Casella Noise Dosimeter Model CEL – 350 Instructions
**IMPORTANT - Retrieve the data from the dosimeters prior to returning to Travelers Laboratory**

Initial Equipment Checks

Visually inspect all equipment upon receipt.
For best results, charge the dosimeters overnight prior to use. The status indicator on the dosimeter will flash
when the unit is charging and is solid
that charging is required before the dosimeter can be used.
Additional assistance is available by:
Viewing the Casella Noise Dosimeter Model CEL-350 Instructional Video at:
Calling the Instrument Laboratory at (800) 842-0355 (Eastern time zone 7:00 am to 5:00 pm)

Dosimeter Calibration (calibrate immediately prior to sampling)

Turn the dosimeter on by pressing the LEFT button.
Remove the windscreen by carefully turning the black plastic ring at the base of
the windscreen counter clockwise. Do not twist the foam windscreen – it will
tear.
Turn the calibrator on with the
emitting a tone before proceeding to the next step.
Insert the calibrator onto the microphone so that o-ring in the calibrator forms a
seal with the dosimeter microphone and the dosimeter screen is still visible.
The display will indicate "CALIBRATE? 114", press the RIGHT button on the
dosimeter to confirm calibration. "CALIBRATING" with an indicator across the
bottom of the screen will appear on the display.
When completed, the display will read "CAL OK" for a few seconds. Press the
RIGHT button until the current sound level data (LAS) is displayed (about 114
dB). Record this data as your PRE sampling calibration verification and remove
the calibrator from the microphone.
Carefully replace the windscreen onto the dosimeter microphone by gently turning the black plastic ring clockwise.
Do not twist the foam windscreen – it will tear.
Repeat calibration for the remainder of dosimeters. The calibrator will turn off automatically.

Collecting Noise Samples

To begin data collection, press the LEFT and RIGHT buttons simultaneously until the display counts down "start 3, 2,
1". Release the buttons when the battery/memory indicators and the recording indicator "►" appear on the display.
Optionally, the dosimeter can be placed in secure mode by holding the RIGHT button and pressing the LEFT button
three times. The locked icon ( ) will be visible in lower left corner of the display.
Attach the dosimeter to the clothing of the employee at the shoulder, near the employee's ear using the clips.
When sampling is done, remove dosimeter from the employee. If it was previously placed in secure mode, remove it
from secure mode by holding the RIGHT button and pressing the LEFT button three times. The locked icon ( ) will
disappear, the recording indicator "►" will still be displayed.
when the charging is complete. "BATTERY FAIL" on the display indicates
blue
red
button. Ensure that the calibrator is
PAGE 1
REDUCE RISK. PREVENT LOSS. SAVE LIVES.
Travelers - Instructional Videos
RISK CONTROL
red

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CEL-350 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Casella CEL-350

  • Page 1  Additional assistance is available by:  Viewing the Casella Noise Dosimeter Model CEL-350 Instructional Video at: Travelers - Instructional Videos  Calling the Instrument Laboratory at (800) 842-0355 (Eastern time zone 7:00 am to 5:00 pm) Dosimeter Calibration (calibrate immediately prior to sampling) ...
  • Page 2 RISK CONTROL Casella Noise Dosimeter Model CEL – 350 Instructions  To stop data collection, press the LEFT and RIGHT buttons simultaneously until the display counts down “stop 3, 2, 1”. Release the two buttons when the battery/memory indicators appear on the display. The recording indicator “►”...
  • Page 3 RISK CONTROL Casella Noise Dosimeter Model CEL – 350 Instructions INDUSTRIAL HYGIENE SURVEY DATA – NOISE DOSIMETRY SAVE THIS DATA SHEET AND EMAIL A COPY TO Tate Berkan at TBerkan@Travelers.com Survey Date: Company Name: Name: Street Address: Email Address: City, State, Zip Code...

Table of Contents