3-4. Replacement of Periodic
Replacement Part

3-4-1. Replacing Lamp

When the lamp does not light, replace the lamp according
to the following procedures.
c
Air-cool the lamp before replacing it.
The lamp is still heated immediately after turning the
power off, so there is a danger of causing burns with
touching it. When replacing the lamp, turn off the power
and wait for 5 minutes. After confirming the lamp power
indicator stops its flashing, turn off the AC power supply
which supplies the power to the FVS-1000, and wait for
about twenty minutes until the lamp is air-cooled, then
replace the lamp.
w
Observe the followings when installing or replacing the
lamp.
If not, the lamp can burst and may result in injury.
. Do not hit the lamp against something, apply the exces-
sive pressure to it or hurt it.
. Do not give a vibration or shock to the lamp.
. Do not touch the glass ball and reflecting mirror of the
lamp with bare hands.
. Do not use the not-specified lamp.
Preparation
Turn off the AC power supply which supplies the power to
the FVS-1000.
n
To prevent an electric shock, be sure to turn off the power
of the AC power supply which supplies the power to the
FVS-1000 before removing.
FVS-1000 MMP1

3-4. Replacement of Periodic Replacement Part

Removal
1. Removing the gate base assembly
Remove the gate base assembly. (Refer to Section 3-2-1.)
Gate base assembly
2. Removing the lamp house assembly
(1) Loosen the four screws (with stopper).
Screws
(with stopper)
Fixing screws
Screws
(with stopper)
3-15