via the following steps: 1.
Page 11
Temperature accuracy 70.0 °C..80.0 °C ±0.2 C 0.0 °C..69.9 °C ±0.1 C -20.0 °C..-0.1 °C ±0.2 C Drift Humidity sensor Typically 1.0 %RH / year 1000465V01 DOC - ECOLOG-PRO 2PTR and 1THR - Operation Manual_EN.docx | Page 11 / 15...
(Cd = cadmium, Hg = mercury, Pb = lead) used. You can dispose of used batteries at collection points in your local community. Please help protect our environment and dispose of batteries properly. 1000465V01 DOC - ECOLOG-PRO 2PTR and 1THR - Operation Manual_EN.docx | Page 12 / 15...
Need help?
Do you have a question about the ECOLOG-PRO 2PTR and is the answer not in the manual?
Questions and answers