HeNe laser beam into the beam receiver and depressing the CALIBRATION button on the front panel. The calibration value is stored in non-volatile memory that is unaffected by power interruption. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 1-1 Burleigh Park, Fishers, NY 14453...
Page 7
GHz, despite the associated low coherence length. The auto-resolution circuitry automatically adjusts the measurement range to be consistent with the coherence length, and the result is displayed with relevant precision. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 1-2 Burleigh Park, Fishers, NY 14453...
If any damage is evident visually or if any loose components can be heard when the instrument is inverted, notify your shipping department and Burleigh Instruments, Inc. It is advisable to save the special carton for future storage or transportation.
Page 9
Figure 2.1. WA-2500 Front Panel. Figure 2.2. WA-2500 Rear Panel. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 2-2 Burleigh Park, Fishers, NY 14453 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
BC-1 optical coupling head. There is a “keyed” position for the connector. Rotate this into position before tightening the knurled lock ring. Figure 3.1. Beam Receiver. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-1 Burleigh Park, Fishers, NY 14453...
NOTE: The beam receiver is also available with a mounting diameter of 30 mm (BC-3). Consult your Burleigh representative for more information. Optical mounts with X-Y as well as q, f adjustments may be convenient in some laboratory configurations. However, basic q, f adjustments are usually sufficient.
Page 12
Once the Wavemeter is aligned with the HeNe tracer beam, it should be possible Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-3 Burleigh Park, Fishers, NY 14453 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
To activate the Average Mode, press the AVERAGE button on the lower front panel. The button lights when it is in the “on” position. Press the button again to return to normal operation. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-4 Burleigh Park, Fishers, NY 14453...
For lasers that are spectrally broad, including multi-line lasers, the measurement will be terminated as soon as the fringe envelope falls below 500 millivolts peak-to-peak. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-5 Burleigh Park, Fishers, NY 14453...
. The Auto-Resolution circuitry computes the appropriate readout based on the accuracy implicit in the number of fringes that can be counted. Insignificant digits are eliminated and replaced by a “_” character on the display. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-6 Burleigh Park, Fishers, NY 14453...
Page 16
12 wavenumbers according to the graph and would be displayed with 4 digit resolution as 654.3_ nm. Figure 3.4. Auto-Resolution Conversion Chart. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-7 Burleigh Park, Fishers, NY 14453 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Hexadecimal) to the “Receive Data” (RXD:pin#5) line. Data transmission can be restarted (Xon) by sending a "^Q”character (11 Hexadecimal) to RXD. An alternative method to control data transmission is provided through the “Clear Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-8 Burleigh Park, Fishers, NY 14453...
6. Adjust the end mirror alignment until the spots observed on the white card are superimposed. Interference fringes will be evident when the spots are superimposed. Adjust the alignment until only one broad fringe exists. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-9 Burleigh Park, Fishers, NY 14453...
Page 19
If you have difficulty obtaining enough signal to activate the bar-graph meter, connect an oscilloscope to the MONITOR and TRIGGER outputs for greater sensitivity in detecting the fringe signal. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-10 Burleigh Park, Fishers, NY 14453...
Page 20
Wavemeter Operating Manual 3. Operating Instructions Figure 3.6. Wavemeter Exploded View. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-11 Burleigh Park, Fishers, NY 14453 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 21
Wavemeter Operating Manual 3. Operating Instructions Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-12 Burleigh Park, Fishers, NY 14453 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
4. Disconnect the 6-conductor ribbon cable from the Detector Board and remove the single mounting screw that secures the Board. Carefully lift the Detector Board from the black plastic guide at its base. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-13 Burleigh Park, Fishers, NY 14453...
8. Introduce a HeNe laser input to the beam receiver and realign the interferometer as described in Section 3.11. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-15 Burleigh Park, Fishers, NY 14453 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 25
Check again the relative timing of the peak of the MONITOR signal envelope and the start of the TRIGGER pulse for the broad linewidth laser. Repeat the above procedure if further correction is necessary. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 3-16 Burleigh Park, Fishers, NY 14453...
This message will appear preceding the countdown from 10 when power is first turned on. If this message appears the memory chip should be replaced. Please consult Burleigh for more information. 232Er Data transmission via the RS-232 interface has been halted by a CTS line “false”...
MONITOR signal, reduce the input power until the amplitude noise following the fringe envelope is less than 0.5 volts peak-to-peak. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 4-3 Burleigh Park, Fishers, NY 14453 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
The adapter clip must be removed and replaced by a second fuse. Figure 4.2. Fuse Holder. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 4-3 Burleigh Park, Fishers, NY 14453 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
However, basic q, f adjustments are usually sufficient. NOTE: The beam receiver is also available with a mounting diameter of 30 mm (Product Code, BC-3). Consult your Burleigh representative for more information. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 4-3 Burleigh Park, Fishers, NY 14453 Artisan Technology Group - Quality Instrumentation ...
Wavemeter jr by a special fiber optic cable that adapts that style of connector to the FC input connector on the WA-2500. Contact your Burleigh representative for more information. Burleigh Instruments, Inc. (04659-M-00 Rev E)
** Input power requirement is specified as power out of the fiber and into the instrument. For a collimated beam with a diameter 2.5 mm, coupling into the standard beam receiver is usually greater than 30%. Burleigh Instruments, Inc. (04659-M-00 Rev E) page 4-3 Burleigh Park, Fishers, NY 14453...
Burleigh’s warranty card. The beam receiver and fiber optic cables are warranted for a period of three months from date of delivery. During the warranty period, Burleigh will repair or, at its option, replace parts which prove defective when the instrument is returned prepaid to Burleigh Instruments, Inc.
// The function returns a 16-bit integer whose high-order byte contains status // bits. The meaning of the low-order byte varies, depending on the service // value. The high-order bits have the following meanings: Burleigh Instruments, Inc. (04659-M-00 Rev E) page A-1 Burleigh Park, Fishers, NY 14453...