Tips; Troubleshooting - Industrial Test Systems Spa eXact EZ Quick Start Manual

Advanced photometer system
Table of Contents

Advertisement

The Spa eXact
®
Each test menu can store 20 results. To retrieve the stored results, go to the desired
test using the MENU key. When the desired test is displayed, press and hold down the
MENU key. Continue holding down the MENU key to scroll the stored results for that
test, starting with the most recent result. The meter will display, from memory, the last
20 readings in sequence beginning with -20, which is the latest result, followed by -19,
which is the 2nd latest result, etc; and fi nally -01, which is the oldest result retained.
Only the last 20 readings are stored in each menu. This meter is able to store 100
results in memory (20 in each menu).
Before testing, rinse the meter sample cell with the sample water 3 times.
Spa water samples should be taken 18 inches below the surface. Submerge the meter
with open cell facing down 18 inches then turn meter upright at that depth to fi ll the cell.
Remove meter from water with the sample for testing.
Always fi ll the cell to capacity (4ml); be careful not to splash liquid over the side.
Test immediately after fi lling the cell with the water sample.
To obtain optimal accuracy when testing outdoors (sunlight), use the cell cover when
zeroing and reading the sample.
Due to the strip slitting process, you may fi nd one or two strips that are noticeably
smaller or larger in width than the normal strips in the bottle. These should be
discarded. Using these strips may give unreliable results.
Meter is not compatible for use with powder pillows, tablets, or liquids from other
manufacturers.
Dip strip for entire countdown.
Each eXact
Strip Micro is valid for ONLY one test. Discard strip after use.
®
Dry the outside of the meter before storage.
Remove batteries before storing for prolonged periods.
Store the meter and test materials out of direct sunlight and away from chemical
storage areas.
Minimize exposure of meter and test reagents to heat above 90°F (32°C).
When installing batteries, verify the O-ring is still attached to the screw before
tightening.
Cleaning the cell with water and brush after each test is recommended for best
accuracy and prior to storage of unit.
Issue
Dim screen or no
response from meter
LO on LCD Display
while zeroing
HI on LCD Display
while reading
LO on LCD Display
while reading
LO fl ashes on LCD
Display, then Err
HI fl ashes on LCD
Display, then Err
Total Chlorine result
less than Free Chlorine
EZ photometer has a 5 minute automatic shutdown timer.

TROUBLESHOOTING

Cause
Low battery
Low battery
Dirty cell
Cloudy sample
Bad LED
Result above
detection level
Result below
detection level
Improper test procedure
for Combined Chlorine
Combined Chlorine result
above detection level
Improper test procedure

TIPS

Solution
Replace batteries
Replace batteries
Clean cell (see pg.15)
Dilute sample or use sieve
Contact ITS.
Re-run test to verify result
Re-run test to verify result
Re-read test procedure
and follow directions carefully
Re-run test to verify result
Dilute and re-run test
Re-read test procedure
and follow carefully
11