Page 1
3119-600 Series Temperature Controlled Chambers M55-16547-EN Revision E Operating Instructions The difference is measurable ®...
Page 2
® is a registered trademark of Illinois Tool Works Inc. (ITW). Other names, logos, icons and marks identifying Instron products and services referenced herein are trademarks of ITW and may not be used without the prior written permission of ITW.
Page 8
Chapter: Introduction Upper Removable Wedge Pullrod Port Power Indicator High Temperature Warning Label Control Panel Heated Circulating Window Internal Light Door (Not Shown) Handle Lower Removable Wedge Figure 1. 3119-600 Chamber Components Caution Eyebolts are included with the chamber to enable initial positioning and installation of the chamber only.
Configuration parameter errors can seriously impair the performance of the controller or cause damage to the chamber. The procedures detailed in this manual have been validated by Instron. You are advised to use only these specific procedures and in the manner documented. You are advised not to change any parameters of the controller other than those detailed in this manual.
See “Safety” on page for illustrations of all hazard labels and their meaning. The following labels are visible on all 3119-600 series chambers: Label Location Purpose Door front High temperature hazard warning label - warns user about above window high temperature in vicinity.
Wherever it is considered that safety is compromised, press the fan button OFF on the touch panel controller or press the mains circuit breaker OFF (O) on the rear panel. Product Support: www.instron.com...
Chapter: Safety Carefully read the operating instructions and observe all Warnings and Cautions. The term Warning is used where a hazard may lead to injury or death. The term Caution is used where a hazard may lead to damage to the equipment or to loss of data. Ensure that the system set-up and the test being performed constitute no hazard to operators or others (see general guidelines on page 27).
Disconnect the unit from the power supply before removing any panel/cover or changing any fuse. Repairs should only be carried out by an Instron Service Engineer, contact your local Instron representative if you suspect an electrical fault. Disconnect equipment from the electrical power supply before removing any electrical safety covers or replacing fuses.
Hazard - The operator MUST consult their supplier’s Material Safety Data Sheet on the particular cryogenic gas being used before using this equipment at sub-ambient temperatures. Instron advise that you receive training in the handling of cryogenic gases from your supplier.
Data Sheets” on page 17. Repairs should only be carried out by an Instron Service Engineer. The glass fibre insulation used in the chamber is retained within its panels and does not represent a hazard in normal use. Contact your local Instron representative if any glass fibre insulation is exposed.
Always de-pressurise pipe work before disconnection. Instron advise that you receive training in the handling of high pressure gases from your gas supplier. Never disconnect high pressure pipe work until you are sure that it is safe to do so.
Material Safety Data Sheets Material Safety Data Sheets This section contains information to allow you to obtain current Material Safety Data Sheets (MSDS) directly from the manufacturers. Instron is not responsible for the content or accuracy of these MSDSs. Material...
Page 18
Chapter: Safety Material Location Manufacturer MSDS information SUPERWOOL 607 SUPERWOOL 607 THERMAL CERAMICS LIMITED Blanket Blanket is contained Tebay Road, Bromborough between all chamber Wirral, Merseyside CH62 3PH outer panels and inner case. United Kingdom Phone: +44 (0) 151 334 4030 SUPERWOOL SUPERWOOL Paper Fax: +44 (0) 151 334 1684...
(see “Rear Input Panel Connections and Labels” on page 22). Door Handle The door handle has a two stage operation, i.e. latch and seal. See “Opening and Closing the Chamber Door” on page for details. Product Support: www.instron.com...
Chapter: Function of Controls and Indicators Control Panel Buttons and Indicators The control panel buttons and indicators are described below. Eurotherm 3208 Temperature Controller Upper Indicators Upper Display Lower Display Lower Indicators Down Button Page Button Up Button Scroll Button Cooling Button Heating Button (Indicator above)
Page 21
“Operating the Chamber” chapter. display Indicator Not used by any procedures detailed in the chamber manuals. If illuminated contact your local Instron representative. Indicator Displays ORANGE when communications between the chamber and a computer are active – see “Operating the Chamber”...
Chapter: Function of Controls and Indicators Table 2. Control Panel Buttons and Indicators Indicator Type Function /Control Heating The button allows the heater to operate on demand from the controller. button and The indicator illuminates GREEN when the heater is ON. indicator Cooling button The button allows the cooling valve to operate on demand from the...
Page 23
Meaning Caution - equipment may be damaged if you disconnect the power cable while power is on. Always switch off the electrical supply to the chamber before disconnecting the power cable from the chamber rear input panel. Product Support: www.instron.com...
Chapter: Function of Controls and Indicators Removable Wedges The chamber is supplied with removable wedges, to allow easy removal of the chamber from the test area without dismantling the load string. Caution The wedge section is fragile and therefore should always be handled with care. Instrumentation Cable Recess Hand Recess Pullrod Aperture...
Configuration parameter errors can seriously impair the performance of the controller or cause damage to the chamber. The procedures detailed in this chapter have been validated by Instron. You are advised to use only these specific procedures and in the manner documented. You are advised not to change any parameters of the controller other than those detailed in this chapter.
Chapter: Operating the Chamber configuration parameters being modified. Instron does not accept liability for configuration changes other than those documented in this manual. When using the procedures described in this chapter for the Eurotherm 3208 controller, the controller will time out and return to the home page if there is a delay of more than 30 seconds between key presses.
2. Turn the handle another 90º counter clockwise. This unlatches the door. 3. Swing the door open. To close the door: 1. Ensure the door handle is turned counterclockwise as far as it will go and is vertical. Product Support: www.instron.com...
Chapter: Operating the Chamber 2. Swing the door closed until the door seals touch the frame. 3. Turn the handle 90º clockwise. This engages the door latch. 4. Turn the handle a further 90º. You will see the door pull into the frame. This compresses the door seal.
For more details on the alarm and setpoint ramp rate settings and how to change them, see page and page respectively. Switching the Chamber Light On and Off Once the chamber is ready for use, you can switch the chamber light on or off as required. Product Support: www.instron.com...
Chapter: Operating the Chamber The interior light is designed for use at any temperature within the range of the chamber. 1. Press the light button to switch on the chamber light. The indicator above the button illuminates GREEN to show the light is ON. 2.
After testing at sub-ambient temperatures and before maintenance, heat the chamber to 100ºC for 30 minutes or until all moisture caused by frost has been driven off. Then return the chamber to ambient as described above. Product Support: www.instron.com...
Page 32
Chapter: Operating the Chamber Caution If you do not allow the chamber to return to ambient temperature after testing below 0ºC, ice may form on the inside of the chamber when you open the door. M55-16547-EN...
Set the chamber to a cooling cycle. The solenoid valve opens and the pressure in the pipe work decays to ambient atmospheric pressure. 2. In all cases, when the chamber reaches ambient temperature turn it OFF using the contact breaker switch on the rear panel of the chamber. Product Support: www.instron.com...
• Arrange for calibration verification, traceable to national or international standards. Contact your local Instron representative. If the chamber is used routinely below 0°C then the thermocouple characteristics can change more rapidly. In this case it is recommended that thermocouple calibration is checked more frequently.
TO ADJUST ALARM 2 SETPOINT A2.LO TO ADJUST DISPLAY UNITS UNITS TO VIEW OPTIONS SETPOINT RATE LIMIT SP.RAT TO ADJUST TIMER CONFIGURATION TM.CFG TO VIEW OPTIONS When TM.CFG = ProG, Figure 6 Figure 5. Eurotherm 3208 Controller Menu Product Support: www.instron.com...
Chapter: Operating the Chamber Changing the Alarm Settings The chamber is delivered with the alarms set 10ºC either side of the operating range of the chamber, i.e. -110ºC to +360ºC. You can change these alarm values if, for example, you always operate within a much smaller range or you wish to protect a delicate specimen.
If you wait a few seconds, the text scrolls across the screen DISPLAY UNITS 2. Press the up or down buttons, , until the units you want are shown in the upper display. There is a choice of %C, %F, %k, (%). nonE PErc Product Support: www.instron.com...
Chapter: Operating the Chamber Caution Do not select as these are not supported and you will not be able to nonE PErc control the chamber temperature. 3. Wait a few seconds and the chosen units will flash once. This indicates the selection is accepted.
To turn on the timer/programmer: 1. From the home page, press the scroll button a number of times until TM.CFG is shown in the lower display. If you wait a few seconds, the text scrolls across the TIMER CONFIGURATION screen. Product Support: www.instron.com...
Page 40
Chapter: Operating the Chamber 2. Press either the up button or the down button until appears in the ProG controller upper display. 3. Wait a few seconds. The display flashes once to show the value is accepted. 4. Press the page button to return to the home page.
If you wait a few seconds, the text scrolls across the screen. TARGET SETPOINT 1 2. Press the scroll button appears in the lower display. Tsp.1 If you wait a few seconds, the text scrolls across the screen. Target setpoint1 Product Support: www.instron.com...
Chapter: Operating the Chamber 3. Press either the up button or the down button to set the target setpoint for section 1 of the program. 4. Press the scroll button appears in the lower display. rmp.1 If you wait a few seconds, the text scrolls across the screen.
7. Verify the installation was successful using the instructions provided with your Instron software. If you are prompted to select a COM port number follow the instructions below. Leave all other COM port settings as their defaults (Baud Rate: 9600 and Address/Unit ID: 1).
Page 44
Expand Ports (COM & LPT) and look for a series of 4 ports named USB Serial Port as shown in Figure 8. Note the highest COM port number in the series of 4. This may be required by supported Instron software. ® Figure 8. Windows...
Analog Output The 0-10V DC retransmission of the chamber temperature can be monitored and recorded using certain Instron equipment and software. If required, the output can be used to calibrate an external device. See “Calibrating an External Device using the Controller’s DC Output”...
Page 46
Chapter: Operating the Chamber • Check all connections and ensure that all covers are in place before re- connecting the power supply. If the fault is not resolved by the following fault finding table, contact your local Instron representative. Table 5. Fault Finding...
Page 47
1.5 bar for LN ). Below these pressures there will be a significant reduction in performance. If the problem persists, contact your local Instron representative. Lights do not illuminate • Check bulbs and replace if necessary. If the problem persists, contact your local Instron representative.
Page 48
Chapter: Operating the Chamber M55-16547-EN...
Appendix: Additional Configuration Parameters Changing User Levels 3208 Controller Access Level Configuration The 3208 controller fitted to the 3119-600 series of chambers has the following levels of access: • Level 1 - The procedures accessible for this level are described in previous chapters.
There are additional menu choices when has been configured to prog. Refer to tm.cfg Figure 6 for details. You are advised to return the chamber controller to level 1 when you have set the level 2 parameters to your requirements. Product Support: www.instron.com...
Page 52
Appendix: Additional Configuration Parameters HOME AUTO - TUNE ENABLE A.TUNE TO ADJUST CONTROL PROPORTIONAL BAND, etc. PARAMETERS TO ADJUST CURRENT RECIPE NUMBER REC.NO TO ADJUST OUTPUT 3 FUNCTION 3.FUNC Read Only ALARM 1 SETPOINT A1.HI TO ADJUST ALARM 2 SETPOINT A2.LO TO ADJUST DISPLAY UNITS...
OFF. a.tUNe The new control terms are permanent until either a new auto tuning process is undertaken or the function is enabled (see page 55). rec.no Product Support: www.instron.com...
Configuration parameter errors can seriously impair the performance of the controller or cause damage to the chamber. Instron recommend that you do not change any of the parameters listed in Figure If you find that the factory settings or the auto tuning function (see...
The upper display will show nonE. If you wait a few seconds, the text scrolls across the screen. current recipe number 3. Press the up or down buttons, to select in the upper display. dcAL Product Support: www.instron.com...
Appendix: Additional Configuration Parameters 4. Press the scroll button once. appears in the lower display. 3.func If you wait a few seconds, the text scrolls across the screen. output 3 function Check that the upper display shows w.SP. If it does not, repeat steps to 4.
Appendix: Hints and Tips for Using Chambers Occasionally CO is a more appropriate choice for the following reasons: can be stored for long periods of time without loss, whereas LN • continually lost even when stored in a Dewar flask that is not in use. This could be an advantage for occasional users of CO has higher cooling power than LN per unit volume, so the chamber air...
• To avoid unwanted pre-loading of specimens due to grip closure or thermal expansion during the soak period, use the “Specimen Protect” feature available with some models of Instron test system. Optimising Your Testing • For optimum window clarity during low temperature testing always switch on the window heaters (if fitted) for 40 minutes prior to running your tests.
Page 60
Appendix: Hints and Tips for Using Chambers • The air circulation from the centrifugal fan within the inner chamber and from the impeller near the motor, is shown in Figure 10. The internal air is directed in a figure of eight from the heaters behind the baffle over the load train pull rods of the testing machine, across the specimen and back to the internal impeller via the alarm and control thermocouples.
Page 61
Certain items such as rubber-coated jaw faces have reduced temperature ranges compared to the rest of the load string. The chamber may be capable of exceeding the temperature rating of the load string so ensure that temperature limits and alarms are set accordingly. Product Support: www.instron.com...
Page 62
Appendix: Hints and Tips for Using Chambers M55-16547-EN...
Page 63
......31 to setpoint ......30 Product Support: www.instron.com...
Page 64
indicators and controls ....20 power indicator ..... . 19 preventive maintenance .
Need help?
Do you have a question about the 3119-600 Series and is the answer not in the manual?
Questions and answers